765-787-0015 info@medisurg.com

Constellation
Alumni Page

THE INDUSTRY’S LEADING, INDEPENDENT EXPERTS IN EYE SURGICAL EQUIPMENT.

Training Alumni

Please email or call with any part orders.

Do not hesitate to call or email Medisurg with any tech support!

317-663-3158

medisurg@medisurg.com

“Great course, great instructor. The staff was great to talk to, ask questions to, and very helpful”

Merrell Williams
Aramark

"The information was taught in a very easy to understand way with lots of hands on and with all of the tools and manual to give me full confidence in servicing the Infinity system."

Cory Ladd
UCSF

"Very fun, hands on, and involved course. Knowledgeable staff that are very kind and thoughtful."

Jarrod Huffer
Trimedx

How we can help you?

Lorem ipsum dolor sit amet, consectetuer adipiscing elit.

855.233.4050

Original Part #
212-0000-XXX
212-1083-XXXAssy, Footpedal, Six Switch
212-5002-XXXAssy, Table Top Console
212-3398-XXXAssy, Display, LCD 17
212-1010-XXXAssy, Module, Host
212-1016-XXXAssy, Power Controller, Tabletop
212-1023-XXXAssy, Module, Pneumatic AK 6.103
212-1037-XXXAssy, Module, U/S-Diat Aqua L 6.124
212-3564-XXXAssy, Module, US/Diat IOL
212-1040-XXXAssy, Filter, Oil/Air Dryer
212-1758-501Assy, Ballast, Illum Xenon
212-3245-XXXAssy, Fluidics
212-3334-552ASSY,PCB,SUPERVISOR MODULE
212-2981-XXXAssy, Gas AGF, External
212-5003-XXXAssy, Base, Premium
212-1007-XXXAssy, Laser
562-1034-XXXAssy, Core Module
562-1564-XXXAssy, Core Module
562-1041-XXXAssy, Engine, Laser
212-1008-XXXAssy, Illuminator, Aux
212-2150-501Assy, Ballast, Aux Illuminator
212-2164-XXXAssy, Power Dist, Cart
212-1004-XXXAssy, Arm, Tray
212-1191-XXXAssy, Arm, Lower
212-1158-XXXAssy, Tray
212-1186-XXXAssy, Arm, Upper
562-1564-501Assy, Module, Core
562-1510-001Assy, Footpedal, Laser

Constellation CR3


Event CodeEvent TypeMessage to UserDetail / Discrepancy
1001FaultFault - 1001 Call Field ServiceThe OSE Gateway exited unex- pectedly.
1002FaultFault - 1002 Call Field ServiceThe OSE Gateway has reported that the Supervisor process has gone down.
1003FaultFault - 1003 Call Field ServiceSomething has failed in the communication between the Host and the Supervisor.
1004FaultFault - 1004 Call Field ServiceThe Main Application exited un- expectedly.
1005FaultFault - 1005 Call Field ServiceThe Host Controller checksum validation failed to pass all files under its control.
1006FaultFault - 1006 Call Field ServiceThe Host has failed to connect to the Supervisor process.
1008FaultFault - 1008 Call Field ServiceThe Host Controller failed to launch an executable image during startup.
1009FaultFault - 1009 Call Field ServiceThe Host Controller could not communicate with the Power Module over the serial link.
1010FaultFault - 1010 Call Field ServiceComponent or Submodule re- ported an incorrect software, hardware, firmware or other.
1011FaultFault - 1011 Call Field ServiceHost Controller communications was lost as the result of a .NET IPC failure.
1012FaultFault - 1012 Call Field ServiceThe Host has not received any heartbeats from the Supervisor within the required timeframe.
1013FaultFault - 1013 Call Field ServiceThe Host has detected that the User Interface thread has not responded within the required timeframe (i.e. the UI is consid- ered locked up).
1014FaultFault - 1014 Call Field ServiceThe system has detected a soft- ware error.
1015FaultFault - 1015 Call Field ServiceThe Host has received a mes- sage from the Supervisor that it doesn’t recognize (this is a spe- cific type of software error).
1016FaultFault - 1016 Call Field ServiceThe Host application’s display fonts can not be loaded.
1017FaultFault - 1017 Call Field ServiceA failure has occurred during the playback of an audio file.
1103ErrorVoice Activation not available.The user tried activate voice ac- tivation, but the voice activation process could not be found.
1107ErrorUnable to load one or more language packs.An error occurred while at- tempting to load a non-English language.
1108ErrorIncompatible version numbers within the Fluidics submodule. Fluidics functions will be dis- abled.The Fluidics module was shut down due to an incompatible version. The log file shows ex- actly what version information was incompatible.
1109ErrorIncompatible version numbers within the Pneumatics sub- module. Pneumatics functions will be disabled.The Pneumatics module was shut down due to an incompat- ible version. The log file shows exactly what version information was incompatible.
1110ErrorIncompatible version numbers within the Ultrasound submod- ule. Ultrasound and Diathermy functions will be disabled.The Ultrasound module was shut down due to an incompat- ible version. The log file shows exactly what version information was incompatible.
1111ErrorIncompatible version numbers within the Auxiliary Illuminator submodule. Auxiliary Illumina- tor functions will be disabled.The Auxiliary Illuminator mod- ule was shut down due to an incompatible version. The log file shows exactly what version information was incompatible.
1112ErrorIncompatible version numbers within the Laser submodule. Laser functions will be dis- abled.The Laser module was shut down due to an incompatible version. The log file shows ex- actly what version information was incompatible.
1200AdvisoryFunction is not allowed when the fluidics subsystem is not functional.The user tried to invoke a com- mand which is only allowed when the fluidics subsystem is functional.
1202AdvisoryFunction is not allowed when the Laser is in Ready Mode.The user tried to invoke a com- mand which is not allowed when the laser is in Ready Mode or Firing.
1203AdvisoryFunction is not allowed when the footswitch treadle is down or buttons are pressed.The user tried to invoke a com- mand which is only allowed when the treadle is up and no buttons pressed.
1204AdvisoryFlow mode is not available un- til the probe or handpiece has been primed.Flow mode is unavailable be- cause a probe or handpiece hasn’t been primed.
1205AdvisoryFlow mode is only avail- able for 20 gauge probes / handpieces.Flow mode is unavailable be- cause a non 20 gauge probe or handpiece is used.
1206AdvisoryPlease prime the Vit probe.The user presses the treadle in a Vit Step when the Vit Probe hasn’t been primed.
1207AdvisoryPlease connect a Phaco hand- piece.The user presses the treadle in a Phaco Step when no Phaco handpiece is connected.
1208AdvisoryPlease tune the Phaco hand- piece.The user presses the treadle in a Phaco Step when the Phaco handpiece isn’t tuned.
1209AdvisoryPlease connect a Frag hand- piece.The user presses the treadle in a Frag Step when no Frag handpiece is connected.
1210AdvisoryPlease tune the Frag hand- piece.The user presses the treadle in a Phaco Step when the Phaco handpiece isn’t tuned.
1211AdvisoryFunction is not allowed when the Ultrasound submodule is not functional.The user tried to invoke a com- mand which is only allowed when the ultrasound subsystem is functional.
1212AdvisoryFunction is not allowed when the cassette is not ready.The user tried to invoke a com- mand which is only allowed when the cassette is ready.
1213AdvisoryInfusion must be on to use VFC Extract. Press [Ignore] to allow VFC. Extract case. without infusion for this caseThe user presses the treadle in a VFC Extract Step when infu- sion is off.
1214AdvisoryPleas connect a VFC syringe.The user presses the treadle in a VFC Step when no VFC Sy- ringe is connected.
1215AdvisoryPlease connect ForcepsThe user presses the treadle in a Forceps Step when no forceps are connected.
1216AdvisoryPleas connect ScissorsThe user presses the treadle in a Scissors Step when no Scis- sors are connected.
1217AdvisoryPleas connect an AGF syringeThe user presses the “Start” but- ton in the Auto Gas Filling dialog when no AGF syringe has been connected.
1218AdvisoryForceps not available during AGF: please try again when AGF is complete.The user presses the treadle in a Forceps Step while Auto Gas Filling is in progress.
1219AdvisoryVFC not available during AGF: please try again when AGF is complete.The user presses the treadle in a VFC Step while Auto Gas Fill- ing is in progress.
1220AdvisoryEjecting the cassette is not allowed while the footswitch treadle is pressed.The user attempts to eject the cassette while the treadle is down.
1221AdvisoryEjecting the cassette is not allowed while infusion is on.The user attempts to eject the cassette while infusion is on.
1222AdvisoryEjecting the cassette is not allowed while irrigation is on.The user attempts to eject the cassette while irrigation is on.
1223AdvisoryEjecting the cassette is not allowed while priming, tuning, or testing.The user attempts to eject the cassette while tests (priming/ tuning/testing) are in progress.
1224AdvisoryCleaning the cassette is not allowed while infusion, irrigation, or FAX is on.The user attempts to start Cas- sette Cleaning when either Infu- sion, Irrigation or FAX is on.
1225AdvisoryCleaning the cassette is not allowed without a functional cassette.The user attempts to start Cas- sette Cleaning when the Cas- sette isn’t available (e.g. not inserted).
1226AdvisoryPriming the cassette is not allowed without a functional cassette.User attempt to prime the cas- sette when preconditions aren’t met (e.g. no cassette inserted, cassette not tested or didn’t pass the test).
1227AdvisoryAn error occurred loading the device settings. The system will revert to default values.A database attempt to load the device settings from the file system has failed. Reverting to defaults.
1228AdvisoryCommand is not allowed while infusion is on.The user tried to invoke a com- mand which is only allowed when infusion is not on.
1229AdvisoryCommand is not allowed while irrigation is on.The user tried to invoke a com- mand which is only allowed when irrigation is not on.
1230AdvisoryThe selected step is not sup- ported with the current cas- sette type.The user attempts to select a Step whose mode/submode isn’t compatible with the current operating mode.
1231AdvisoryThe connected probe is not supported with the current cassette type.The user connects a probe (with RFID) that’s not compatible with the current operating mode (e.g. an Ultra Vit probe when an Anterior cassette is connected or Ultra Vit Anterior probe when a Posterior cassette is con- nected).
1232AdvisoryPlease connect footswitch.The Supervisor has reported that no footswitch is connected.
1233AdvisoryThe connected footswitch is not supported.The Supervisor has detected the connection of an unknown footswitch.
1234AdvisoryFlow limit is currently not avail- able due to flow mode being unavailabe.Vacuum mode flow limit de- pends on the availability of flow mode. If flow mode becomes unavailable then the flow limit function can not be performed by the fluidics submodule.
1235AdvisoryIV Pole pressure mode is cur- rently not available.User attempts to select IV Pole as the infusion source but the IV Pole is not available (not config- ured, not working or Cassette is of Premium type)
1236AdvisoryGravity pressure mode is cur- rently not available.User attempts to select Gravity as the infusion source when a Premium Cassette is inserted.
1237AdvisoryIOP Compensation is currently not available.User attempts to turn IOP Com- pensation on when it’s not avail- able.
1238AdvisoryAn error occurred saving the device settings.User attempt to save device set- tings fails.
1239AdvisoryProportional reflux mode is currently not available.User attempts to toggle into pro- portional reflux mode when the fluid level in the cassette cham- ber is out of range or the fluid level gets out of range while in proportional reflux.
1240AdvisoryExtraction flow mode is cur- rently not available.User attempts to toggle into flow mode when the fluid level in the cassette chamber is out of range or the fluid level gets out of range while in flow mode.
1241AdvisoryInfusion source is getting low: please check the bottle.The infusion container fluid level is getting low.
1242AdvisoryCommand is not allowed due to instrument not being avail- able.The user tried to invoke a com- mand which is not allowed when no instrument is available for the command to work with.
1243AdvisoryInfusion / Irrigation source is empty: please press [Change] and replace the bottle.The infusion fluid container is empty but there’s still fluid in the cassette chambers.
1244AdvisoryCommand is not allowed while priming, tuning, or testing.The user attempts to do some- thing, e.g. change Steps, that’s not allowed when tests are in progress.
1245AdvisoryPlease connect handpiece.The user attempts to perform an action, e.g. priming, that requires a handpiece but no handpiece is connected. Note: This is only applicable to handpieces the system can de- tect connection for (e.g. Phaco)
1247AdvisoryConnected handpiece is not supported with the current cassette type.User connects a handpiece that’s not compatible with the current Cassette type. (e.g., a Frag handpiece is connected when an Anterior Cassette is inserted)
1248AdvisoryIs the inserted cassette new?The user has inserted a Cas- sette which the system can’t de- termine whether it’s new or the same Cassette that was previ- ously ejected.
1249AdvisoryAn error occurred parsing the log file.An error occurred parsing one of more lines of the log file.
1250AdvisoryMulti-Cut is not available when proportional scissors are selected. Press [Multi-Cut] to indicate that multi-cut scissors are currently connected.The user has selected scis- sors of type Proportional and attempts to select the Multi- Cut submode or attempt to use momentary cutting in Extrusion Mode.
1251AdvisoryPress [Multi-Cut] to indicate that multi-cut scissors are currently connected.The user tried to turn Infusion, Irrigation or FAX on while the cassette is being cleaned.
1252AdvisoryThe scanned barcode is not recognized.The user has scanned an item that the system doesn’t recog- nize.
1255AdvisoryIlluminator fiber is not con- nectedUser has attempted to turn on an illuminator with no illuminator fiber connected.
1256AdvisoryProportional reflux mode is not available when the footswitch treadle is de- pressed.User attempts to enter Propor- tional Reflux mode when the treadle is down
1257AdvisoryThe report’s header or foot- er has too many rows to fit on the page.The user has added too many rows to a header or footer table being edited in an End Case report. As a result, the table will not be printed out in its entirety on the various pages compris- ing the report.
1258AdvisoryThe report’s current table has too many rows to fit on the page.The user has added too many rows to the current table being edited in an End Case report. As a result, the table will not be printed out in its entirety on the various pages comprising the report.
1259AdvisoryCommand is not allowed while proportional diathermy is ac- tive.The user attempts to enter pro- portional reflux while propor- tional diathermy is active.
1260AdvisoryCommand is not allowed while proportional reflux is active.The user attempts to enter pro- portional diathermy while pro- portional reflux is active.
1261AdvisoryCassette can’t be ejected while being cleaned.The user attempts to eject the cassette while it’s being cleaned.
1262AdvisoryInvalid handpiece tip.The system attempts to select a tip that’s not valid for the specif- ic handpiece. This will only hap- pen if an incorrect tip is speci- fied for a scanned pak, i.e. the user will never be able to select an invalid tip from the UI.
1263AdvisoryPort can’t be selected: there is no probe connected.The user tries to make a laser port the active port but there’s no probe connected to that port.
1264AdvisoryPort can’t be selected: the probe type for the port isn’t valid.The user tries to make a laser port the active port but the cur- rently selected probe type for that port is invalid.
1266AdvisoryA laser probe is not connected to the active port.Shown in any of the following situations: • Laser Step entered and no probe connected to the active port • Probe removed from active port • User tries to go to Ready Mode when no probe is connected to the active port
1267AdvisoryA valid laser probe is not se- lected for the active port.Shown in any of the following situations: Laser Step entered and no valid probe type selected for the active port. User tries to go to the Ready Mode when no valid prove type is selected for the active port.
1268AdvisoryThe laser remote interlock is open.Shown in any of the following situations: • Laser Step entered and the Interlock is open • Interlock opened in a Laser Step • User tries to go to Ready Mode when the Interlock is open
1269AdvisoryLaser Dr. Filters are not con- nected to the console. Are all necessary Dr. Filters properly installed and con- nected?Shown in any of the following situations: • Laser Step entered and the Dr. Filters haven’t been verified for the active port • Endo Probe inserted to the active port while in a Laser Step • User tries to go to Ready Mode when Dr. Filters haven’t been verified for the active port.
1270AdvisoryOne Laser Dr. Filter is con- nected to the console. Are all necessary Dr. Filters properly installed and con- nected?Same as above
1271AdvisoryTwo Laser Dr. Filters are con- nected to the console. Are all necessary Dr. Filters properly installed and con- nected?Same as above
1273AdvisoryLaser Dr. Filter 1 is disen- gagedThe user has disengaged the (connected) Dr. Filter 1 while in a Laser Step.
1275AdvisoryLaser Dr. Filter 2 is disen- gagedSame as 1273 but for Dr. Filter 2
1276AdvisoryCommand is not allowed when Laser is firing.User has attempted to change one of the laser’s settings while the laser is firing.
1277AdvisoryPort can’t be selected: it’s not functional.User has attempted to select a laser port that’s not functional.
1278AdvisoryCannot go to Laser Ready Mode while the current screen is being displayed.User has pressed the Ready button from the footswitch when a “conflicting screen” is active.
1279AdvisoryNo laser footswitch is con- nected.User enters a Laser Step when no laser footswitch is connected or the laser footswitch is discon- nected in a Laser Step.
1280AdvisoryUnable to write to the report file.A problem has occurred when trying to write to the specified Report file.
1281AdvisoryUnable to read from the report file.A problem has occurred when trying to read from the specified Report file.
1285AdvisoryThe report file and its associ- ated CRC value do not match.The Report file and its associ- ated CRC value do not match.
1288AdvisoryCan’t eject the cassette while reflux is active.The user attempts to eject the cassette while reflux (of any type) is active. For proportional reflux, this ad- visory is displayed when the user attempts cassette ejection while proportional reflux mode is active with the treadle not depressed. Cassette ejection while the treadle is depressed will generate advisory 1220.
1289AdvisoryExtraction flow mode is not available when FAX is turned on.FAX was turned on while in Flow Mode or a Posterior Step with Flow Mode preference was selected when FAX was on. The system then changed to Vacu- um Mode.
1290AdvisoryThe surgical function is cur- rently unavailable.The user has requested surgi- cal functionality that is currently unavailable. This is a default ad- visory that is only displayed for cases in which a more meaning- ful explanation is not available.
1296AdvisoryThe barcode reader failed to initialize.The bar code reader failed to initialize during startup.
1297AdvisoryThe system is currently low on free disk space. Please backup or remove non-critical files.During startup, free space was detected at less than 15%.
1298AdvisoryA printer is not installed. The printer is out of paper.The user tried to print but no printer is installed. The user sent a print job but the printer is out of, or ran out of paper.
1299AdvisoryThe printer is off line.The user requested a print job, but the printer is off line.
1302AdvisoryAn unknown error occurred while printing.Generated by any printer error other the errors above.
1303AdvisoryThe Illuminators are currently turned off from the footswitch and cannot be turned on or off at this time.The user has attempted to turn on or off an endo illuminator when the footswitch Momentary Endo Illuminators Off function is active.
1304AdvisoryThe Laser is currently unable to deliver the maximum power level of 2 Watts. The maximum Laser power (Watts) currently available is:Is generated when the power available from the laser drops below the maximum level (2 Watts).
1305AdvisoryAn error occurred loading a surgeon. The system will re- vert to default values.A database attempt to load a surgeon from the file system has failed. Reverting to defaults.
1306AdvisoryAn error occurred setting the printer port. The system will revert to the previous value.Attempt to set the printer IP ad- dress failed.
1307AdvisoryAn error occurred loading a procedure. The system will re- vert to default values.A database attempt to load a procedure from the file system has failed. Reverting to de- faults.
1308AdvisoryAn error occurred loading the system settings. The system will revert to default values.A database attempt to load the system settings from the file system has failed. Reverting to defaults.
1309AdvisoryAn error occurred saving a surgeon’s settings. Changes will be reverted at the start of the next case.A database attempt to save a surgeon to the file system has failed.
1310AdvisoryAn error occurred saving a procedure. Changes will be reverted at the start of the next case.A database attempt to save a procedure from the file system has failed. Reverting to de- faults.
1311AdvisoryAn error occurred saving the system settings. The system will use default values.A database attempt to save the system settings from the file system has failed. Reverting to defaults.
1312AdvisoryAn error has occurred initializ- ing Video Overlay. Video Over- lay functions will be disabled.An error happened during Vid- eoOverlay initialization.
1313AdvisoryAn error has occurred in the Video Overlay component. Video Overlay functions will be disabled.An error happened during Vid- eoOverlay processing.
1314AdvisoryThere was an error saving sur- geon data.There was an error saving sur- geon data.
1317AdvisoryThere was an error renaming the surgeon.There was an error renaming the surgeon.
1318AdvisoryThere was an error renaming the procedure.There was an error renaming the procedure.
1319AdvisoryThere was an error renaming the case data.There was an error renaming the case data.
1320AdvisoryThere was an error deleting the surgeon.There was an error deleting the surgeon.
1321AdvisoryThere was an error deleting the procedure.There was an error deleting the procedure.
1322AdvisoryThere was an error deleting the case data.There was an error deleting the case data.
1323AdvisoryThere was an error loading the surgeon.There was an error loading the surgeon.
1324AdvisoryThere was an error loading the procedure.There was an error loading the procedure.
1325AdvisoryThere was an error loading the case data.There was an error loading case data.
1326AdvisoryUnable to load the case report template.There was an error loading the Case Report Template.
1329AdvisoryUnable to save the techni- cian’s log.There was an error saving the Technician’s Log.
1330AdvisoryUnable to load the video table.There was an error loading the Video Table.
1331AdvisoryUnable to play the video.There was an error loading and\ or playing the Video file.
1332AdvisoryUnable to load helpThere was an error opening the Help pdf file.
1333AdvisoryUnable to save the case Info.There was an error saving the Case Info file.
1334AdvisoryUnable to update the log fileThere was an error writing to the Log File.
1335AdvisoryUnable to read the log fileThere was an error reading from the Log File.
1336AdvisoryUnable to write the incident file.There was an error writing to the Incident File.
1337AdvisoryUnable to read the incident fileThere was an error reading from the Incident File.
1338Advisoryunable to write the system metric fileThere was an error writing to the System Metrics File.
1339AdvisoryUnable to read the system metrics fileThere was an error reading from the System Metrics File.
1340AdvisoryUnable to writ to the removable drive.There was an error while at- tempting to write to the removable drive.
1341AdvisoryFurther increasing the output level in air can damage fiber tips. Would you like to con- tinue?The user has exceeded Thresh- old 1 (see SRS-5687.1) illumi- nator value by either turning on the illuminator, changing the setpoint while the illuminator is on, or changing Procedures while the illuminator is on.
1342AdvisoryFurther increasing the output level will reduce exposure time by 35%. Would you like to con- tinue?The user has exceeded Thresh- old 2 (see SRS-5687.3) illumi- nator value by either turning on the illuminator, changing the setpoint, or changing Proce- dures.
1343AdvisoryFurther increasing the output level in air can damage fiber tips. Also, further increasing the output level will reduce exposure time by 35%. Would you like to continue?The user has exceeded both Threshold 1 and Threshold 2 (see SRS-5687.1 and SRS- 5687.3) illuminator value by ei- ther turning on the illuminator, changing the setpoint while the illuminator is on, or changing Procedures while the illumina- tor is on.
1344AdvisoryOnly two illuminators can be turned on simultaneously.User attempts to turn on an illu- minator port when two ports are already on. This is not allowed.
1345AdvisoryAGF not allowed while For- ceps or VFC is in use.User attempts to start Auto Gas Filling when either Forceps or VFC is in use (i.e. selected and treadle down).
1346AdvisoryAn error occurred trying to ac- cess the wireless network.An error occurs when the user attempts to access the wireless network.
1349AdvisoryA RAID hard drive has failed or is missingOne Disk in a redundant RAID volume is missing or has failed but the volume is still function- al.
1350AdvisoryThe probe in the current laser port is not supported.The type of probe in the current port is not supported by NGVS but might be supported by NGL (slit lamp for example).
1355AdvisoryCommand is not allowed when the Pneumatics submodule is not functional.The user tried to invoke a com- mand which is only allowed when the Pneumatics subsys- tem is functional.
1356AdvisoryCommand is not allowed when an Anterior Only Cassette is being used.The user tried to invoke a com- mand which is not allowed with an Anterior Only Cassette.
1357AdvisoryCannot switch to Infusion: the cassette has not been primed.The user tried to switch to Infu- sion when the cassette was not primed.
1358AdvisoryCurrent probe type is unrecog- nized. Please select a valid probe type.User has attempted to turn on an illuminator when the probe type is “unrecognized”.
1359AdvisoryPower module communication error. When powering down the system, you will have to press the “Options/Shutdown” button on the screen.Host controller failed to commu- nicate with power module.
1361AdvisoryCommand is not allowed when Infusion backup pressure is active.The user tried to invoke a com- mand which is only allowed when Infusion backup pressure is not active.
1362AdvisoryCommand is not allowed when the probe type has been iden- tified by RFID.The user tried to change the type of a laser probe that has been identified by RFID.
1363AdvisoryCommand is not allowed when the probe type has been pre- viously identified by the user. Disconnect and then recon- nect the probe to change its type.The user tried to change the type of a laser probe that has been identified by a user selec- tion. Only one selection is allowed after connection.
1364AdvisoryIOP Compensation is not allowed without a premium cas- sette inserted. Please insert a premium cassette and retry.The user tried to turn on IOP compensation when a premium cassette is not inserted.
1365AdvisoryIOP Compensation is not allowed with an uncalibrated cassette. Please calibrate cas- sette and retry.The user tried to turn on IOP compensation with a cassette that has not been calibrated.
1366AdvisoryCommand is not allowed when FAX is on and the cassette has not been primed.The user tried to perform a Test Instrument command when FAX is on and the cassette is not primed.
1367AdvisoryFAX is not allowed when the cassette is being primed.The user tried to turn on FAX when the cassette is being primed.
1369AdvisorySetting values is not allowed when the cassette is in the chamber overflow condition.The cassette chamber is in an overflow condition and pressure cannot be controlled.
1370AdvisoryTest instrument is not allowed in this mode.User has attempted to perform a test instrument command in Setup or End Case.
1371AdvisoryThe remote control battery is low.User has pressed on a remote control button and the remote’s battery is running out of power.
1372AdvisoryInfusion pressure will drop to zero during calibration. Con- tinue?User has pressed the “Cali- brate” button on the advisory popup brought up when there’s been an IV Pole error. Since I/V Pole calibration will change the current Infusion pressure, the user is warned before calibra- tion is started.
1751InformationPower recovery is in progress. Please wait until surgical func- tions become available.AC power has been restored and the recovery process has started.
1752InformationFault - 2100 Call Field ServiceA two second heart beat reply from the host was missed.
1753InformationCommunications failure with the Fluidics submodule. Fluid- ics functions will be disabled.The supervisor cannot establish communicate with the Fluidics submodule.
2100FaultCommunications failure with the Fluidics submodule. Fluid- ics functions will be disabled.The supervisor lost communi- cate with the Fluidics submod- ule
14300InformationUnable to perform this function without a primed cassette.This advisory is produced when extraction is attempted without a primed cassette.
2201InformationUnable to aspirate. Please turn on infusion.This advisory is produced when extraction is attempted without infusion on.
2202AdvisoryUnable to aspirate while infu- sion/irrigation is unavailable.This advisory is produced when extraction is attempted without infusion being functional.
2203AdvisoryPlease wait: draining cassette.This advisory is produced when extraction is attempted when the extraction chamber is in a overflow condition.
2204AdvisoryUnable to turn on infusion with- out a primed cassette.This advisory is produced when infusion is attempted with out a primed cassette.
2205AdvisoryUnable to turn on infusion with- out a calibrated IV pole.This advisory is produced when infusion is attempted with out a calibrated IV Pole.
2208AdvisoryUnable to turn on infusion with- out sufficient source pressure.This advisory is produced when infusion is attempted with out source air pressure
2209AdvisoryUnable to turn on irrigation without a primed cassette.This advisory is produced when irrigation is attempted with out a primed cassette.
2210AdvisoryUnable to turn on infusion with- out a calibrated IV pole.This advisory is produced when irrigation is attempted with out a calibrated IV Pole.
2211AdvisoryUnable to turn on irrigation without sufficient source pres- sure.This advisory is produced when irrigation is attempted with out suitable source pressure.
2212AdvisoryIrrigation unavailable: out of fluid. Please change the bot- tle.This advisory is produced when irrigation is attempted with out sufficient irrigation fluid avail- able
2213AdvisoryFAX unavailable. Please insert a cassette.This advisory is produced when F/AX is attempted with out a cassette inserted.
2214ErrorExtraction setpoint timeout. In- fusion/Irrigation and Extraction functions will be disabled.This error is produced when the extraction is active and the extraction proxy fails to get a footswitch update within 20 msec
2215AdvisoryMicro reflux is currently not available.This error is produced when the the user attempt to activate mi- cro reflux and the fluid level is out of range.
2250ErrorCommunications failure with the Pneumatics submodule. Pneumatics functions will be disabled.The supervisor cannot com- municate with the Pneumatics submodule.
2251ErrorCommunications failure with the Pneumatics submodule. Pneumatics functions will be disabled.This error is produced when the Supervisor losses communica- tion with the pneumatics mod- ule.
2252AdvisoryUnable to turn on cutting with- out sufficient source pressure.This advisory is produced when cutter or utility operation is requested with out sufficient source pressure.
2253AdvisoryUnable to turn on cutting with excessive source pressure.This advisory is produced when cutter or utility operation is re- quested with source pressure that is too high.
2254ErrorPneumatics setpoint timeout. Pneumatics functions will be disabled.This error is produced when the pneumatics is active and the pneumatics proxy fails to get a footswitch update within 20 msec.
2300ErrorCommunications failure with the Ultrasound submodule. Ultrasound functions will be disabled.The supervisor cannot commu- nicate with the Ultrasound sub- module.
2301ErrorCommunications failure with the Ultrasound submodule. Ultrasound functions will be disabled.The supervisor lost communi- cation with the Ultrasound sub- module.
2302ErrorUS setpoint timeout. US and Diathermy functions will be disabled.This error is produced when the US is active and the US proxy fails to get a footswitch update within 20 msec.
2350ErrorCommunications failure with the Laser submodule. Laser functions will be disabled.The supervisor cannot commu- nicate with the Laser submod- ule.
2351AdvisoryCommunications failure with the Laser submodule. Laser functions will be disabled.The supervisor lost communica- tion with the Laser submodule.
2400AdvisoryPlease insert the Table Top Il- luminator drawer.Supervisor detected a non –zero setpoint with the Table Top Illuminator drawer out.
2401AdvisoryThe lamp in the Table Top Illu- minator needs to be replaced. Please contact Field Service.Supervisor detected a non–zero setpoint with the Table Top Illu- minator Lamp bad.
2500AdvisoryCommunications failure with the Power Control submodule. Please contact Field Service.Supervisor detected a serial I/O error when trying to talk to the Power Module.
2550ErrorFootswitch error. Footswitch treadle functions will be dis- abled.The footswitch has an error.
2600AdvisoryIV Pole over-current error. IV Pole functions will be dis- abled.IV Pole motor is drawing too much current.
2700ErrorCommunications failure with the Auxiliary Illuminator sub- module. Auxiliary Illuminator functions will be disabled.The supervisor cannot commu- nicate with the Auxiliary Illumi- nator submodule.
2701AdvisoryCommunications failure with the Auxiliary Illuminator sub- module. Auxiliary Illuminator functions will be disabled.The supervisor lost communica- tion with the Auxiliary Illuminator submodule.
2702AdvisoryThe lamp in the Auxiliary Illu- minator needs to be replaced. Please contact Field Service.- Supervisor detected a non– zero setpoint with the Auxiliary Illuminator Lamp bad.
2750FaultFault - 2750 Call Field ServiceThe Supervisor could not assert reset control over all submod- ules. A module that was sup- posed to be reset responded with its module information.
2751FaultFault - 2751 Call Field ServiceThe Supervisor got duplicate slot ID for the submodules.
3100AdvisoryCould not calibrate the fluid level sensors. Please eject and reinsert, or replace the cassette.Level Sensor calibration failed. The calibration done status bit was not set by the FPGA within the specified timeout limit or the calculated required max/min pixel gain was outside the valid range.
3200ErrorCassette latch error. Please remove and reinsert the cas- sette.Cassette latch hardware feed- back signals indicate a hard- ware problem.
3201AdvisoryInvalid cassette ID detected. Please eject and insert a valid cassette.An invalid cassette id was read by the cassette id sensors.
3202AdvisoryCassette test failed. Please eject and reinsert, or replace cassette.Cassette pressure and/or vac- cum tests failed.
3203AdvisoryThe cassette was not properly latched into position. Please remove and reinsert the cas- sette.The cassette latch optical po- sition sensor indicates that the latch did not reach its locked position.
3204ErrorCassette ID sensor error. Infu- sion/irrigation and extraction functions will be disabled.Cassette ID sensor test failed. The sensor output voltage is not within the expected range.
3205AdvisoryThe Fluidics module fan is not working.The Fluidics module fan tachom- eter indicates that the Fluidics module fan is not operating.
3300AdvisoryDraining cassette. Please wait.The cassette aspiration cham- ber is full of fluid.
3302AdvisoryN/ANot enough or too much fluid in the aspiration chamber to al- low aspiration flow mode and reflux. NOTE: This advisory is never explicitly displayed. Instead ad- visory 1239 or 1240 might be displayed if appropriate. See 11239 and 11240 for details.
3304AdvisoryLeak test failure. Please con- firm the irrigation tubing; as- piration tubing , and the test chamber are properly connect- ed to the handpiece.A leak in the aspiration or irriga- tion tubing was detected during flow check of a handpiece.
3305AdvisoryPriming of the aspiration probe was unsuccessful. Please at- tempt to re-prime the probe.Probe prime was unsuccess- ful. The required volume of fluid was not transferred through the aspiration tubing set within the specified timeout period.
3306AdvisoryPriming of the aspiration handpiece was unsuccessful. Please attempt to re-prime the handpiece.Handpiece prime was unsuc- cessful. The required volume of fluid was not transferred through the aspiration tubing set within the specified timeout period.
3307ErrorAspiration level sensor prob- lem detected. Infusion/irriga- tion and extraction functions will be disabled.The level sensor status signals indicate a level sensor hard- ware problem.
3308AdvisoryFlow check failure: measured flow restriction is too high. Ex- traction and Ultrasound func- tions in Phaco/Frag will be disabled.The handpiece failed the flow- check. Too high vacuum level was required to achieve the ref- erence flow level.
3309AdvisoryFlow check failure: aspiration chamber could not be filled with fluid. Extraction and Ultra- sound functions in Phaco/Frag will be disabled.The handpiece failed the flow- check. Too high vacuum level was required to achieve the ref- erence flow level.
3325AdvisoryDrain bag is almost full. Please replace bag and press [Done].~50 cc remaining volume in drain bag.
3326AdvisoryDrain bag is full. Please re- place bag and press [Done].~0 cc remaining volume in drain bag.
3327AdvisoryDrain bag is critically full. Please replace bag and press [Done].Drain bag filled > 50 cc above capacity.
3329ErrorDrain pump problem detected. Infusion/irrigation and extrac- tion functions will be disabled.The commanded pump rate does not correspond to the ac- tual pump rate measured by the optical encoder.
3330AdvisoryThe drain pump fan is not working.The drain pump fan tachometer indicates that the drain pump fan is not operating.
3331AdvisoryCould not drain the aspiration chamber. Please remove and insert a new cassette.The aspiration chamber could not be drained within the speci- fied timeout period.
3350ErrorExtraction pressure transducer offset error. Infusion/irrigation and extraction functions will be disabled.Extraction pressure transducer 0 offset out of range.
3351ErrorExtraction pressure transducer discrepancy error. Infusion/irri- gation and extraction functions will be disabled.A discrepancy between the pri- mary and redundant extraction pressure transducers was de- tected.
3352ErrorExtraction isolation valve error. Infusion/irrigation and extrac- tion functions will be disabledExtraction isolation valve hardware error.
3353ErrorReflux valve error. Infusion/irri- gation and extraction functions will be disabled.Extraction reflux valve hardware error.
3354ErrorExtraction output valve (S1) error. Infusion/irrigation and extraction functions will be dis- abled.Extraction output pincher valve (S1) error
3355ErrorExtraction output valve (S11) error. Infusion/irrigation and extraction functions will be dis- abled.Extraction output port 1 pincher valve error.
3356ErrorExtraction output valve (S22) error. Infusion/irrigation and extraction functions will be dis- abled.Extraction output port 2 pincher valve error.
3357ErrorExtraction cross-connection valve (SC) error. Infusion/irri- gation and extraction functions will be disabled.Extraction “normally open” cross-connection valve error.
3358ErrorExtraction cross-connection valve (SC2) error. Infusion/irri- gation and extraction functions will be disabled.Extraction “normally closed” cross-connection valve error.
3359AdvisorySuction pressure surges de- tected. Vacuum will be dis- abled. Please release the footswitch treadle to reset.Extraction pressure oscillations detected.
3360AdvisorySuction flow surges detected. Flow will be disabled. Please release the footswitch treadle to reset.Extraction flow oscillations detected.
3361AdvisorySuction pressure is too high. Vacuum will be disabled. Please release the footswitch treadle to reset.Extraction pressure overshoot detected.
3362AdvisoryAspiration flow too high. Flow will be disabled. Please re- lease the footswitch treadle to reset.Extraction flow overshoot detected.
3363ErrorExtraction pressure transducer reference voltage out of range. Infusion/irrigation and extrac- tion functions will be disabled.Extraction transducer reference voltage error.
3400ErrorInfusion source container pres- sure transducer offset error. In- fusion/irrigation and extraction functions will be disabled.Infusion source container pressure transducer 0 offset out of range.
3401ErrorInfusion source container iso- lation valve error. Infusion/irri- gation and extraction functions will be disabled.Infusion source container isola- tion valve error.
3402ErrorInfusion source container transducer discrepancy error. Infusion/irrigation and extrac- tion functions will be disabled.A dispcrepancy between the source container and LPAS pressure transducers was detected.
3403ErrorInfusion source container pressure too high. Infusion/irri- gation and extraction functions will be disabled.Infusion source container pres- sure too high.
3420ErrorInfusion pressure transducer offset error. Infusion/irrigation and extraction functions will be disabled.Infusion pressure transducer 0 offset out of range
3421ErrorInfusion pressure transducer discrepancy error. Infusion/irri- gation and extraction functions will be disabled.A discrepancy between the primary and redundant infusion pressure transducers was detected.
3422ErrorInfusion isolation valve error. Infusion/irrigation and extrac- tion functions will be disabled.Infusion isolation valve error..
3423ErrorInfusion FAX valve error. Infu- sion/irrigation and extraction functions will be disabled.Infusion FAX valve error
3424ErrorInfusion input valve error. Infu- sion/irrigation and extraction functions will be disabled.Infusion chamber input pincher valve error.
3425ErrorInfusion output valve error. In- fusion/irrigation and extraction functions will be disabled.Infusion chamber output pincher valve error.
3426ErrorIrrigation output valve error. In- fusion/irrigation and extraction functions will be disabled.Irrigation output pincher valve error.
3427ErrorInfusion pressure surges de- tected. Infusion/irrigation and extraction functions will be disabled.Infusion pressure oscillations detected.
3428ErrorInfusion pressure too high. In- fusion/irrigation and extraction functions will be disabled.Infusion high pressure detect- ed.
3429AdvisoryLow infusion pressure de- tected. Please check infusion connections. Select 30 mmHg backup pressure, or ignore low pressure condition. Infusion pressure too low.
3431AdvisoryLow irrigation pressure detected. Please check irrigation connections.Irrigation pressure too low.
3433ErrorInfusion chamber isolation valve error. Infusion/irrigation and extraction functions will be disabled.Infusion chamber isolation valve detected to be stuck closed during power up diagnostics.
3434AdvisoryInfusion chamber overflow error. Control of infusion pressure has possibly been lost. The current pressure, in mmHg, could be as high as:Infusion chamber overflow error. The fluid level in the infusion chamber reached above the overflow level.
3436ErrorFAX valve error. Infusion/irrigation and extraction functions will be disabled.F/AX valve detected to be stuck closed during power up diagnostics.
3437ErrorIinfusion level sensor error. Infusion/irrigation and extraction functions will be disabled.The level sensor status signals indicate a level sensor hardware problem.
3438AdvisoryNo more infusion fluid available. Press [Change] to change the infusion bottle.The infusion chamber in the cassette is empty.
3440AdvisoryNo more irrigation fluid available. Press [Change] to change the infusion/irrigation bottle.The irrigation chamber in the cassette is empty.
3442AdvisoryIrrigation chamber overflow detected. Control of irrigation pressure has possibly been lost. The current pressure, in mmHg, could be as high as:Irrigation chamber overflow error. The fluid level in the irrigation chamber reached above the overflow level.
3460ErrorInfusion backup valve error. Infusion/irrigation, extraction, and Ultrasound functions will be disabled.Infusion backup valve error.
3461ErrorInfusion LPAS pump error detected. Infusion/irrigation, extraction, and Ultrasound functions will be disabled.The infusion LPAS pump tachometer indicate that the LPAS pump is not operating correctly.
3462AdvisoryInfusion flow sensor communication error. IOP Compensation functions will be disabled.Infusion flow sensor communication error.
3464AdvisoryInfusion flow data invalid: IOP Compensation functions will be disabled. Check infusion tubing for air bubblesInfusion flow sensor readings are not valid.
3466AdvisoryInfusion flow sensor accuracy error. IOP Compensation functions will be disabled.Infusion flow sensor accuracy error. Flow sensor and chamber fluid volume measurements are out of range of each other. Detected during flow sensor calibration.
3467AdvisoryInfusion flow sensor signal amplitude is low. IOP Compensation functions will be disabled. Please eject and reinsert the cassette.Infusion flow sensor signal amplitude error. Detected during flow sensor calibration
3469AdvisoryTubing calibration offset error. IOP Compensation functions will be disabled. Please position the infusion cannula at the height of the center of the cassette and re-prime.Infusion tubing calibration 0 offset advisory. The infusion cannula was not positioned within the correct vertical range of the cassette during calibration.
3470AdvisoryCalibration verification error: calculated pressures using the acquired calibration profile are not within the expected range for the selected infusion cannula. IOP Compensation functions will be disabled. Please re-prime.Infusion tubing calibration check point advisory. The pressure drop calculated using the acquired calibration profile was not within the expected range.
3471AdvisoryNoisy calibration flow readings. IOP Compensation functions will be disabled. Please re-prime.Infusion tubing calibration standard deviation advisory. The standard deviation between the acquired flow measurements and the calculated calibration profile was larger than the specified max limit.
3472AdvisoryInfusion chamber leak detected. Please eject and replace the cassette.A leak in the infusion chambers was detected during priming.
3473ErrorInfusion pressure transducer reference voltage out of range. Infusion/irrigation and extraction functions will be disabled.Infusion transducer reference voltage error.
3474ErrorInfusion NIFS valve error detected. Infusion/irrigation and extraction functions will be disabled.Infusion NIFS valve error.
3475AdvisoryInfusion prime failed.Infusion prime failed. SEE CR4 3475 Details.
3476AdvisoryThe infusion chambers did not fill with fluid. Please check the infusion bottle and connections or press [Change] to replace the infusion bottle.Infusion chamber did not fill within the specified timeout period. Source container is out of fluid.
3477ErrorInfusion LPAS pressure transducer offset error. Infusion/irrigation and extraction functions will be disabled.Infusion LPAS pressure transducer 0 offset out of range.
3478ErrorInfusion LPAS pressure error. Infusion/irrigation and extraction functions will be disabled.Infusion LPAS source pressure too high.
3479AdvisoryInfusion LPAS pressure is low. Infusion pressure loss is possible.Infusion LPAS pump output pressure is low.
3481AdvisoryInfusion flow sensor is disconnected. IOP Compensation functions will be disabled.Infusion flow sensor is disconnected or the sensor connection has failed.
4100ErrorU/S voltage failure (+5 analog). Ultrasound and Diathermy functions will be disabled.US Kernel Analog 5 volt feedback is bad. US Submodule is non-functional.
4101ErrorU/S voltage failure (+2.5). Ultrasound and Diathermy functions will be disabled.US Kernel +2.5 volt feedback is bad. US submodule is non-functional.
4102ErrorU/S voltage failure (-2.5). Ultrasound and Diathermy functions will be disabled.US Kernel -2.5 volt feedback is bad. US submodule is non-functional.
4103ErrorU/S voltage failure (+8.5). Ultrasound and Diathermy functions will be disabled.US Kernel 2.5 volt feedback is bad. US submodule is non-functional.
4111ErrorUS failure: SPI driver write timeout. Ultrasound and Diathermy functions will be disabled.SPI driver timed out waiting for a write to complete. US submodule is non-functional.
4200AdvisoryHandpiece EEPROM CRC is invalid. Please replace the handpiece.The handpiece EEPROM CRC is invalid. Handpiece needs to be replaced.
4201AdvisoryOnly one US handpiece may be connected at a time. Please remove one of the handpieces.Two US handpieces are connected. One of the handpieces must be removed before the other handpiece can be used.
4202AdvisoryHandpiece current is too low. Please replace handpiece and re-tune.U/S handpiece current is too low. A short circuit in the handpiece can cause this.
4203AdvisoryHandpiece voltage is too low. Please replace handpiece and re-tune.U/S handpiece voltage is too low. An open circuit in the handpiece can cause this.
4204AdvisoryHandpiece power is too high. Please replace handpiece and re-tuneU/S handpiece power output is too high.
4206AdvisoryHandpiece power DC2DC output was out of range. Please release the footswitch treadle to reset. If problem persists, please contact Field Services.While powering the handpiece, the DC2DC voltage for handpiece power is out of range.
4207AdvisoryHandpiece was removed while powered. Please reconnect handpiece and re-tune.User has disconnected the handpiece while it’s being powered.
4208AdvisoryData in handpiece EEPROM is out of range. Please replace the handpiece.Data contained in the handpiece is out of range.
4209AdvisoryUnknown US handpiece connected. Please connect a known handpiece.The system has detected that a US handpiece was connected but it cannot determine its type.
4210AdvisoryUnsupported US handpiece connected. Please connect a supported handpiece.The system has detected the connection of a recognized US handpiece but that handpiece is not supported by this system. Handpiece is not tuned.
4220AdvisoryTune failure – attempted while handpiece was in air. Please re-tune the handpiece.The handpiece was tuned while in air. Handpiece is not tuned.
4221AdvisoryTune failure: handpiece was removed before tuning. Please connect a handpiece and re-tune.A handpiece tune was requested but no handpiece is connected.
4222AdvisoryTune failure: handpiece is an unknown type. Please connect a known handpiece and re-tune.A handpiece tune was requested but an unknown type of handpiece is connected.
4223AdvisoryTune failure: handpiece has a loose tip. Please tighten the tip and re-tune.The handpiece tip was loose when tuned. Handpiece is not tuned.
4224AdvisoryTune failure: handpiece current is low. Please replace handpiece and re-tune.The handpiece current was too low (open circuit). Handpiece is not tuned.
4225AdvisoryTune failure: handpiece voltage is low. Please replace handpiece and re-tune.The handpiece voltage was too low when tuned (short circuit). Handpiece is not tuned.
4226AdvisoryTune failure: handpiece frequency order error. Please re-tune.The series (low impedance) and parallel (high impedance) frequencies were out of order when the handpiece was tuned.
4228AdvisoryTune failure: handpiece series frequency margin error. Please re-tuneThe series tune frequency was too close to the tune start frequency while tuning the handpiece. The handpiece is not tuned.
4229AdvisoryTune failure: handpiece parallel frequency margin error. Please re-tune.The parallel tune frequency was too close to the tune end frequency while tuning the handpiece. The handpiece is not tuned.
4231AdvisoryTune failure: handpiece frequency bandwidth too low. Please re-tune.The difference between the series and parallel tune frequencies was too small while tuning the handpiece. The handpiece is not tuned.
4232AdvisoryTune failure: handpiece frequency bandwidth too high. Please re-tuneThe difference between the series and parallel tune frequencies was too large while tuning the handpiece. The handpiece is not tuned.
4234AdvisoryTune failure: handpiece DC2DC output out of range. Please re-tune. If problem persists, please contact Field Service.The DC2DC voltage was out of range while tuning the handpiece. The handpiece is not tuned.
4235AdvisoryTune failure: handpiece removed while tuning. Please connect handpiece and re-tune.The handpiece was removed while tuning. The handpiece is not tuned.
4240AdvisoryThe requested Frag continuous power is too high. The Power level will be limited.A request for more than 60% frag power was made while not in a pulsed mode. The power will be limited to 60%.
4250ErrorUltrasound failure: ADC calibration. Ultrasound functions will be disabled.The ADC feedback reading with power off for DC2DC voltage, handpiece voltage, or handpiece current was too high.
4300ErrorDiathermy failure: DC2DC output was out of range. Diathermy functions will be disabled.While powering the handpiece, the DC2DC voltage for handpiece power is out of range.
4301AdvisoryDiathermy power is too high. Please release the footswitch treadle / button and try again.Too much power was being delivered to the diathermy handpiece. Diathermy power is turned off. The operator must release the treadle/switch and the depress the treadle/switch to re-activate power.
5100ErrorBallast failure (voltage). Illuminator functions will be disabled.Ballast transducer has reading that is out of specified safety range.
5101ErrorBallast failure (current). Illuminator functions will be disabled.Ballast transducer has reading that is out of specified safety range.
5102AdvisoryLamp calibration data is corrupted: lamp needs to be calibrated. Please contact Field Service.The calibration data saved in the flash contains incorrect CRC checksum.
5103AdvisoryFailure to turn lamp on: lamp needs to be replaced. Please contact Field Service.Lamp status reported by hardware is not the same as software status.
5104AdvisoryThe lamp has exceeded its rated life. Please replace the lamp.The lamp has exceeded its rated life. The new lamp is expected.
5105AdvisoryThe lamp has exceeded its rated maximum life. Please install a new lamp immediately.The lamp has exceeded its rated safe life. The new lamp should be installed immediately.
5107AdvisoryThe Table-Top Illuminator drawer is ejected. Please close the drawer to continue.The module has been pulled out of drawer.
5109InformationThe calibration data for the Illuminator has changed: due to recalibration or replacement with a new unit.The calibration data is changed.
5200AdvisoryIlluminator optics temperature is high. The lamp will be turned off if the temperature continues to rise.Optics temperature is too high. Lamp is going to be shut down if temperature continues to rise.
5201ErrorIlluminator optics temperature has exceeded its limit. Illuminator functions will be disabled.Illuminator is shut down because Optics temperature is too high.
5202AdvisoryIlluminator optics fan is at full speed. Optics unit may be overheating.Warning that the Optics fan is full on.
5203ErrorIlluminator optics thermo-cut-off has been triggered. Illuminator functions will be disabled.Lamp is turned off because thermo cut-off.
5204AdvisoryCommunication failure with the Illuminator optics fan. The fan may not work properly.Communication Error with optics fan.
5300AdvisoryIlluminator ballast temperature is high. The lamp will be turned off if the temperature continues to rise.Ballast temperature is too high. Lamp is going to be shut down if temperature continues to rise.
5301ErrorIlluminator ballast temperature has exceeded its limit. Illuminator functions will be disabled.Illuminator is shut down because Ballast temperature is too high.
5302AdvisoryIlluminator ballast fan is at full speed. Ballast unit may be overheating.Warning that the Ballast fan is full on
5303ErrorIlluminator ballast thermo-cut-off has been triggered. Illuminator functions will be disabled.Lamp is turned off because thermo cut-off.
5304AdvisoryCommunication failure with the Illuminator ballast fan. The fan may not work properly.Communication Error with ballast fan.
5400ErrorLamp louver failure at port 1: unable to move to home position. Port 1 will be disabled.Step motor at port 1 failed to move to home position.
5401ErrorLamp Louver failure at port 2: unable to move to home position. Port 2 will be disabled.Step motor at port 2 failed to move to home position.
5402ErrorLamp Louver failure at port 1: unable to move to specified position. Port 1 will be disabled.Step motor at port 1 failed to move to specified position.
5403ErrorLamp Louver failure at port 2: unable to move to specified position. Port 2 will be disabled.Step motor at port 2 failed to move to specified position.
6101ErrorPressure reading is too high. Cutting and Pneumatics functions will be disabled.Pneumatics pressure is too high.
6201ErrorAir Pressure valves have high transition faults. Cutting and Pneumatics functions will be disabled.AirPressure valves have high level faults (fail to open).
6202ErrorAir Pressure valves have low transition faults. Cutting and Pneumatics functionals will be disabled.AirPressure valves have low level faults (fail to close).
6203AdvisoryAir Pressure inlet filter may be dirty and needs to be replaced. Please contact Field Service.The unusual high pressure drop cross the Air Filter. This is caused by either the Air Filter is too dirty and need to be replaced or the system may be leaking air.
6204AdvisoryInlet pressure is too low for the system. Please adjust the inlet pressure to higher than 58 psi.The Air source (wall pressure) is too low (below 58 psi) to turn on. On startup, Pneumatics turns on the pressure automatically if the pressure is high enough (above 58 psi).
6206AdvisoryInlet pressure is too high for the system. Please adjust the inlet pressure to less than 140 psi.The Air source (wall pressure) is too high (above 120 psi) to turn on. On startup, Pneumatics turns on the pressure automatically if the pressure is high enough (above 58 psi).
6208ErrorInlet pressure is unstable. Pneumatics functions will be disabled. Please restart the system.The Air source pressure (wall pressure) is unstable to cause the system unexpected to turn On/Off pressure.
6301ErrorCutting errors; valves have high transition faults (failed to open). Cutting functions will be disabled.Cutters valves have high level faults (fail to open).
6302ErrorCutting errors; valves have low transition faults (failed to close). Cutting functions will be disabled.Cutters valves have low level faults (fail to close).
6303ErrorCutting error: redundant transducers discrepancy error. Cutting functions will be disabled. Please contact Field Service.Cutters redundant transducers discrepancy error.
6304ErrorCutting error: redundant transducers calibration. Cutting functions will be disabled. Please contact Field Service.Cutters redundant transducers calibration error.
6305AdvisoryCutting pressure is oscillating beyond the specified range. Please release the footswitch treadle to reset.Cutters actual pressure is oscillating beyond the specified tolerance.
6306AdvisoryCutting pressure is surging beyond the specified tolerance. Please release the footswitch treadle to reset.Cutters actual pressure is surging beyond the specified tolerance.
6401ErrorUtilities valves have high transition faults (failed to open). Pneumatics functions will be disabled.Utilities valves have high level faults (fail to open).
6402ErrorUtilities valves have low transition faults (failed to close). Pneumatics functions will be disabled.Utilities valves have low level faults (fail to close).
6403ErrorUtilities redundant transducers discrepancy error. Pneumatics functions will be disabled. Please contact Field Service.Utilities redundant transducers discrepancy error.
6404ErrorUtilities redundant transducers calibration error. Pneumatics functions will be disabled. Please contact Field Service.Utilities redundant transducers calibration error.
6405ErrorVacuum redundant transducers discrepancy error. Pneumatics functions will be disabled. Please contact Field Service.Vacuum redundant transducers discrepancy error.
6406ErrorVacuum redundant transducers calibration error. Pneumatics functions will be disabled. Please contact Field Service.Vacuum redundant transducers calibration error.
6407AdvisoryUtilities pressure is oscillating beyond the specified range. Please release the footswitch treadle to reset.Utilities actual pressure is oscillating beyond the specified tolerance.
6408AdvisoryUtilities pressure is surging beyond the specified tolerance. Please release the footswitch treadle to reset.Utilities actual pressure is surging beyond the specified tolerance.
6501ErrorAuto Gas valves have high transition faults (failed to open). Pneumatics functions will be disabled.Auto Gas valves have high level faults (fail to open).
6502ErrorAuto Gas valves have low transition faults (failed to close). Pneumatics functions will be disabled.Auto Gas valves have low level faults (fail to close).
6503ErrorAuto Gas redundant transducers discrepancy error. Pneumatics functions will be disabled. Please contact Field Service.Auto Gas redundant transducers discrepancy error.
6504ErrorAuto Gas redundant transducers calibration error. Pneumatics functions will be disabled. Please contact Field Service.Auto Gas redundant transducers calibration error.
6505AdvisoryC3F8 bottle may be empty and needs to be replaced. Please press [Replaced] to confirm bottle replacement.Gas 1 bottle may be empty and needs to be filled up.
6506AdvisorySF6 bottle may be empty and needs to be replaced. Please press [Replaced] to confirm bottle replacement.Gas 2 bottle may be empty and needs to be filled up.
7100ErrorBallast failure (voltage). Illuminator functions will be disabled.Ballast transducer has reading that is out of specified safety range.
7101ErrorBallast failure (current). Illuminator functions will be disabled.Ballast transducer has reading that is out of specified safety range.
7102AdvisoryLamp calibration data is corrupted: lamp needs to be calibrated. Please contact Field ServiceThe calibration data saved in the flash contains incorrect CRC checksum.
7103AdvisoryFailure to turn lamp on: lamp needs to be replaced. Please contact Field Service.Lamp status reported by hardware is not the same as software status.
7104AdvisoryThe lamp has exceeded its rated life: lamp needs to be replaced. Please contact Field Service.The lamp has exceeded its rated life. The new lamp is expected.
7105AdvisoryThe lamp has exceeded its rated maximum life: lamp needs to be replaced. Please contact Field Service.The lamp has exceeded its rated safe life. The new lamp should be installed immediately.
7200AdvisoryIlluminator optics temperature is high. The lamp will be turned off if the temperature continues to rise.Optics temperature is too high. Lamp is going to be shut down if temperature continues to rise.
7201ErrorIlluminator optics temperature has exceeded its limit. Illuminator functions will be disabled.Illuminator is shut down because Optics temperature is too high.
7202AdvisoryIlluminator optics fan is at full speed. Optics unit may be overheating.Warning that the Optics fan is full on.
7203ErrorIlluminator optics thermo-cut-off has been triggered. Illuminator functions will be disabled.Lamp is turned off because thermo cut-off.
7204AdvisoryCommunication failure with the Illuminator optics fan. The fan may not work properly.Communication Error with optics fan.
7300AdvisoryIlluminator ballast temperature is high. The lamp will be turned off if the temperature continues to rise.Ballast temperature is too high. Lamp is going to be shut down if temperature continues to rise.
7301ErrorIlluminator ballast temperature has exceeded its limit. Illuminator functions will be disabled.Illuminator is shut down because Ballast temperature is too high.
7302AdvisoryIlluminator ballast fan is at full speed. Ballast unit may be overheating.Warning that the Ballast fan is full on.
7303ErrorIlluminator ballast thermo-cut-off has been triggered. Illuminator functions will be disabled.Lamp is turned off because thermo cut-off.
7304AdvisoryCommunication failure with the Illuminator ballast fan. The fan may not work properly.Communication Error with ballast fan.
7400ErrorLamp louver failure at port 3: unable to move to home position. Port 3 will be disabled.Step motor at port 1 failed to move to home position.
7401ErrorLamp louver failure at port 4: unable to move to home position. Port 4 will be disabled.Step motor at port 2 failed to move to home position.
7402ErrorLamp louver failure at port 3: unable to move to specified position. Port 3 will be disabled.Step motor at port 1 failed to move to specified position.
7403ErrorLamp louver failure at port 4: unable to move to specified position. Port 4 will be disabled.Step motor at port 2 failed to move to specified position.
8100ErrorLaser controller software error. Laser functions will be disabled.Need description.
8101ErrorLaser controller shutter open for too long error. Laser functions will be disabled.Need description.
8102ErrorLaser controller shutter unexpected error. Laser functions will be disabled.Need description.
8103ErrorLaser controller shutter open between firing error. Laser functions will be disabled.Need description.
8104ErrorLaser controller mirror incorrect position error. Laser functions will be disabled.Need description.
8105ErrorLaser controller mirror position unexpected error. Laser functions will be disabled.Need description.
8106ErrorLaser controller power over the limit error. Laser functions will be disabled.Need description.
8107AdvisoryLaser controller maximum current exceeded error. Laser functions will be disabled.Need description.
8108ErrorLaser controller port power detection error. Laser functions will be disabled.Need description.
8109ErrorLaser controller power detected at wrong port error. Laser functions will be disabled.Need description.
8110ErrorLaser controller power reading mismatch error. Laser functions will be disabled.Need description.
8111ErrorLaser controller LBO crystal limit error. Laser functions will be disabled.Need description.
8112ErrorLaser controller pump temperature limit error. Laser functions will be disabled.Need description.
8113ErrorLaser controller power detect digital I/O mismatch error. Laser functions will be disabled.Need description.
8114ErrorLaser controller footswitch digital I/O mismatch error. Laser functions will be disabled.Need description.
8115ErrorLaser controller footswitch no NC (normally closed) error. Laser functions will be disabled.Need description.
8117AdvisoryLaser controller internal parameter error.Need description.
8118ErrorLaser controller process hunt error. Laser functions will be disabled.Need description.
8119ErrorLaser controller process attach error. Laser functions will be disabled.Need description.
8120ErrorLaser controller start up time out error. Laser functions will be disabled.Need description.
8121ErrorLaser controller JEM CRC error. Laser functions will be disabled.Need description.
8122ErrorLaser controller kernel CRC error. Laser functions will be disabled.Need description.
8123ErrorLaser controller flash file error. Laser functions will be disabled.Need description.
8124AdvisoryLaser LBO diode over temperature error.Need description.
8125AdvisoryLaser LBO diode under temperature error.Need description.
8126AdvisoryLaser controller crystal over-temperature error.Need description.
8127AdvisoryLaser controller crystal under-temperature error.Need description.
8129AdvisoryLaser controller probe connection error.Need Description
8131ErrorLaser controller power monitor POST error. Laser functions will be disabled.Need Description
8132ErrorLaser controller +2.5 volt out-of-range error. Laser functions will be disabled.Need Description
8134ErrorLaser controller 12 volt power error. Laser functions will be disabled.Need Description
8135ErrorLaser shutter timer expired error. Laser functions will be disabled.Need Description
8138ErrorLaser controller diode thermal electric cooler error. Laser functions will be disabled.Need Description
8139ErrorLaser probe port process POST error. Laser functions will be disabled.Need Description
8140ErrorLaser controller startup timeout error (laser engine). Laser functions will be disabled.Need Description
8141ErrorLaser controller startup timeout error (shutter control process). Laser functions will be disabled.Need Description
8142ErrorLaser controller startup timeout error (supervisor process). Laser functions will be disabled.Need Description
8143ErrorLaser controller startup timeout error (peripheral management process). Laser functions will be disabled.Need Description
8144ErrorLaser controller startup timeout error (probe port process). Laser functions will be disabled.Need Description
8145ErrorLaser controller startup timeout error (laser system controller). Laser functions will be disabled.Need Description
8147ErrorLaser controller TMP crystal startup timeout error. Laser functions will be disabled.Need Description
8148ErrorLaser controller TMP diode startup timeout error. Laser functions will be disabledNeed Description
8149ErrorLaser controller Dr Filter error. Laser functions will be disabled.Need Description
8150ErrorLaser controller invalid maximum power error. Laser functions will be disabled.Need Description
8153ErrorLaser controller load module CRC error. Laser functions will be disabled.Need Description
8154ErrorLaser controller module requires calibration. Laser functions will be disabled.Need Description
8200AdvisoryLaser controller probe removed while firing.Need Description
8202AdvisoryLaser controller Dr. Filter disengaged while firing.Need Description
8203AdvisoryLaser controller Dr. Filter disconnected while firing.Need Description
8204AdvisoryLaser controller interlock opened while firing.Need Description
8205AdvisoryLaser controller footswitch disconnected while firing.Need Description
8207AdvisoryLaser controller port changed in Ready State.Need Description
8208AdvisoryLaser controller Dr. Filter disengaged in Ready State.Need Description
8209AdvisoryLaser controller Dr. Filter disconnected in Ready State.Need Description
8210AdvisoryLaser controller interlock opened in Ready State.Need Description
8211AdvisoryLaser controller footswitch removed in Ready State.Need Description
8212AdvisoryLaser controller footswitch engaged when Ready requested.Need Description
8213AdvisoryLaser controller user data CRC error.Need Description
8214AdvisoryLaser controller Ready State denied: the filter is disengaged.Need Description
8215AdvisoryLaser controller Ready State denied: there is no interlock.Need Description
8216AdvisoryLaser controller Ready State denied: there is no valid probe for the active port.Need description.
8217AdvisoryLaser controller Ready State denied: the footswitch is depressed.Need description.
8218AdvisoryAre all necessary Dr. Filters properly installed and connected?User requests going to Ready Mode when the doctor filters haven’t been verified.
8219AdvisoryLaser controller Ready State denied: there is no footswitch present.User presses the Laser Ready button when no footswitch is connected.
8220AdvisoryLaser controller Dr. Filter connected while in Transition, Ready or Firing State.User connects an engaged Dr. Filter while in Transition / Ready or Firing mode and and Endo probe is connected to the active port.

Constellation CR4


Event CodeEvent TypeMessage to UserDetail / DiscrepancySystem Action
X001ErrorSub module system software error detected.OSE_ERROR_CODE; Internal OSE RTOS error.Application Software - Reload; Controller PCB.
X005ErrorSub module failure (+5V).KRNL_5_VOLTS_IS_BAD; Kernel +5 Voltage is bad.Controller PCB
X006ErrorSub module failure (+3.3V).KRNL_3_3_VOLTS_IS_BAD; Kernel +3.3 Voltage is badController PCB
X008ErrorSub module application failed to load.KRNL_PRG_LOAD_FAILED; The kernel failed loading a load module component. Attempted to load application code from Host and it failed.Application Software - Reload; Controller PCB
X009ErrorSub module application failed to start.KRNL_PRG_START_FAILED; The kernel failed starting a load module component. Software has downloaded, but module failed to runController PCB
X013ErrorSub module software failure (watchdog timeout).KRNL_WD_TIME_OUT; Watchdog had a task timeout.Controller PCB
X015ErrorSub module failure (voltage checker).KRNL_VOLTAGE_CHECKER_UNDERRUN; Kernel Voltage checker ADC underrun.Controller PCB
X017ErrorSub module failure (+24V).KRNL_24_VOLTS_IS_BAD; Kernel +24 Voltage is bad.Power Cable; Controller PCB
X018ErrorSub module failure (+1.5V).KRNL_1_5_VOLTS_IS_BAD; Kernel +1.5 Voltage is bad.•       Controller PCB
X024ErrorSub module failure (+1.2V).KRNL_1_2_VOLTS_IS_BAD Kernel +1.2 Voltage is bad.•       Controller PCB
X025ErrorSub module failure (reference voltage A).KRNL_REF_A_IS_BAD; Kernel Voltage Reference A is bad.•       Controller PCB
X026ErrorSub module failure (reference voltage B).KRNL_REF_B_IS_BAD; Kernel Voltage Reference B is bad.•       Controller PCB
X027ErrorSub module failure (reference voltage C).KRNL_REF_C_IS_BAD; Kernel Voltage Reference C is bad.•       Controller PCB
X028ErrorSub module failure (+2.5V).KRNL_2_5_VOLTS_IS_BAD; Kernel +2.5 Voltage is bad.•       Controller PCB
X029ErrorSub module failure (FPGA).KRNL_FPGA_NOT_READY; The FPGA ready bit is not set.•       Controller PCB
X030ErrorSub module failure (communication error).KRNL_SUPERVISOR_COMM_LOSS; The Kernel Status Handler lost communication with the Supervisor module.Ethernet Cable; Supervisor PCB
X033ErrorSub module failure (watchdog test).KRNL_WD_INIT_TEST_FAILED; The watchdog startup test failed on power up.•       Controller PCB
X035ErrorSub module failure (ECC memory error).KRNL_RAM_RUNTIME_FAILED; Bad SDRAM ECC errors were detected.•       Controller PCB
X037ErrorSub module failure (Ethernet receiver).KRNL_FCC_INTERNAL_RX_ERROR; CRC Error on Ethernet communication.Ethernet Cable; Controller PCB
X038ErrorSub module failure (Ethernet transmitter).KRNL_FCC_INTERNAL_TX_ERROR; CRC Error on Ethernet communication.Ethernet Cable; Controller PCB
X039ErrorSub module failure (+12V).KRNL_12V_VOLTAGE_IS_BAD; Kernel 12 Voltage is bad error (Pneumatics and Fluidics modules only).•       Controller PCB
X050ErrorSub module failure (software).CMN_SW_ERR - A sub module software error was detected. Similar to 1014 for the Host system. NOTE: Table Top Only will show 2050 not 5050 because Table Top communicates through Supervisor)Application Software - Reload; Controller PCB
X051ErrorSub module failure (invalid parameter).CMN_PARAM_ERR - An invalid parameter value was passed to a function.•       Controller PCB
X053ErrorSub module failure (ADC timeout).CMN_ADC_TIMEOUT - Timeout waiting for the ADC done bit to get set.•       Controller PCB
X054ErrorSub module failure (DAC timeout).CMN_DAC_TIMEOUT - Timeout waiting for the DAC busy bit to be cleared.•       Controller PCB
X055ErrorSub module failure (critical data invalid).CMN_CRITICAL_DATA_ERR - Critical data parameter error. The data and the inverted value do not match.•       Controller PCB
X056ErrorSub module failure (watchdog non- operational).CMN_WD_ATTACH_ERR - Received an attach signal indicating that the kernel watchdog process died.•       Controller PCB
X057ErrorSub module failure (supervisor timeout)CMN_SUPERVISOR_TIMEOUT_ERR - Supervisor set point message timeout. The supervisor did not send set point messages at the correct rate when a surgical function was active.Ethernet cables; Supervisor; Controller PCB; Footswitch Cable
X059ErrorSub module failure (access violation).CMN_ACCESS_ERR - A process tried to access a resource without sufficient privileges.•       Controller PCB
X061AdvisorySub module failure (unknown message)CMN_UNKNOWN_SIG_ERR - A signal not recognized by a process was received. The extra parameter contains a pointer to the unknown signal.•       Controller PCB
X062ErrorSubmodule failure (triggered to demonstrate recovery functionality).Error intentionally generated by the Host to test submodule error recovery functionality. Used for user training purposes.•       N/A
X099ErrorSub module failure (no communication with Host)CMN_SUBSYSTEM_UNAVAILABLE_ERR - The Host did not receive a Sub module information message within approximately 3 1/2 minutes after the splash screen was removed after power up or restart. NOTE: Laser and Auxiliary Illuminators are exempt from this error due to the fact that their power state/ejection state can change at anytime and be reset.•       Controller PCB
1001FaultFault - 1001; Call Field ServiceOseGatewayDownFlt - The OSE Gateway exited unexpectedly.•   Isolated occurrence - Cycle power; Host; Supervisor
1002FaultFault - 1002; Call Field ServiceSupervisorProcessDownFlt - The OSE Gateway has reported that the Supervisor process has gone down.•   Isolated occurrence - Cycle power; Supervisor
1003FaultFault - 1003; Call Field ServiceSupervisorCommErrorFlt - Something has failed in the communication between the Host and the Supervisor.•   Isolated occurrence - Cycle power; Host; Supervisor
1004FaultFault - 1004; Call Field ServiceMainApplicationDownFlt - The Main Application exited unexpectedly.•   Isolated occurrence - Cycle power; Application Software - Install; SDRAM - Reseat or Replace; Host; Supervisor
1005FaultFault - 1005; Call Field ServiceCrcChecksumFailureFlt – The Host Controller checksum validation failed to pass all files under its control.•   Isolated occurrence - Cycle power; Host; Supervisor
1006FaultFault - 1006; Call Field ServiceMainApplicationFailed ToConnectToSupervisorFlt The Host has failed to connect to the Supervisor process.Isolated occurrence - Cycle power; Signal/Ethernet Cable W17; Isolated occurrence - Cycle power; Host; Supervisor; 24V Supply to Supervisor
1007FaultFault - 1007; Call Field ServiceHostControllerSoftwareErrorFlt - The Host Controller encountered an unexpected error.•   Isolated occurrence - Cycle power; Host; Supervisor
1008FaultFault - 1008; Call Field ServiceHostControllerExeLaunchFlt - The Host Controller failed to launch an executable image during startup.•   Isolated occurrence - Cycle power; Reset Power Controller
1009FaultFault - 1009; Call Field ServicePowerModuleCommFailureFlt - The Host Controller could not communicate with the Power Module over the serial link•   Isolated occurrence - Cycle power; Host; Power Controller PCB
1010FaultFault - 1010; Call Field ServiceVersionCheckFailureFlt- Component or Submodule reported an incorrect software, hardware, firmware or other versionIsolated occurrence - Cycle power; Verify correct Software OS and Application; are installed; Reinstall OS or Application Software.
1011FaultFault - 1011; Call Field ServiceIpcProxyMissingFlt – Host Controller communications was lost as the result of a .NET IPC failure•   Isolated occurrence - Cycle power; Host; Supervisor
1012FaultFault - 1012; Call Field ServiceMissingSupervisorHeartbeatFlt The Host has not received any heartbeats from the Supervisor within the required timeframe.Supervisor; Ethernet cable W17; Host
1013FaultFault - 1013; Call Field ServiceUiThreadUnresponsiveFlt - The Host has detected that the User Interface thread has not responded within the required timeframe (i.e. the UI is considered locked up)•   Isolated occurrence - Cycle power; Host; Supervisor
1014FaultFault - 1014; Call Field ServiceSoftwareErrorFlt The system has detected a software error.•   Isolated occurrence - Cycle power; Host; Supervisor
1015FaultFault - 1015; Call Field ServiceUnknownSupervisorMsgFlt The Host has received a message from the Supervisor that it doesn’t recognize (this is a specific type of software error)•   Isolated occurrence - Cycle power; Host; Supervisor
1016FaultFault - 1016; Call Field ServiceApplicationFontsFailureFlt The Host application’s display fonts can not be loaded.•   Application Software - Reinstall
1017FaultFault - 1017; Call Field ServiceAudioPlaybackFlt A failure has occurred during the playback of an audio file.•   Isolated occurrence - Cycle power; Host
1018FaultFault - 1018; Call Field ServiceThe Main Application failed to communicate with the Host Controller.•   Isolated occurrence - Cycle power; Host
1019FaultFault - 1019; Call Field ServiceThe Host Controller failed to communicate with the Main Application.•   Isolated occurrence - Cycle power; Host
1020FaultFault - 1020; Call Field ServiceThe Main Application detected an invalid configuration for the host or submodules.•   Isolated occurrence - Cycle power; Host
1021FaultFault - 1021; Call Field ServiceFault intentionally generated by the Host to test the fault recovery functionality. Used for user training purposes.
1022FaultFault - 1022; Call Field ServiceFault intentionally generated by the Host to demonstrate the system behavior when an unrecoverable fault is triggered. Used for user training purposes.
1107ErrorUnable to load one or more language packs.InvalidLanguageLoadAttemptErr An error occurred while attempting to load a non-English language.Verify correct version of Language; Translation Tool is being used.
1108ErrorIncompatible version numbers within the Fluidics submodule. Fluidics functions will be disabled.IncompatibleVersionFluidicsErr - The Fluidics module was shut down due to an incompatible version. The log file shows exactly what version information was incompatible.Software corruption or incompatibility; Fluidics Controller PCB
1109ErrorIncompatible version numbers within the Pneumatics submodule. Pneumatics functions will be disabled.IncompatibleVersionPneumaticsErr The Pneumatics module was shut down due to an incompatible version. The log file shows exactly what version information was incompatible.Software corruption or incompatibility; Pneumatics Controller PCB
1110ErrorIncompatible version numbers within the Ultrasound submodule. Ultrasound and Diathermy functions will be disabled.IncompatibleVersionUltrasoundErr The Ultrasound module was shut down due to an incompatible version. The log file shows exactly what version information was incompatible.Software corruption or incompatibility; U/S Controller PCB
1111ErrorIncompatible version numbers within the Auxiliary Illuminator submodule. Auxiliary Illuminator functions will be disabled.IncompatibleVersionAuxIllumErr The Auxiliary Illuminator module was shut down due to an incompatible version. The log file shows exactly what version information was incompatible.Software corruption or incompatibility; Laser Controller PCB
1112ErrorIncompatible version numbers within the Laser submodule. Laser functions will be disabled.IncompatibleVersionLaserErr The Laser module was shut down due to an incompatible version. The log file shows exactly what version information was incompatible.Software corruption or incompatibility; Aux Controller PCB
1200AdvisoryFunction is not allowed when the fluidics subsystem is not functional.CommandNotAllowedFluidicsNotFunctionalAdv The user tried to select Fluidics function which is only allowed when the fluidics subsystem is functional.•   User Advisory - See Discrepancy description
1202AdvisoryFunction is not allowed when the Laser is in Ready Mode or firing.CommandNotAllowedWhenLaserReadyOrFiringAdv The user tried to invoke a command which is not allowed when the laser is in Ready Mode or Firing.•   User Advisory - See Discrepancy description
1203AdvisoryFunction is not allowed when the footswitch treadle is down or buttons are pressed.CommandNotAllowedWhenTreadleDownAdv The user tried to invoke a command which is only allowed when the treadle is up and no buttons pressed.•   User Advisory - See Discrepancy description
1204AdvisoryFlow mode is not available until the probe or handpiece has been primed.ExtrFlowModeUnavailProbeHpNotPrimedAdv Flow mode is unavailable because a probe or handpiece hasn’t been primed.•   User Advisory - See Discrepancy description
1205AdvisoryFlow mode is only available for 20 gauge probes / handpieces.ExtrFlowModeUnavailProbeHpNot20GaugeAdv Flow mode is unavailable because a non 20 gauge probe or handpiece is used.•   User Advisory - See Discrepancy description
1207AdvisoryPlease connect a Phaco handpiece.NoPhacoHpConnectedAdv The user presses the treadle in a Phaco Step when no Phaco handpiece is connected.•   User Advisory - See Discrepancy description
1208AdvisoryPlease tune the Phaco handpiece.PhacoHpNotTunedAdv The user presses the treadle in a Phaco Step when the Phaco handpiece isn’t tuned.•   User Advisory - See Discrepancy description
1209AdvisoryPlease connect a Frag handpiece.NoFragHpConnectedAdv The user presses the treadle in a Frag Step when no Frag handpiece is connected.•   User Advisory - See Discrepancy description
1210AdvisoryPlease tune the Frag handpiece.FragHpNotTunedAdv The user presses the treadle in a Phaco Step when the Phaco handpiece isn’t tuned.•   User Advisory - See Discrepancy description
1211AdvisoryFunction is not allowed when the Ultrasound submodule is not functional.CommandNotAllowedUltrasoundNotFunctionalAdv The user tried to invoke a command which is only allowed when the ultrasound subsystem is functional.•   User Advisory - See Discrepancy description
1212AdvisoryFunction is not allowed when the cassette is not ready.CommandNotAllowedCassetteNotReadyAdvThe user tried to invoke a command which is only allowed when the cassette is ready.•   User Advisory - See Discrepancy description
1213AdvisoryInfusion must be on to use VFC Extract. Press [Ignore] to allow VFC Extract without infusion for this case.InfusionNeededInInVfcExtractAdv The user presses the treadle in a VFC Extract Step when infusion is off.•   User Advisory - See Discrepancy description
1214AdvisorySystem hasn’t detected VFC. If already connected, press “Connected” button.NoVfcSyringeConnectedAdv The user presses the treadle in a VFC Step when no VFC Syringe is connected.•   User Advisory - See Discrepancy description
1215AdvisorySystem hasn’t detected Forceps. If already connected, press “Connected” button.NoForcepsConnectedAdv The user presses the treadle in a Forceps Step when no forceps are connected.•   User Advisory - See Discrepancy description
1216AdvisorySystem hasn’t detected Scissors. If already connected, press “Connected” button.NoScissorsConnectedAdv The user presses the treadle in a Scissors Step when no Scissors are connected.•   User Advisory - See Discrepancy description
1217AdvisoryPlease connect an AGF syringe.NoAgfSyrinceConnectedAdv - The user presses the “Start” button in the Auto Gas Filling dialog when no AGF syringe has been connected.•   User Advisory - See Discrepancy description
1218AdvisoryForceps not available during AGF: please try again when AGF is complete.ForcepsNotAvailableDuringAgfAdv The user presses the treadle in a Forceps Step while Auto Gas Filling is in progress.•   User Advisory - See Discrepancy description
1219AdvisoryVFC not available during AGF: please try again when AGF is complete.VfcNotAvailableDuringAgfAdv The user presses the treadle in a VFC Step while Auto Gas Filling is in progress.•   User Advisory - See Discrepancy description
1220AdvisoryEjecting the cassette is not allowed while the footswitch treadle is pressed.CassetteEjectRejectedTreadleDownAdv The user attempts to eject the cassette while the treadle is down.•   User Advisory - See Discrepancy description
1221AdvisoryEjecting the cassette is not allowed while infusion or FAX is on.CassetteEjectRejectedInfOnAdv The user attempts to eject the cassette while infusion is on.•   User Advisory - See Discrepancy description
1222AdvisoryEjecting the cassette is not allowed while irrigation is on.CassetteEjectRejectedIrrOnAdv The user attempts to eject the cassette while irrigation is on.•   User Advisory - See Discrepancy description
1223AdvisoryEjecting the cassette is not allowed while priming, tuning, or testing.CassetteEjectRejectedTestInProgrAdv The user attempts to eject the cassette while tests (priming/ tuning/testing) are in progress.•   User Advisory - See Discrepancy description
1224AdvisoryCleaning the cassette is not allowed while infusion, irrigation, or FAX is on.CassetteCleanNotAllowedWhenInfOrIrrIsOnAdv The user attempts to start Cassette Cleaning when either Infusion, Irrigation or FAX is on.•   User Advisory - See Discrepancy description
1225AdvisoryCleaning the cassette is not allowed without a functional cassette.CassetteCleanNotAllowedCassetteNotAvailableAdv The user attempts to start Cassette Cleaning when the Cassette isn’t available (e.g. not inserted).•   User Advisory - See Discrepancy description
1226AdvisoryPriming the cassette is not allowed without a functional cassette.User attempt to prime the cassette when preconditions aren’t met (e.g. no cassette inserted, cassette not tested or didn’t pass the test)•   User Advisory - See Discrepancy description
1227AdvisoryAn error occurred loading the device settings. The system will revert to default values.LoadDeviceSettingsFailureAdv A database attempt to load the device settings from the file system has failed. Reverting to defaults.•   User Advisory - See Discrepancy description
1228AdvisoryCommand is not allowed while infusion is on.CommandNotAllowedInfusionIsOnAdv The user tried to invoke a command which is only allowed when infusion is not on.•   User Advisory - See Discrepancy description
1229AdvisoryCommand is not allowed while irrigation is on.CommandNotAllowedIrrigationIsOnAdv The user tried to invoke a command which is only allowed when irrigation is not on.•   User Advisory - See Discrepancy description
1230AdvisoryThe selected step is not supported with the current cassette type.IncompatibleOperatingModeAdv The user attempts to select a Step whose mode/submode isn’t compatible with the current operating mode.•   User Advisory - See Discrepancy description
1231AdvisoryThe connected probe is not supported with the current cassette type.VitProbeNotCompatibleAdv The user connects a probe (with RFID) that’s not compatible with the current operating mode (e.g. an Ultra Vit probe when an Anterior cassette is connected or Ultra Vit Anterior probe when a Posterior cassette is connected)•   User Advisory - See Discrepancy description
1232AdvisoryPlease connect footswitch.NoFootswitchAdv The Supervisor has reported that no footswitch is connected.•   User Advisory - See Discrepancy description
1233AdvisoryThe connected footswitch is not supported.UnknownFootswitchAdv The Supervisor has detected the connection of an unknown footswitch.•   User Advisory - See Discrepancy description
1234AdvisoryFlow limit is currently not available due to flow mode being unavailable.ExtrVacModeFlowLimitUnavailableAdv Vacuum mode flow limit depends on the availability of flow mode. If flow mode becomes unavailable then the flow limit function can not be performed by the fluidics submodule.•   User Advisory - See Discrepancy description
1236AdvisoryGravity pressure mode is currently not available.GravityBasedPressureUnavailableAdv User attempts to select Gravity as the infusion source when a Premium Cassette is inserted.•   User Advisory - See Discrepancy description
1237AdvisoryIOP Control is currently not available.IopCompensationUnvailableAdv User attempts to turn IOP Control on when it’s not available.•   User Advisory - See Discrepancy description
1238AdvisoryAn error occurred saving the device settings.SaveDeviceSettingsFailureAdv User attempt to save device settings fails.•   User Advisory - See Discrepancy description
1239AdvisoryProportional reflux mode is currently not available.PropRefluxModeUnavailFluidLvlNotInRangeAdv Displayed in one of the following situations: 1.User attempts to toggle into proportional reflux mode when the fluid level in the cassette chamber is out of range or the fluid 2.Fluid level gets out of range while in proportional reflux. 3.Treadle is depressed while in proportional reflux and the fluid level is out of range.•   User Advisory - See Discrepancy description
1240AdvisoryExtraction flow mode is currently not available.ExtrFlowModeUnavailFluidLvlNotInRangeAdv The fluid level in the cassette chamber is out of range and the treadle is depressed.•   User Advisory - See Discrepancy description
1241AdvisoryInfusion source is getting low: please check the fluid container.InfContainerNearEmptyAdv The infusion container fluid level is getting low.•   User Advisory - See Discrepancy description
1242AdvisoryCommand is not allowed due to instrument not being available.Command is not allowed due to instrument not being available.•   User Advisory - See Discrepancy description
1243AdvisoryInfusion / Irrigation source is empty: please press [Change] and replace the fluid container.InfContainerOnReserveAdv The infusion fluid container is empty but there’s still fluid in the cassette chambers.•   User Advisory - See Discrepancy description
1244AdvisoryCommand is not allowed while priming, tuning, or testing.CommandNotAllowedWhenTestInProgressAdv The user attempts to do something, e.g. change Steps, that’s not allowed when tests are in progress.•   User Advisory - See Discrepancy description
1245AdvisoryPlease connect handpiece.NoHpConnectedAdv The user attempts to perform an action, e.g. priming, that requires a handpiece but no handpiece is connected. Note: This is only applicable to handpieces the system can detect connection for (e.g. Phaco)•   User Advisory - See Discrepancy description
1247AdvisoryConnected handpiece is not supported with the current cassette type.HpNotCompatibleWithOperatingModeAdv User connects a handpiece that’s not compatible with the current Cassette type. (e.g., a Frag handpiece is connected when an Anterior Cassette is inserted)•   User Advisory - See Discrepancy description
1248AdvisoryIs the inserted cassette new?IsInsertedCassetteNewAdv The user has inserted a Cassette which the system can’t determine whether it’s new or the same Cassette that was previously ejected.•   User Advisory - See Discrepancy description
1249AdvisoryAn error occurred parsing the log file.LogFileParseFailureAdv An error occurred parsing one of more lines of the log file.•   User Advisory - See Discrepancy description
1250AdvisoryMulti-Cut is not available when proportional scissors are selected. Press [Multi-Cut] to indicate that multi-cut scissors are currently connected.ScissorsMultiCutUnavailWhenPropScissSelAdv The user has selected scissors of type Proportional and attempts to select the Multi-Cut submode or attempt to use momentary cutting in Extrusion Mode. The advisory lets the user change both the Scissors type and the Scissors submode.•   User Advisory - See Discrepancy description
1251AdvisoryCommand is not allowed while the cassette is being cleaned.The user tried to turn Infusion, Irrigation or FAX on while the cassette is being cleaned.•   User Advisory - See Discrepancy description
1252AdvisoryThe scanned barcode is not recognized.UnknownBarcodeScannedAdv The user has scanned an item that the system doesn’t recognize.•   User Advisory - See Discrepancy description
1255AdvisoryIlluminator fiber is not connected.NoIlluminatorConnectedAdv User has attempted to turn on an illuminator with no illuminator fiber connected.•   Probe not fully inserted •   Fiber Detection Cable mounting screw too Short •   Fiber Detection Cable assembly •   RFID PCB
1256AdvisoryProportional reflux mode is not available when the footswitch treadle is depressed.PropRefluxUnavailableTreadleDownAdv User attempts to enter Proportional Reflux mode when the treadle is down•   User Advisory - See Discrepancy description
1257AdvisoryThe report’s header or footer has too many rows to fit on the page.ReportExcessiveHeaderFooterSizeAdv The user has added too many rows to a header or footer table being edited in an End Case report. As a result, the table will not be printed out in its entirety on the various pages comprising the report.•   User Advisory - See Discrepancy description
1258AdvisoryThe report’s current table has too many rows to fit on the page.ReportExcessiveTableSizeAdv The user has added too many rows to the current table being edited in an End Case report. As a result, the table will not be printed out in its entirety on the various pages comprising the report.•   User Advisory - See Discrepancy description
1259AdvisoryCommand is not allowed while proportional diathermy is active.CommandNotAllowedInProportionalDiathermyAdv The user attempts to enter proportional reflux while proportional diathermy is active.•   User Advisory - See Discrepancy description
1260AdvisoryCommand is not allowed while proportional reflux is active.CommandNotAllowedInProportionalRefluxAdv The user attempts to enter proportional diathermy while proportional reflux is active.•   User Advisory - See Discrepancy description
1261AdvisoryCassette can’t be ejected while being cleaned.CassetteEjectRejectedCassetteIsCleaningAdv The user attempts to eject the cassette while it’s being cleaned.•   User Advisory - See Discrepancy description
1263AdvisoryPort can’t be selected: there is no probe connected.LsrActivePortSelRejectedNoProbeConnAdv The user tries to make a laser port the active port but there’s no probe connected to that port.•   User Advisory - See Discrepancy description
1264AdvisoryPort can’t be selected: the probe type for the port isn’t valid.LsrActivePortSelRejectedNoValidProbeSelAdv The user tries to make a laser port the active port but the currently selected probe type for that port is invalid.•   User Advisory - See Discrepancy description
1266AdvisoryA laser probe is not connected to the active port.LsrNoProbeConnAdv Shown in any of the following situations: · Laser Step entered and no probe connected to the active port · Probe removed from active port · User tries to go to Ready Mode when no probe is connected to the active port•   User Advisory - See Discrepancy description
1267AdvisoryA valid laser probe is not selected for the active port.LsrNoValidProbeSelAdv Shown in any of the following situations: · Laser Step entered and no valid probe type selected for the active port · User tries to go to Ready Mode when no valid probe type is selected for the active port•   User Advisory - See Discrepancy description
1268AdvisoryThe laser remote interlock is open.LsrInterlockOpenAdv Shown in any of the following situations: · Laser Step entered and the Interlock is open · Interlock opened in a Laser Step · User tries to go to Ready Mode when the Interlock is open•   User Advisory - See Discrepancy description
1269AdvisoryLaser Dr. Filters are not connected to the console. Are all necessary Dr. Filters properly installed and connected?LsrDoctorFilterUnverified0ConnAdv Shown in any of the following situations: Laser Step entered and the Dr. Filters haven’t been verified for the active port · Endo Probe inserted to the active port while in a Laser Step. User tries to go to Ready Mode when Dr. Filters haven’t been verified for the active port.•   User Advisory - See Discrepancy description
1270AdvisoryOne Laser Dr. Filter is connected to the console. Are all necessary Dr. Filters properly installed and connected?LsrDoctorFilterUnverified1ConnAdv Same as above•   User Advisory - See Discrepancy description
1271AdvisoryTwo Laser Dr. Filters are connected to the console. Are all necessary Dr. Filters properly installed and connected?LsrDoctorFilterUnverified2ConnAdv Same as above•   User Advisory - See Discrepancy description
1273AdvisoryLaser Dr. Filter 1 is disengagedLsrDoctorFilter1DisengagedAdv The user has disengaged the (connected) Dr. Filter 1 while in a Laser Step.•   User Advisory - See Discrepancy description
1275AdvisoryLaser Dr. Filter 2 is disengagedLsrDoctorFilter2DisengagedAdv Same as 1273 but for Dr. Filter 2•   User Advisory - See Discrepancy description
1276AdvisoryCommand is not allowed when Laser is firing.LsrCommandNotAllowedWhenFiringAdv User has attempted to change one of the laser’s settings while the laser is firing.•   User Advisory - See Discrepancy description
1277AdvisoryPort can’t be selected: it’s not functional.LsrActivePortSelRejectedPortFailureAdv User has attempted to select a laser port that’s not functional.•   User Advisory - See Discrepancy description
1278AdvisoryCannot go to Laser Ready Mode while the current screen is being displayed.LsrReadyModeRejectedScreenActiveAdv User has pressed the Ready button from the footswitch when a “conflicting screen” is active.•   User Advisory - See Discrepancy description
1279AdvisoryNo laser footswitch is connected.LsrNoFootswitchConnectedAdv User enters a Laser Step when no laser footswitch is connected or the laser footswitch is disconnected in a Laser Step.•   User Advisory - See Discrepancy description
1280AdvisoryUnable to write to the report file.ReportFileWriteErrorAdv A problem has occurred when trying to write to the specified Report file.•   User Advisory - See Discrepancy description
1281AdvisoryUnable to read from the report file.ReportFileReadErrorAdv A problem has occurred when trying to read from the specified Report file.•   User Advisory - See Discrepancy description
1285AdvisoryThe report file and its associated CRC value do not match.ReportFileCrcMismatchErrorAdv The Report file and its associated CRC value do not match.•   User Advisory - See Discrepancy description
1288AdvisoryCan’t eject the cassette while reflux is active.CassetteEjectRejectedRefluxInUseAdv The user attempts to eject the cassette while reflux (of any type) is active. For proportional reflux, this advisory is displayed when the user attempts cassette ejection while proportional reflux mode is active with the treadle not depressed. Cassette ejection while the treadle is depressed will generate advisory 1220.•   User Advisory - See Discrepancy description
1289AdvisoryExtraction flow mode is not available when FAX is turned on.ExtrFlowModeUnavailFaxInUseAdv FAX was turned on while in Flow Mode or a Posterior Step with Flow Mode preference was selected when FAX was on.•   User Advisory - See Discrepancy description
1290AdvisoryThe surgical function is currently unavailable.SurgicalFunctionProxyUnavailableAdv The user has requested surgical functionality that is currently unavailable. This is a default advisory that is only displayed for cases in which a more meaningful explanation is not available.•   User Advisory - See Discrepancy description
1292AdvisoryAn error occurred accessing the video recorder. Please verify it is on and connected.VideoRecorderNotCommunicatingAdv An error occurred accessing the video recorder.•   Check connection to Video Recorder.
1296AdvisoryThe barcode reader failed to initialize.BarcodeReaderInitFailureAdv The bar code reader failed to initialize during startup.•   User Advisory - See Discrepancy description
1297AdvisoryThe system is currently low on free disk space. Please backup or remove non-critical files.LowFreeSpace During startup, free space was detected at less than the desired minimum %.•   User Advisory - See Discrepancy description
1298AdvisoryA printer must be configured in System Settings before printing.PrinterNotConfiguredAdv The user tried to print but has not configured the printer in system settings•   User Advisory - See Discrepancy description
1299AdvisoryThe current printer configuration could not be saved.DefaultPrintSettingsErrorAdv An error occurred attempting to save the default printer with the current configuration•   User Advisory - See Discrepancy description
1302AdvisoryAn unknown error occurred while printing.PrinterGeneralErrorAdv Generated by any printer error other the errors above.•   User Advisory - See Discrepancy description
1303AdvisoryThe Illuminators are currently turned off from the footswitch and cannot be turned on or off at this time.IlluminatorMasterSwitchOffAdv The user has attempted to turn on or off an endo illuminator when the footswitch Momentary Endo Illuminators Off function is active.•   User Advisory - See Discrepancy description
1304AdvisoryThe Laser is currently unable to deliver the maximum power level of 2 Watts. The maximum Laser power (Watts) currently available is:LsrMaxPowerAvailableAdv Is generated when the power available from the laser drops below the maximum level (2 Watts).•   User Advisory - See Discrepancy description
1305AdvisoryAn error occurred loading a surgeon. The system will revert to default values.LoadSurgeonFailureAdv A database attempt to load a surgeon from the file system has failed. Reverting to defaults.•   User Advisory - See Discrepancy description
1306AdvisoryAn error occurred setting the printer port. The system will revert to the previous value.PrinterPortConfigErrorAdv Attempt to set the printer IP address failed.•   User Advisory - See Discrepancy description
1307AdvisoryAn error occurred loading a procedure. The system will revert to default values.LoadProcedureFailureAdv A database attempt to load a procedure from the file system has failed. Reverting to defaults.•   User Advisory - See Discrepancy description
1308AdvisoryAn error occurred loading the system settings. The system will revert to default values.LoadSystemSettingsFailureAdv A database attempt to load the system settings from the file system has failed. Reverting to defaults.•   User Advisory - See Discrepancy description
1309AdvisoryAn error occurred saving a surgeon’s settings. Changes will be reverted at the start of the next case.SaveSurgeonFailureAdv A database attempt to save a surgeon to the file system has failed.•   User Advisory - See Discrepancy description
1310AdvisoryAn error occurred saving a procedure. Changes will be reverted at the start of the next case.SaveProcedureFailure A database attempt to save a procedure from the file system has failed. Reverting to defaults.•   User Advisory - See Discrepancy description
1311AdvisoryAn error occurred saving the system settings. The system will use default values.SaveSystemSettingsFailureAdv A database attempt to save the system settings from the file system has failed. Reverting to defaults.•   User Advisory - See Discrepancy description
1312AdvisoryAn error has occurred initializing Video Overlay. Video Overlay functions will be disabled.VideoOverlayInitializationErrorAdv An error happened during VideoOverlay initialization.•   User Advisory - See Discrepancy description
1313AdvisoryAn error has occurred in the Video Overlay component. Video Overlay functions will be disabled.VideoOverlayErrorAdv An error happened during VideoOverlay processing.•   User Advisory - See Discrepancy description
1314AdvisoryThere was an error saving surgeon data.VcdSaveSurgeonAdv There was an error saving surgeon data.•   User Advisory - See Discrepancy description
1317AdvisoryThere was an error renaming the surgeon.VcdRenameSurgeonAdv There was an error renaming the surgeon.•   User Advisory - See Discrepancy description
1318AdvisoryThere was an error renaming the procedure.VcdRenameProcedureAdv There was an error renaming the procedure.•   User Advisory - See Discrepancy description
1319AdvisoryThere was an error renaming the case data.VcdRenameCaseAdv There was an error renaming the case data.•   User Advisory - See Discrepancy description
1320AdvisoryThere was an error deleting the surgeon.VcdDeleteSurgeonAdv There was an error deleting the surgeon.•   User Advisory - See Discrepancy description
1321AdvisoryThere was an error deleting the procedureVcdDeleteProcedureAdv There was an error deleting the procedure•   User Advisory - See Discrepancy description
1322AdvisoryThere was an error deleting the case data.VcdDeleteCaseAdv There was an error deleting the case data•   User Advisory - See Discrepancy description
1323AdvisoryThere was an error loading the surgeon.VcdLoadSurgeonAdv There was an error loading the surgeon.•   User Advisory - See Discrepancy description
1325AdvisoryThere was an error loading the case data.VcdLoadCaseAdv There was an error loading the case data.•   User Advisory - See Discrepancy description
1326AdvisoryUnable to load the case report template.VcdLoadCaseReportTemplateAdv There was an error loading the Case Report Template.•   User Advisory - See Discrepancy description
1329AdvisoryUnable to save the technician’s log.VcdSaveTechniciansLogAdv There was an error saving the Technicians Log.•   User Advisory - See Discrepancy description
1330AdvisoryUnable to load the video table.VideoInfoTableReadFailureAdv There was an error loading the Video Table.•   User Advisory - See Discrepancy description
1331AdvisoryUnable to play the video.VideoDfuFailureAdv There was an error loading and\or playing the Video file.•   User Advisory - See Discrepancy description
1332AdvisoryUnable to load Help.HelpFileReadFailureAdv There was an error opening the Help pdf file.•   User Advisory - See Discrepancy description
1333AdvisoryUnable to save the case info.SaveCaseInfoFailureAdv There was an error saving the Case Info file.•   User Advisory - See Discrepancy description
1334AdvisoryUnable to update the log file.LogFileWriteFailureAdv There was an error writing to the Log File.•   User Advisory - See Discrepancy description
1335AdvisoryUnable to read the log file.LogFileReadFailureAdv There was an error reading from the Log File.•   User Advisory - See Discrepancy description
1336AdvisoryUnable to write the incident file.IncidentFileWriteFailureAdv There was an error writing to the Incident File.•   User Advisory - See Discrepancy description
1337AdvisoryUnable to read the incident file.IncidentFileReadFailureAdv There was an error reading from the Incident File.•   User Advisory - See Discrepancy description
1338AdvisoryUnable to write the system metrics file.SystemMetricFileWriteFailureAdv There was an error writing to the System Metrics File.•   User Advisory - See Discrepancy description
1339AdvisoryUnable to read the system metrics file.SystemMetricFileReadFailureAdv There was an error reading from the System Metrics File.•   User Advisory - See Discrepancy description
1340AdvisoryUnable to write to the removable drive.ErrorAccessingRemovableDriveAdv There was an error while attempting to write to the removable drive.•   User Advisory - See Discrepancy description
1341AdvisoryFurther increasing the output level in air can damage fiber tips. Would you like to continue?IllumSafeThreshold1ExceededAdv The user has exceeded Threshold 1 illuminator value by either turning on the illuminator, changing the setpoint while the illuminator is on, or changing Procedures while the illuminator is on.•   User Advisory - See Discrepancy description
1342AdvisoryFurther increasing the output level will reduce exposure time by 35%. Would you like to continue?IllumSafeThreshold2ExceededAdv The user has exceeded Threshold 2 illuminator value by either turning on the illuminator, changing the setpoint, or changing Procedures.•   User Advisory - See Discrepancy description
1343AdvisoryFurther increasing the output level in air can damage fiber tips. Also, further increasing the output level will reduce exposure time by 35%. Would you like to continue?IllumSafeThreshold1And2ExceededAdv The user has exceeded both Threshold 1 and Threshold 2 illuminator value by either turning on the illuminator, changing the setpoint while the illuminator is on, or changing Procedures while the illuminator is on.•   User Advisory - See Discrepancy description
1344AdvisoryOnly two illuminators can be turned on simultaneously.IllumMaxTwoIlluminatorsTurnedOnAdv User attempts to turn on an illuminator port when two ports are already on. This is not allowed.•   User Advisory - See Discrepancy description
1345AdvisoryAGF not allowed while Forceps or VFC is in use.AgfNotAllowedForcepsOrVfcInUseAdv User attempts to start Auto Gas Filling when either Forceps or VFC is in use (i.e. selected and treadle down).•   User Advisory - See Discrepancy description
1346AdvisoryAn error occurred trying to access the wireless network.WirelessNetworkAccessAdv An error occurs when the user attempts to access the wireless network.•   User Advisory - See Discrepancy description
1349AdvisoryA RAID hard drive has failed or is missingRaidDriveFailedOrMissingAdv One Disk in a redundant RAID volume is missing or has failed but the volume is still functional.•   User Advisory - See Discrepancy description
1350AdvisoryThe probe in the current laser port is not supported.LsrReadyModeRejectedInvalidProbeTypeAdv The type of probe in the current port is not supported by CONSTELLATION* VISION SYSTEM EH1 but might be supported by PurePoint* Laser (slit lamp for example).•   User Advisory - See Discrepancy description
1355AdvisoryCommand is not allowed when the Pneumatics submodule is not functional.CommandNotAllowedPneumaticsNotFunctionalAdv The user tried to invoke a command which is only allowed when the Pneumatics subsystem is functional.•   User Advisory - See Discrepancy description
1356AdvisoryCommand is not allowed when an Anterior Only Cassette is being used.CommandNotAllowedWithAnteriorOnlyCassetteAdv The user tried to invoke a command which is not allowed with an Anterior Only Cassette.•   User Advisory - See Discrepancy description
1357AdvisoryCannot switch to Infusion: the cassette has not been primed.CantSwitchToInfusionCassetteNotPrimedAdv The user tried to switch to Infusion when the cassette was not primed.•   User Advisory - See Discrepancy description
1358AdvisoryCurrent probe type is unrecognized. Please select a valid probe type.IllumTurnOnRejectedUnrecognizedTypeAdv User has attempted to turn on an illuminator when the probe type is “unrecognized”•   User Advisory - See Discrepancy description
1359AdvisoryPower module communication error. When powering down the system, you will have to press the “Options/Shutdown” button on the screen.PowerModuleCommFailureAdv Host controller failed to communicate with power module.•   User Advisory - See Discrepancy description
1361AdvisoryCommand is not allowed when Infusion backup pressure is active.The user tried to invoke a command which is only allowed when Infusion backup pressure is not active.•   User Advisory - See Discrepancy description
1362AdvisoryCommand is not allowed when the probe type has been identified by RFID.The user tried to change the type of a laser probe that has been identified by RFID.•   User Advisory - See Discrepancy description
1363AdvisoryCommand is not allowed when the probe type has been previously identified by the user. Disconnect and then reconnect the probe to change its type.The user tried to change the type of a laser probe that has been identified by a user selection. Only one selection is allowed after connection.•   User Advisory - See Discrepancy description
1364AdvisoryIOP Control is not allowed without a premium cassette inserted. Please insert a premium cassette and retry.IopNotAllowedWhenUsingNonPremiumCassetteAdv The user tried to turn on IOP Control when a premium cassette is not inserted.•   User Advisory - See Discrepancy description
1365AdvisoryIOP Control is not allowed with an uncalibrated cassette. Please calibrate cassette and retry.IopNotAllowedWhenCassetteNotCalibratedAdv The user tried to turn on IOP Control with a cassette that has not been calibrated.•   User Advisory - See Discrepancy description
1366AdvisoryCommand is not allowed when FAX is on and the cassette has not been primed.The user tried to perform a Test Instrument command when FAX is on and the cassette is not primed.•   User Advisory - See Discrepancy description
1367AdvisoryFAX is not allowed when the cassette is being primed.FaxNotAllowedWhenCassetteIsPrimingAdv The user tried to turn on FAX when the cassette is being primed.•   User Advisory - See Discrepancy description
1369AdvisorySetting values is not allowed when the cassette is in the chamber overflow condition.CommandNotAllowedDuringChamberOverflowAdv The cassette chamber is in an overflow condition and pressure cannot be controlled.•   User Advisory - See Discrepancy description
1370AdvisoryTest instrument is not allowed in this mode.TestInstrumentNotAllowedInThisModeAdv User has attempted to perform a test instrument command in Setup or End Case.•   User Advisory - See Discrepancy description
1371AdvisoryUpdate is only available from End Case.UpdateOnlyAvailableInEndCaseAdv User has attempted to perform a update command in Setup or Surgery.•   User Advisory - See Discrepancy description
1372AdvisoryCleaning the cassette is not allowed when the inlet pressure is out of range. Please adjust the inlet pressure between 58 psi and 120 psi.CassetteCleanNotAllowedAirPressureOutOfRange The user attempts to start Cassette Cleaning when the Air source (wall pressure) is either below 58 psi or above 120 psi.•   User Advisory - See Discrepancy description
1751InformationThe remote control battery is low.RemoteControlBatteryLow User has pressed on a remote control button and the remote’s battery is running out of power•   User Advisory - See Discrepancy description
1752InformationInfusion pressure will drop to zero during calibration. Continue?CalibrationBringsInfusionToZeroInfo User has pressed the “Calibrate” button on the advisory popup brought up when there’s been an IV Pole error. Since I/V Pole calibration will change the current Infusion pressure, the user is warned before calibration is started.•   User Advisory - See Discrepancy description
1753InformationPower recovery in progress. Please wait until surgical functionality becomes available.PowerRecoveryInProgressInfo AC power has been restored and the recovery process has started.•   User Advisory - See Discrepancy description
1373AdvisoryThe Default Doctor can not be modified.CommandNotAllowedDefaultDoctorAdv - The user attempts to save, modify or add a procedure when the default doctor is selected.•   User Advisory - See Discrepancy description
1374AdvisoryThe Accurus* Classic procedure can not be modified.CommandNotAllowedAccurusClassicAdv - The user attempts to modify an Accurus Classic procedure.•   User Advisory - See Discrepancy description
1375AdvisoryThis command is not available in End Case.CommandNotAllowedInEndCaseAdv - The user attempts to change either the doctor or procedure in End Case.•   User Advisory - See Discrepancy description
1376AdvisoryThis command is not available when the laser is firing.CommandNotAllowedLaserFiringAdv The user attempt to reset the shot count and energy metrics when the Laser is firing.•   User Advisory - See Discrepancy description
1377AdvisoryNo changes have been made to the current procedure.CommandNotAllowedNoProcChanges The user attempts to save a Procedure when no procedure changes have been made.•   User Advisory - See Discrepancy description
1378AdvisoryThis command is not available when system is in the surgery screen.CommandNotAllowedInSurgeryMode The command is not available when system is in the Surgery screen.•   User Advisory - See Discrepancy description
1379AdvisoryConnected vitreous probe is not valid. The probe will be used as a low-speed probe.VitProbeInvalidAdv A connected vitreous probe has valid checksum but one or more invalid parameters.•   User Advisory - See Discrepancy description
1380AdvisoryThe CPU battery is bad. Please contact Field Service.CPUBatteryBadAdv•   User Advisory - See Discrepancy description
1381AdvisoryUnable to copy to the removable drive.RemovableDriveCopyErrorAdv•   User Advisory - See Discrepancy description
1382AdvisoryUnable to delete from the removable drive.RemovableDriveDeleteErrorAdv•   User Advisory - See Discrepancy description
1383AdvisoryCommand is not allowed when the inlet pressure is out of range. Please adjust the inlet pressure between 58 psi and 120 psi.CommandNotAllowedAirPressureOutOfRange
The user tried to invoke a command when the Air source (wall pressure) is either below 58 psi or above 120 psi.
•   User Advisory - See Discrepancy description
1384AdvisoryEnsure proper scissors tip attachment to pneumatic handpiece. Prior to use in the eye, depress foot pedal\button to ensure proper tip function.TestScissorsTipAttachmentAdv
User has chosen to use scissors and is reminded to confirm attachment and functionality before use.
•   User Advisory - See Discrepancy description
1385AdvisoryEnsure proper forceps tip attachment to pneumatic handpiece. Prior to use in the eye, depress foot pedal to ensure proper tip function.TestForcepsTipAttachmentAdv
User has chosen to use forceps and is reminded to confirm attachment and functionality before use.
•   User Advisory - See Discrepancy description
1386AdvisoryFluidics submodule recovery unsuccessful. Fluidics functions will remain disabled.Fluidics submodule recovery unsuccessful. Fluidics functions will remain disabled.•   User Advisory - See Discrepancy description
1387AdvisoryUS Diathermy submodule recovery unsuccessful. US Diathermy functions will remain disabled.US Diathermy submodule recovery unsuccessful. US Diathermy functions will remain disabled.•   User Advisory - See Discrepancy description
1388AdvisoryTableTop Illuminator submodule recovery unsuccessful.
TableTop Illuminator functions will remain disabled.
Displayed after the Table Top Illuminator submodule has unsuccessfully recovered from an error.•   User Advisory - See Discrepancy description
1389AdvisoryPneumatics submodule recovery unsuccessful. Pneumatics functions will remain disabled.Displayed after the Pneumatics submodule has unsuccessfully recovered from an error.•   User Advisory - See Discrepancy description
1390AdvisoryAuxiliary Illuminator submodule recovery unsuccessful.
Auxiliary Illuminator functions will remain disabled.
Displayed after the Auxiliary Illuminator submodule has unsuccessfully recovered from an error.•   User Advisory - See Discrepancy description
1391AdvisoryLaser submodule recovery unsuccessful. Laser functions will remain disabled.LaserSubmoduleErrorRecoveryUnsuccessfulAdv Displayed after the Laser submodule has unsuccessfully recovered from an error.•   User Advisory - See Discrepancy description
1392AdvisoryCommand not allowed without a calibrated AutoSert* handpiece.IolCmdNotAllowedHandpieceNotReadyAdv
The user pressed the treadle in an AutoSert* Step when no AutoSert* handpiece was connected or the handpiece was connected but not calibrated.
•   User Advisory - See Discrepancy description
1394AdvisoryThe AutoSert* handpiece is already prepared. The command is not allowed.IolCmdNotAllowedHandpiecePreparedAdv User pressed the “Load Plunger” or “Preload IOL” button when the AutoSert* handpiece was “Prepared”•   User Advisory - See Discrepancy description
1395AdvisoryThe current AutoSert* operation was cancelled.IolCmdCancelledEnteringEndCaseAdv
User selected End Case state when one of the following AutoSert* setup commands were in progress:
• Load Plunger
• Preload IOL
•   User Advisory - See Discrepancy description
1396AdvisoryPreload IOL command not allowed until the AutoSert* plunger has been fully retracted.IolCmdNotAllowedLoadPlungerStartedAdv
User pressed the [Preload IOL] button when a Load Plunger command was previously started but not yet completed (and currently paused).
•   User Advisory - See Discrepancy description
1397AdvisoryLoad Plunger command not allowed until the AutoSert* plunger has been fully retracted.IolCmdNotAllowedPreloadIolStartedAdv
User pressed the [Load Plunger] button when a Preload IOL command was previously started but not yet completed (and currently paused).
•   User Advisory - See Discrepancy description
1399AdvisoryFunction is not allowed when the IOL Injection submodule is not functional.CommandNotAllowedIOLInjectionNotFunctionalAdv The user tried to invoke a command which is only allowed when the IOL Injection subsystem is functional.•   User Advisory - See Discrepancy description
1401Advisory20 cc of undiluted gas has been dispensed. After removing the syringe from the system, adjust the plunger to obtain the desired gas concentration before injection.AgfUndilutedGasAdv
Warns user that there are 20 cc of undiluted gas in the syringe.
•   User Advisory - See Discrepancy description
1402AdvisoryFluidics recovery complete. Cassette status not fully restored.FluidicsSubmoduleRecoveryPartialSuccessAdv The Fluidics submodule recovered, but the system could not fully restore the state due to missing instruments (no valid cassette reported)•   User Advisory - See Discrepancy description
1403AdvisoryTable Top Illuminator recovery complete. Illuminator probe status not fully restored.The Table Top Illuminator load module recovered, but the system could not fully restore the state due to missing instruments (illuminator probe(s) not detected)•   User Advisory - See Discrepancy description
1404AdvisoryPneumatics recovery complete. Vitreous probe status not fully restored.PneumaticsSubmoduleRecoveryPartialSuccessAdv The Pneumatics submodule recovered, but the system could not fully restore the state due to missing instruments (vitreous probe not detected)•   User Advisory - See Discrepancy description
1405AdvisoryAux Illuminator recovery complete. Illuminator probe status not fully restored.AuxIllumSubmoduleRecoveryPartialSuccessAdv The Auxiliary Illuminator submodule recovered, but the system could not fully restore the state due to missing instruments (illuminator probe(s) not detected)•   User Advisory - See Discrepancy description
1406AdvisoryLaser recovery complete. Laser probe status not fully restored.LaserSubmoduleRecoveryPartialSuccessAdv The Laser submodule recovered, but the system could not fully restore the state due to missing instruments (laser probe(s) not detected).•   User Advisory - See Discrepancy description
1407AdvisoryFault recovery complete. AutoSert* Handpiece status not fully restored.IolFunctionalityNotRestoredAfterFaultRecoveryAdv An IOL handpiece was inserted when a Fault occurred. After recovery the system will not restore the handpiece state and the user is notified via this advisory.•   User Advisory - See Discrepancy description
2100FaultFault - 2100
Call Field Service
SUP_HOST_HEARTBEAT_TIMEOUT – A two second heart beat reply from the host was missed.• Cable W17
• Host Module
• Supervisor
2200ErrorCommunications failure with the Fluidics submodule. Fluidics functions will be disabled.SUP_FLUIDICS_HUNT_FAIL_ERR - The supervisor cannot establish communicate with the Fluidics submodule.• Cable W24
• Supervisor
• Fluidics Controller PCB
2201ErrorCommunications failure with the Fluidics submodule. Fluidics functions will be disabled.SUP_FLUIDICS_COMM_FAIL_ERR The supervisor lost communicate with the Fluidics submodule• Cable W24
• Supervisor
• Fluidics Controller PCB
2202AdvisoryUnable to perform this function without a primed cassette.SUP_CASSETTE_NOT_READY_ADV - This advisory is produced when extraction is attempted without a primed cassette.•   User Advisory - See Discrepancy description
2203AdvisoryUnable to aspirate. Please turn on infusion.SUP_INFUSION_IRRIGATION_OFF_ADV - This advisory is produced when extraction is attempted without infusion on.•   User Advisory - See Discrepancy description
2204AdvisoryUnable to aspirate while infusion/irrigation is unavailable.SUP_INFUSION_UNAVAILABLE_ADV - This advisory is produced when extraction is attempted without infusion being functional.•   User Advisory - See Discrepancy description
2205AdvisoryPlease wait: draining cassette.SUP_EXT_CHMBR_OVER_ADV - This advisory is produced when extraction is attempted when the extraction chamber is in a overflow condition.•   User Advisory - See Discrepancy description
2206AdvisoryUnable to turn on infusion without a primed cassette.SUP_INF_CASSETTE_NOT_READY_ADV - This advisory is produced when infusion is attempted with out a primed cassette.•   User Advisory - See Discrepancy description
2207AdvisoryUnable to turn on infusion without a calibrated IV pole.SUP_INF_IV_POLE_NOT_CALIB_ADV - This advisory is produced when infusion is attempted with out a calibrated IV Pole.•   User Advisory - See Discrepancy description
2208AdvisoryUnable to turn on infusion without sufficient source pressure.SUP_INF_NO_AIR_PRESSURE_ADV - This advisory is produced when infusion is attempted with out source air pressure•   User Advisory - See Discrepancy description
2209AdvisoryUnable to turn on irrigation without a primed cassette.SUP_IRR_CASSETTE_NOT_READY_ADV - This advisory is produced when irrigation is attempted with out a primed cassette.•   User Advisory - See Discrepancy description
2211AdvisoryUnable to turn on irrigation without sufficient source pressure.SUP_IRR_NO_AIR_PRESSURE_ADV - This advisory is produced when irrigation is attempted with out suitable source pressure.•   User Advisory - See Discrepancy description
2212AdvisoryIrrigation unavailable: out of fluid. Please change the bottle.SUP_IRR_NO_FLUID_ADV - This advisory is produced when irrigation is attempted with out sufficient irrigation fluid available•   User Advisory - See Discrepancy description
2213AdvisoryFAX unavailable. Please insert a cassette.SUP_FAX_NO_CASSETTE_ADV - This advisory is produced when FAX is attempted with out a cassette inserted.•   User Advisory - See Discrepancy description
2214ErrorExtraction setpoint timeout. Infusion/Irrigation and Extraction functions will be disabled.SUP_EXT_NO_ACTIVE_UPDATE_ERR – This error is produced when the extraction is active and the extraction proxy fails to get a footswitch update within the required timeframe.•   Supervisor
•   Footswitch
•   Fluidics Controller PCB
2215AdvisoryMicro reflux is currently not available.SUP_EXT_NO_REFLUX_FLUID_ADV – This error is produced when the user attempts to activate micro reflux and the fluid level is out of range.•   User Advisory - See Discrepancy description
2216AdvisoryPlease change the Bottle to regain extraction.SUP_EXT_NO_INFUSE_FLUID_ADV – This error is generated when aspiration is attempted when the infusion in the cassette is empty•   User Advisory - See Discrepancy description
2217AdvisoryExtraction is not allowed while in 30 mmHg backup mode.SUP_EXT_INFUSE_IN_BACKUP_ADV – This advisory occurs when Extraction is attempted while in manual backup.•   User Advisory - See Discrepancy description
2218AdvisoryIrrigation is not allowed while in 30 mmHg backup mode.SUP_IRR_INFUSE_IN_BACKUP_ADV – This advisory occurs when Irrigation is attempted while in manual backup.•   User Advisory - See Discrepancy description
2250ErrorCommunications failure with the Pneumatics submodule. Pneumatics functions will be disabled.SUP_PNEUMATICS_HUNT_FAIL_ERR – The supervisor cannot communicate with the Pneumatics submodule.•   Cable W21
•   Supervisor
•   Pneumatic Controller
2251ErrorCommunications failure with the Pneumatics submodule. Pneumatics functions will be disabled.SUP_PNEUMATICS_COMM_FAIL_ERR – This error is produced when the Supervisor losses communication with the pneumatics module•   Cable W21
•   Supervisor
•   Pneumatic Controller
2252AdvisoryUnable to enable function without sufficient source pressure.SUP_PNEU_NO_AIR_PRESSURE_ADV - This advisory is produced when cutter or utility operation is requested with out sufficient source pressure.•   User Advisory - See Discrepancy description
2253AdvisoryUnable to enable function with excessive source pressure.SUP_PNEU_HIGH_AIR_PRESSURE_ADV - This advisory is produced when cutter or utility operation is requested with source pressure that is too high.•   User Advisory - See Discrepancy description
2254ErrorPneumatics setpoint timeout. Pneumatics functions will be disabled.SUP_PNEU_NO_ACTIVE_UPDATE_ERR - This error is produced when the pneumatics is active and the pneumatics proxy fails to get a footswitch update within the required timeframe.•   Supervisor
•   Footswitch
•   Pneumatics Controller PCB
2300ErrorCommunications failure with the Ultrasound submodule. Ultrasound functions will be disabled.SUP_ULTRASOUND_HUNT_FAIL_ERR – Thesupervisor cannot communicate with the Ultrasound submodule.•   Cable W20
•   Supervisor
•   U/S Controller PCB
2301ErrorCommunications failure with the Ultrasound submodule. Ultrasound functions will be disabled.SUP_ULTRASOUND_COMM_FAIL_ERR - The supervisor lost communication with the Ultrasound submodule•   Cable W20
•   Supervisor
•   U/S Controller PCB
2302ErrorUS setpoint timeout. US, IOL injector and Diathermy functions will be disabled.SUP_US_NO_ACTIVE_UPDATE_ERR - This error is produced when the US is active and the US proxy fails to get a footswitch update within required timeframe.•   Supervisor
•   Footswitch
•   Fluidics Controller PCB
2350ErrorCommunications failure with the Laser submodule. Laser functions will be disabled.SUP_LASER_HUNT_FAIL_ERR - The supervisor cannot communicate with the Laser submodule.• Communication Cable
• Supervisor
• Laser Core Module
2351AdvisoryCommunications failure with the Laser submodule. Laser functions will be disabled.SUP_LASER_COMM_FAIL_ADV - The supervisor lost communication with the Laser submodule• This message is normally displayed
whenever the key is turned off. If displayed
any other time check:
• Communication Cable
• Supervisor
• Laser Core Module
2400AdvisoryPlease insert the Table Top Illuminator drawer.SUP_ILLUM_DRAWER_OUT_ADV – Supervisor detected a non –zero setpoint with the Table Top Illuminator drawer out•   User Advisory - See Discrepancy description
2401AdvisoryThe lamp in the Table Top Illuminator needs to be replaced. Please contact Field Service.SUP_TT_ILLUM_LAMP_BAD_ADV - Supervisor detected a non–zero setpoint with the Table Top Illuminator Lamp bad.•   User Advisory - See Discrepancy description
2500AdvisoryCommunications failure with the Power Control submodule. Please contact Field Service.SUP_POWER_MONITOR_COMM_FAIL_ERR – Supervisor detected a serial I/O error when trying to talk to the Power Module• Reset needed for Power Controller - Use
PCB reset switch or press and hold in
Standby switch to power down
• W48 Cable Assy
• Power Controller PCB
• Supervisor
2550ErrorFootswitch error. Footswitch treadle functions will be disabled.SUP_FOOTSWITCH_ERR – The footswitch has an error.• Footswitch
• Supervisor
• Cable W22
• Cable W01 (Base Footswitch Cable)
2700ErrorCommunications failure with the Auxiliary Illuminator submodule. Auxiliary Illuminator functions will be disabled.SUP_AUX_ILLUMINATOR_HUNT_FAIL_ERR – The supervisor cannot communicate with the Auxiliary Illuminator submodule.• Supervisor
• Aux illuminator PCB
2701AdvisoryCommunications failure with the Auxiliary Illuminator submodule. Auxiliary Illuminator functions will be disabled.SUP_AUX_ILLUM_COMM_FAIL_ADV - The supervisor lost communication with the Auxiliary Illuminator submodule• Supervisor
• Aux illuminator PCB
2702AdvisoryThe lamp in the Auxiliary Illuminator needs to be replaced. Please contact Field Service.SUP_AUX_ILLUM_LAMP_BAD_ADV - Supervisor detected a non–zero setpoint with the Auxiliary Illuminator Lamp bad.•   User Advisory - See Discrepancy description
2750FaultFault - 2750
Call Field Service
SUP_RESET_FAIL_ERR – The Supervisor could not assert reset control over all submodules. A module that was supposed to be reset responded with its module information.•   Supervisor
•   Module not responding
2751FaultFault - 2751
Call Field Service
SUP_SLOT_ID_ERR – The Supervisor got duplicate slot ID for the submodules.•   Supervisor
•   Module not responding
3100AdvisoryCould not calibrate the fluid level sensors. Please eject and reinsert, or replace the cassette.FLD_CMN_LEVEL_SENSOR_CAL_ERR - Level Sensor calibration failed. Possible reasons:
1) the calibration done status bit was not set by the FPGA within the specified timeout limit
2) the calculated required max/min pixel gain was outside the valid range
3) the brightness level of the level sensor LEDs could not be adjusted.
• Cassette
• Level Sensor
• Fluidics Controller PCB
• Fluidics Module
3200ErrorCassette latch error. Infusion/irrigation and extraction functions will be disabled.CASSETTE_LATCH_ERR - Cassette latch hardware feedback signals indicate a hardware problem.• Cassette
• Mechanical interference with latch
mechanism
• Latch Cylinder
• Fluidics Controller PCB
• Fluidics Module
3201AdvisoryInvalid cassette ID detected. Please eject and insert a valid cassette.CASSETTE_ID_INVALID_ERR - An invalid cassette id was read by the cassette id sensors.• Cassette
• Cassette ID Sensors
• Suction PCB
• Fluidics Controller PCB
• Fluidics Module
3202AdvisoryCassette test failed. Please eject and reinsert, or replace cassette.CASSETTE_TEST_FAILED_ERR Cassette pressure and/or vacuum tests failed.• Cassette
• Vacuum or Pressure leak
• Fluidics Controller
• FluidIcs Module
3203AdvisoryThe cassette was not properly latched into position. Please remove and reinsert the cassette.CASSETTE_NOT_LATCHED_ADV The cassette latch optical position sensor indicates that the latch did not reach its locked position.• Cassette
• Mechanical interference with latch
mechanism
• Latch Cylinder
• Fluidics Controller PCB
• Fluidics Module
3204ErrorCassette ID sensor error. Infusion/irrigation and extraction functions will be disabled.CASSETTE_ID_SENSOR_ERR Cassette ID sensor test failed. The sensor output voltage is not within the expected range.• Cassette
• Cassette ID Sensors
• Suction PCB
• Fluidics Controller PCB
• Fluidics Module
3205AdvisoryThe Fluidics module fan is not working.FLD_MODULE_FAN_TAC_ADV The Fluidics module fan tachometer indicates that the Fluidics module fan is not operating.• Fluidics Controller PCB
• Fan
• Fluidics Module
3207InformationCassette cannot be loaded. Please adjust the inlet pressure between 58 psi and 120 psi.CASS_NO_AIR_PRSR_INFO A cassette was inserted in the receiver mechanism while the wall pressure is outside its valid operating range.•   User Information - See Discrepancy description
3300AdvisoryDraining cassette. Please wait.EXT_CHAMBER_OVERFLOW_ADV The cassette aspiration chamber is full of fl• Cassette not draining to drain bag
• Fluid Level Sensors
• Fluidics Controller PCB
• Fluidics Module
3302AdvisoryFault - 3302
Call Field Service
EXT_NOT_IN_OP_RANGE_ADV Not enough or too much fluid in the aspiration chamber to allow aspiration flow mode and reflux. NOTE: This advisory is never explicitly displayed. Instead advisory 1239 or 1240 might be displayed if appropriate. See 1239 and 1240 for details.•   User Advisory - See Discrepancy description
3304AdvisoryLeak test failure. Please confirm the irrigation tubing, aspiration tubing, and the test chamber are properly connected to the handpiece.EXT_HP_TUBING_LEAK_ADV A leak in the aspiration or irrigation tubing was detected during flow check of a handpiece.•   User Advisory - See Discrepancy description
3305AdvisoryPriming of the aspiration probe was unsuccessful. Please attempt to re-prime the probe.EXT_PROBE_PRIME_FAILED_ADV Probe prime was unsuccessful. The required volume of fluid was not transferred through the aspiration tubing set within the specified timeout period.•   User Advisory - See Discrepancy description
3306AdvisoryPriming of the aspiration handpiece was unsuccessful. Please attempt to re-prime the handpiece.EXT_HP_PRIME_FAILED_ADV Handpiece prime was unsuccessful. The required volume of fluid was not transferred through the aspiration tubing set within the specified timeout period.•   User Advisory - See Discrepancy description
3307ErrorAspiration level sensor problem detected. Infusion/ irrigation and extraction functions will be disabled.EXT_LVL_STATUS_ERR The level sensor status signals indicate a level sensor hardware problem.• Aspiration Level Sensor
• Cable W93
• Fluidics Controller PCB
• Fluidics Module
3308AdvisoryFlow check failure: measured flow restriction is too high. Extraction and Ultrasound functions in Phaco/Frag will be disabled.EXT_FLOW_CHECK_FAILED_ADV The handpiece failed the flowcheck. Too high vacuum level was required to achieve the reference flow level.•   User Advisory - See Discrepancy description
3309AdvisoryFlow check failure: aspiration chamber could not be filled with fluid. Extraction and Ultrasound functions in Phaco/Frag will be disabled.EXT_FILL_TIMEOUT_ADV The handpiece failed the flowcheck. Too high vacuum level was required to achieve the reference flow level.•   User Advisory - See Discrepancy description
3325AdvisoryDrain bag is almost full. Please replace bag and press [Done].EXT_BAG_NEAR_FULL_ADV ~400 cc of fluid has been pumped into the drain bag.•   User Advisory - See Discrepancy description
3326AdvisoryDrain bag is full. Please replace bag and press [Done].EXT_BAG_FULL_ADV - ~450 cc of fluid has been pumped into the drain bag.•   User Advisory - See Discrepancy description
3327AdvisoryDrain bag is critically full. Please replace bag and press [Done].EXT_BAG_CRITICALLY_FULL_ADV ~500 cc of•   User Advisory - See Discrepancy description
fluid has been pumped into the drain bag.
3329ErrorDrain pump problem detected. Infusion/irrigation and extraction functions will be disabled.EXT_PUMP_ROTATION_ERR The commanded pump rate does not correspond to the actual pump rate measured by the optical encoder.• Drain Pump
• Fluidics Controller PCB
• Fluidics Module
3330AdvisoryThe drain pump fan is not working.EXT_PUMP_FAN_STATUS_ERR The drain pump fan tachometer indicates that the drain pump fan is not operating.• Fan
• Fluidics Controller PCB
• Fluidics Module
3331AdvisoryCould not drain the aspiration chamber. Please remove and insert a new cassette.EXT_CHAMBER_DRAIN_TIMEOUT_ADV The• Cassette
• Drain Pump
• Aspiration Level Sensor
• Fluidics Controller PCB
• Fluidics Modu
3350ErrorExtraction pressure transducer offset error. Infusion/irrigation and extraction functions will be disabled.EXT_PT_OFFS_ERR Extraction pressure transducer 0 offset out of range.• Cable W86
• Fluidics Controller PCB
• Suction PCB
• Transducer MT7 and/or MT8
• Fluidics Module
3351ErrorExtraction pressure transducer discrepancy error. Infusion/irrigation and extraction functions will be disabled.EXT_PT_DISCR_ERR - A discrepancy between the primary and redundant extraction pressure transducers was detected.• Transducer MT7 and/or MT8
• Fluidics Controller PCB
• Suction PCB
• Cable W86
• Fluidics Module
3352ErrorExtraction isolation valve error. Infusion/irrigation and extraction functions will be disabled.EXT_ISO_VLV_ERR Extraction isolation valve hardware error.• L14
• Fluidics Controller PCB
• Suction PCB
• Cable W86
• Fluidics Module
3353ErrorReflux valve error. Infusion/irrigation and extraction functions will be disabled.EXT_REFLUX_VLV_ERR Extraction reflux valve hardware error.• L15
• Fluidics Controller PCB
• Suction PCB
• Cable W86
• Fluidics Module
3354ErrorExtraction output valve (S1) error. Infusion/ irrigation and extraction functions will be disabled.EXT_OUTPUT_VLV_ERR Extraction output pincher valve (S1) error• 5 Valve PCB, Yellow Lower
• Cable W90
• Fluidics Controller PCB
• Fluidics Module
3355ErrorExtraction output valve (S11) error. Infusion/ irrigation and extraction functions will be disabled.EXT_PORT1_VLV_ERR Extraction output port 1 pincher valve error.• 3 Valve PCB, Green
• Cable W90
• Fluidics Controller PCB
• Fluidics Module
3356ErrorExtraction output valve (S22) error. Infusion/ irrigation and extraction functions will be disabled.EXT_PORT2_VLV_ERR Extraction output port 2 pincher valve error.• 3 Valve PCB, Green
• Cable W90
• Fluidics Controller PCB
• Fluidics Module
3357ErrorExtraction cross-connection valve (SC) error. Infusion/irrigation and extraction functions will be disabled.EXT_CROSS_NO_VLV_ERR Extraction “normally open” cross-connection valve error.• 5 Valve PCB, Red Upper
• Cable W90
• Fluidics Controller PCB
• Fluidics Module
3358ErrorExtraction cross-connection valve (SC2) error. Infusion/irrigation and extraction functions will be disabled.EXT_CROSS_NC_VLV_ERR Extraction “normally closed” cross-connection valve error.• 5 Valve PCB, Red Upper
• Cable W90
• Fluidics Controller PCB
• Fluidics Module
3359AdvisorySuction pressure surges detected. Vacuum will be disabled. Please release the footswitch treadle to reset.EXT_PRSR_OSCILLATION_ADV Extraction pressure oscillations detected.• Bubbles in tubing - Check Cassette
connections and/or reprime
• Transducers MT7 & MT8
• Fluidics Controller PCB
• Suction PCB
• Cable W86
• Fluidics Module
3360AdvisorySuction flow surges detected. Flow will be disabled. Please release the footswitch treadle to reset.EXT_FLOW_OSCILLATION_ADV Extraction flow oscillations detected.• Bubbles in tubing - Check Cassette
connections and/or reprime
• Transducers MT7 & MT8
• Fluidics Controller PCB
• Suction PCB
• Cable W86
• Fluidics Module
3361AdvisorySuction pressure is too high. Vacuum will be disabled. Please release the footswitch treadle to reset.EXT_PRSR_OVERSHOOT_ADV Extraction pressure overshoot detected.• Bubbles in tubing - Check Cassette
connections and/or reprime
• Transducers MT7 & MT8
• Fluidics Controller PCB
• Suction PCB
• Cable W86
• Fluidics Module
3362AdvisoryAspiration flow too high. Flow will be disabled. Please release the footswitch treadle to reset.EXT_FLOW_OVERSHOOT_ADV Extraction flow overshoot detected.• Bubbles in tubing - Check Cassette
connections and/or reprime
• Fluidics Controller PCB
• Suction PCB
• Cable W86
• Fluidics Module
3363ErrorExtraction pressure transducer reference voltage out of range. Infusion/irrigation and extraction functions will be disabled.EXT_TRANSDUCER_REFERENCE_ERR Extraction transducer reference voltage error.• Transducers MT7 & MT8
• Fluidics Controller PCB
• Fluidics Module
3364AdvisoryAspiration manifold input pressure is too high.EXT_75_PSI_REGULATOR_ADV Extraction 75 psi pressure regulator advisory. The aspiration manifold 75 psi regulator is out of regulation. A pressure higher than 220 mmHg was measured during the regulator pressure test which is part of the cassette test.• Relief Valve
• Transducers MT7 & MT8
• Fluidics Controller PCB
• Fluidics Module
3400ErrorInfusion source container pressure transducer offset error. Infusion/irrigation and extraction functions will be disabled.PRSR_SRC_PT_OFFS_ERR Infusion source container pressure transducer 0 offset out of range.• Fluidics Infusion PCB
• Cable W88
• Fluidics Controller PCB
• Fluidics Module
3401ErrorInfusion source container isolation valve error. Infusion/irrigation and extraction functions will be disabled.PRSR_SRC_ISO_VLV_ERR Infusion source container isolation valve error.• Infusion 5 Valve PCB
• Fluidics Infusion PCB
• Cable W88
• Fluidics Controller PCB
• Fluidics Module
3402ErrorInfusion source container transducer discrepancy error. Infusion/irrigation and extraction functions will be disabled.PRSR_SRC_PT_DISCR_ERR A dispcrepancy between the source container and LPAS pressure transducers was detected.• Fluidics Infusion PCB
• Cable W88
• Fluidics Controller PCB
• Fluidics Module
3403ErrorInfusion source container pressure too high. Infusion/irrigation and extraction functions will be disabled.PRSR_SRC_HI_ERR Infusion source container pressure too high.• Proportional Valve - PV3 or PV6
• Fluidics Infusion PCB
• Cable W88
• Fluidics Controller PCB
• Fluidics Module
3420ErrorInfusion pressure transducer offset error. Infusion/ irrigation and extraction functions will be disabled.INF_PT_OFFS_ERR Infusion pressure transducer 0 offset out of range• Cable W88
• Fluidics Infusion PCB
• Fluidics Controller PCB
• Fluidics Module
3421ErrorInfusion pressure transducer discrepancy error. Infusion/irrigation and extraction functions will be disabled.INF_PT_DISCR_ERR A dispcrepancy between the primary and redundant infusion pressure transducers was detected.• Cable W88
• Fluidics Infusion PCB
• Fluidics Controller PCB
• Fluidics Module
3422ErrorInfusion isolation valve error. Infusion/irrigation and extraction functions will be disabled.INF_ISO_VLV_ERR Infusion isolation valve error.• Infusion 5 Valve PCB
• Cable W88
• Fluidics Infusion PCB
• Fluidics Controller PCB
• Fluidics Modu
3423ErrorInfusion FAX valve error. Infusion/irrigation and extraction functions will be disabled.INF_FAX_VLV_ERR Infusion FAX valve error.• Infusion 5 Valve PCB
• Cable W88
• Fluidics Infusion PCB
• Fluidics Controller PCB
• Fluidics Modu
3424ErrorInfusion input valve error. Infusion/irrigation and extraction functions will be disabled.INF_INPUT_VLV_ERR Infusion chamber input pincher valve error.• 5 Valve PCB, Red Upper
• Cable W90
• Fluidics Controller PCB
• Fluidics Module
3425ErrorInfusion output valve error. Infusion/irrigation and extraction functions will be disabled.INF_OUTPUT_VLV_ERR Infusion chamber output pincher valve error.• 5 Valve PCB, Red Upper
• Cable W90
• Fluidics Controller PCB
• Fluidics Module
3426ErrorIrrigation output valve error. Infusion/irrigation and extraction functions will be disabled.IRR_OUTPUT_VLV_ERR Irrigation output pincher valve error.• Cassette (fluid can interfere with sensor
• detection and cylinder movement)
• CY4-CO Cylinder
• Optical position sensor
• 3 Valve PCB, Green
• Cable W90
• Fluidics Controller PCB
• Low pressure or pressure leak.
• Fluidics Module
3427ErrorInfusion pressure surges detected. Infusion/ irrigation and extraction functions will be disabled.INF_OSCILLATION_ERR Infusion pressure oscillations detected.• Infusion Proportional Valve
• Fluidics Infusion PCB
• Cable W88
• Fluidics Controller PCB
• Fluidics Module
3428ErrorInfusion pressure too high. Infusion/irrigation and extraction functions will be disabled.INF_HI_PRSR_ERR Infusion high pressure detected.• Infusion Proportional Valve
• Fluidics Infusion PCB
• Cable W88
• Fluidics Controller PCB
• Fluidics Module
3429AdvisoryLow infusion pressure detected. Please check infusion connections. Select 30 mmHg liquid backup pressure, or ignore low pressure condition.INF_LO_PRSR_ADV Infusion pressure too low.•   User Advisory - See Discrepancy description
3431AdvisoryLow irrigation pressure detected. Please check irrigation connections.IRR_LO_PRSR_ADV Irrigation pressure too low.•   User Advisory - See Discrepancy description
3433ErrorInfusion chamber isolation valve error. Infusion/ irrigation and extraction functions will be disabled.INF_ISO_VLV_STUCK_CLOSED_ERR Infusion chamber isolation valve detected to be stuck closed during power up diagnostics.• Infusion 5 Valve PCB
• Cable W88
• Fluidics Infusion PCB
• Fluidics Controller PCB
• Fluidics Module
3434AdvisoryInfusion chamber overflow error. Control of infusion pressure has possibly been lost. The current pressure, in mmHg, could be as high as:INF_CHAMBER_OVERFLOW_ADV Infusion chamber overflow error. The fluid level in the infusion chamber reached above the overflow level.•   User Advisory - See Discrepancy description
3436ErrorFAX valve error. Infusion/irrigation and extraction functions will be disabled.INF_FAX_VLV_STUCK_CLOSED_ERR FAX
valve detected to be stuck closed during power up
diagnostics.
• Infusion 5 Valve PCB
• Cable W88
• Fluidics Infusion PCB
• Fluidics Controller PCB
• Fluidics Module
3437ErrorInfusion level sensor error. Infusion/irrigation and extraction functions will be disabled.INF_LVL_STATUS_ERR The level sensor status
signals indicate a level sensor hardware problem
• Infusion Level Sensor
• Cable W94
• Fluidics Controller PCB
• Fluidics Module
3438AdvisoryNo more infusion fluid available. Press [Change] to change the fluid container.INF_FLUID_OUT_ADV The infusion chamber in the cassette is empty.•   User Advisory - See Discrepancy description
3440AdvisoryNo more irrigation fluid available. Press [Change] to change the fluid container..IRR_FLUID_OUT_ADV The irrigation chamber in the cassette is empty.•   User Advisory - See Discrepancy description
3442AdvisoryIrrigation chamber overflow detected. Control of irrigation pressure has possibly been lost. The current pressure, in mmHg, could be as high as:IRR_CHAMBER_OVERFLOW_ADV Irrigation chamber overflow error. The fluid level in the irrigation chamber reached above the overflow level.•   User Advisory - See Discrepancy description
3444AdvisoryLow infusion pressure. Please check connections. Select 30 mmHg air backup, or ignore low pressure condition. Backup mode will not allow extraction.INF_FAX_MS_LO_PRSR_ADV Infusion pressure too low while in FAX with a
Manual Stopcock cassette.
•   User Advisory - See Discrepancy description
3460ErrorInfusion backup valve error. Infusion/irrigation, extraction, and Ultrasound functions will be disabled.INF_BACKUP_VLV_ERR Infusion backup valve error.• Infusion 5 Valve PCB
• Cable W88
• Fluidics Infusion PCB
• Fluidics Controller PCB
• Fluidics Module
3461ErrorInfusion LPAS pump error detected. Infusion/ irrigation, extraction, and Ultrasound functions will be disabled.INF_LPAS_TAC_ERR The infusion LPAS pump tachometer indicate that the LPAS pump is not operating correctly.• LPAS Pump
• Fluidics Controller PCB
• Fluidics Module
3462AdvisoryInfusion flow sensor communication error. IOP Control functions will be disabled.INF_FLOW_SENSOR_COM_ERR Infusion flow sensor communication error.• NIFS Flow Sensor
• Fluidics Controller PCB
• Fluidics Module
3464AdvisoryInfusion flow data invalid: IOP Control functions will be disabled. Check infusion tubing for air bubbles.INF_FLOW_DATA_INVALID_ADV Infusion flow sensor readings are not valid.• Bubbles
• Cassette Connections
• Flow Sensor
• Fluidics Controller PCB
• Fluidics Module
3466AdvisoryInfusion flow sensor accuracy error. IOP Control functions will be disabled.INF_FLOW_ACCURACY_ADV Infusion flow sensor accuracy error. Flow sensor and chamber fluid volume measurements are out of range of each other. Detected during flow sensor calibration.• Cassette
• Infusion Level Sensor
• Cable W94
• Flow Sensor
• Flow PCB
• Fluidics Controller PCB
• Fluidics Module
3467AdvisoryInfusion flow sensor signal amplitude is low. IOP Control functions will be disabled. Please eject and reinsert the cassette.INF_FLOW_SIGNAL_AMPLITUDE_ADV Infusion flow sensor signal amplitude error. Detected during flow sensor calibration• Wrong infusion cannula selected
• Cassette
• Flow Sensor
• Flow PCB
• Fluidics Controller PCB
• Fluidics Module
3469AdvisoryTubing calibration offset error. IOP Control functions will be disabled. Please position the infusion cannula at the height of the center of the cassette and re-prime.INF_TUBING_CAL_OFFSET_ADV Infusion tubing calibration 0 offset advisory. The infusion cannula was not positioned within the correct vertical range of the cassette during calibration.• Cassette
• Flow Sensor
• Flow PCB
• Fluidics Controller PCB
• Fluidics Module
3470AdvisoryCalibration verification error: calculated pressures are not within the expected range. IOP Control functions will be disabled. Please re-prime.INF_TUBING_CAL_CHECK_PT_ADV Infusion tubing calibration check point advisory. The pressure drop calculated using the acquired calibration profile was not within the expected range.• Cassette
• Flow Sensor
• Flow PCB
• Fluidics Controller PCB
• Fluidics Module
3471AdvisoryNoisy calibration flow readings. IOP Control functions will be disabled. Please re-prime.INF_TUBING_CAL_STDDEV_ADV Infusion tubing calibration standard deviation advisory.
The standard deviation between the acquired flow measurements and the calculated calibration profile was larger than the specified max limit.
• Wrong infusion cannula selected
• Cassette
• Flow Sensor
• Flow PCB
• Fluidics Controller PCB
• Fluidics Module
3472AdvisoryInfusion chamber leak detected. Please eject and replace the cassette.INF_CHAMBER_LEAK_ADV – A leak in the infusion chambers was detected during priming.•   Cassette
3473ErrorInfusion pressure transducer reference voltage out of range. Infusion/irrigation and extraction functions will be disabled.INF_TRANSDUCER_REFERENCE_ERR Infusion transducer reference voltage error.• Cable W88
• Fluidics Infusion PCB
• Fluidics Controller PCB
• Fluidics Module
3474ErrorInfusion NIFS valve error detected. Infusion/ irrigation and extraction functions will be disabled.INF_NIFS_VLV_ERR Infusion NIFS valve error.• Cable W88
• Fluidics Infusion PCB
• Fluidics Controller PCB
• Fluidics Module
3475AdvisoryInfusion prime failed.INF_PRIME_FAILED_ADV Infusion prime failed.• Infusion line clamped
• Cassette drip chamber check valve stuck
• Cassette
• Cassette Connectors
• Stuck AI or BI pincher
• Loss of 45psi
• 5 Valve PCB, Yellow Lower
• Cable W90
• Fluidics Controller PCB
• Infusion Level Sensors
• Faulty MT5 “S1” Pressure Sensor
• Fluidics Module
3476AdvisoryThe infusion chambers did not fill with fluid. Please check the infusion bottle and connections or press [Change] to replace the infusion bottle.INF_CHAMBER_FILL_TIMEOUT_ADV Infusion chamber did not fill within the specified timeout period. Source container is out of fluid.• User Advisory - See Discrepancy
description
3477ErrorInfusion LPAS pressure transducer offset error. Infusion/irrigation and extraction functions will be disabled.INF_LPAS_PT_OFFS_ERR Infusion LPAS pressure transducer 0 offset out of range.• Cable W88
• Fluidics Infusion PCB
• Fluidics Controller PCB
• Fluidics Module
3478ErrorInfusion LPAS pressure error. Infusion/irrigation and extraction functions will be disabled.INF_LPAS_PRSR_ERR Infusion LPAS source pressure too high.• Proportional valve
• Cable W88
• Fluidics Infusion PCB
• Fluidics Controller PCB
• Fluidics Module
3479AdvisoryInfusion LPAS pressure is low. Infusion pressure loss is possible.INF_LPAS_PRSR_LOW_ADV Infusion LPAS pump output pressure is low.• Proportional valve
• Cable W88
• Fluidics Infusion PCB
• Fluidics Controller PCB
• Fluidics Module
3481AdvisoryInfusion flow sensor is disconnected. IOP Control functions will be disabled.INF_FLOW_SENSOR_DISCONNECT_ADV Infusion flow sensor is disconnected or the sensor connection has failed.•   User Advisory - See Discrepancy description
3482AdvisoryAuto infusion valve failed to open during prime. Please replace infusion tubing and attempt to re- prime.The AIV valve flow check performed as part of the cassette prime sequence failed.•   User Advisory - See Discrepancy description
4100ErrorU/S voltage failure (+5 analog). Ultrasound and Diathermy functions will be disabled.US_5_VOLT_ANALOG_ERR - US Kernel Analog 5 volt feedback is bad. US Submodule is non- functional.•   U/S Controller PCB
4101ErrorU/S voltage failure (+2.5). Ultrasound and Diathermy functions will be disabled.US_PLUS_2_5_VOLT_ERR - US Kernel +2.5 volt feedback is bad. US submodule is non-functional.•   U/S Controller PCB
4102ErrorU/S voltage failure (-2.5). Ultrasound and Diathermy functions will be disabled.US_MINUS_2_5_VOLT_ERR US Kernel -2.5 volt feedback is bad. US submodule is non-functional.•   U/S Controller PCB
4103ErrorU/S voltage failure (+8.5). Ultrasound and Diathermy functions will be disabled.US_8_5_VOLT_POWER_ERR US Kernel 2.5 volt feedback is bad. US submodule is non-functional.•   U/S Controller PCB
4111ErrorUS failure: SPI driver write timeout. Ultrasound and Diathermy functions will be disabled.SPI_WRITE_TIMEOUT_ERR SPI driver timed out waiting for a write to complete. US submodule is non-functional.•   U/S Controller PCB
4200AdvisoryHandpiece EEPROM CRC is invalid. Please replace the handpiece.US_HP_BAD_CRC_ADV The handpiece EEPROM CRC is invalid. Handpiece needs to be replaced.•   User Advisory - See Discrepancy description
4201AdvisoryOnly one US handpiece may be connected at a time. Please remove one of the handpieces.US_HP_MULTIPLE_ADV Two US handpieces are connected. One of the handpieces must be removed before the other handpiece can be used.•   User Advisory - See Discrepancy description
4202AdvisoryHandpiece current is too low. Please replace handpiece and re-tune.US_HP_CURRENT_ERR U/S handpiece current is too low. A short circuit in the handpiece can cause this.•   User Advisory - See Discrepancy description
4203AdvisoryHandpiece voltage is too low. Please replace handpiece and re-tune.US_HP_VOLTAGE_ERR U/S handpiece voltage is too low. An open circuit in the handpiece can cause this.•   User Advisory - See Discrepancy description
4204AdvisoryHandpiece power is too high. Please replace handpiece and re-tune.US_HP_EXCESSIVE_POWER_ERR U/S handpiece power output is too high.•   User Advisory - See Discrepancy description
4206AdvisoryHandpiece power DC2DC output was out of range. Please release the footswitch treadle to reset. If problem persists, please contact Field Services.US_HP_DC2DC_ERR While powering the handpiece, the DC2DC voltage for handpiece power is out of range.•   User Advisory - See Discrepancy description
4207AdvisoryHandpiece was removed while powered. Please reconnect handpiece and re-tune.US_HP_REMOVED_WHILE_POWERED_ADV User has disconnected the handpiece while it’s being powered.•   User Advisory - See Discrepancy description
4208AdvisoryData in handpiece EEPROM is out of range. Please replace the handpiece.US_HP_OUT_OF_RANGE_DATA_ADV Data contained in the handpiece is out of range.•   User Advisory - See Discrepancy description
4209AdvisoryUnknown US handpiece connected. Please connect a known handpiece.US_HP_UNKNOWN_HANDPIECE_ADV The system has detected that a US handpiece was connected but it cannot determine its type.•   User Advisory - See Discrepancy description
4210AdvisoryUnsupported US handpiece connected. Please connect a supported handpiece.US_HP_UNSUPPORTED_HANDPIECE_ADV The system has detected the connection of a recognized US handpiece but that handpiece is not supported by this system. Handpiece is not tuned.•   User Advisory - See Discrepancy description
4220AdvisoryTune failure – attempted while handpiece was in air. Please re-tune the handpiece.US_TUNE_IN_AIR_ADV The handpiece was tuned while in air. Handpiece is not tuned.•   User Advisory - See Discrepancy description
4221AdvisoryTune failure: handpiece was removed before tuning. Please connect a handpiece and re-tune.US_TUNE_NO_HANDPIECE_ADV A handpiece tune was requested but no handpiece is connected.•   User Advisory - See Discrepancy description
4222AdvisoryTune failure: handpiece is an unknown type. Please connect a known handpiece and re-tune.US_TUNE_UNKNOWN_HP_ADV A handpiece tune was requested but an unknown type of handpiece is connected.•   User Advisory - See Discrepancy description
4223AdvisoryTune failure: handpiece has a loose tip. Please tighten the tip and re-tune.US_TUNE_LOOSE_TIP_ADV The handpiece tip was loose when tuned. Handpiece is not tuned.•   User Advisory - See Discrepancy description
4224AdvisoryTune failure: handpiece current is low. Please replace handpiece and re-tune.US_TUNE_HP_CURRENT_ADV The handpiece current was too low (open circuit). Handpiece is not tuned.•   User Advisory - See Discrepancy description
4225AdvisoryTune failure: handpiece voltage is low. Please replace handpiece and re-tune.US_TUNE_HP_VOLTAGE_ADV The handpiece voltage was too low when tuned (short circuit). Handpiece is not tuned.•   User Advisory - See Discrepancy description
4226AdvisoryTune failure: handpiece frequency order error. Please re-tune.US_TUNE_FREQ_ORDER_ADV The series (low impedance) and parallel (high impedance) frequencies were out of order when the handpiece was tuned.•   User Advisory - See Discrepancy description
4228AdvisoryTune failure: handpiece series frequency margin error. Please re-tune.US_TUNE_FS_MARGIN_ERR The series tune frequency was too close to the tune start frequency while tuning the handpiece. The handpiece is not tuned.•   User Advisory - See Discrepancy description
4229AdvisoryTune failure: handpiece parallel frequency margin error. Please re-tune.US_TUNE_FP_MARGIN_ERR The parallel tune frequency was too close to the tune end frequency while tuning the handpiece. The handpiece is not tuned.•   User Advisory - See Discrepancy description
4231AdvisoryTune failure: handpiece frequency bandwidth too low. Please re-tune.US_TUNE_BW_LOW_ERR The difference between the series and parallel tune frequencies was too small while tuning the handpiece. The handpiece is not tuned.•   User Advisory - See Discrepancy description
4232AdvisoryTune failure: handpiece frequency bandwidth too high. Please re-tuneUS_TUNE_BW_HIGH_ERR The difference between the series and parallel tune frequencies was too large while tuning the handpiece. The handpiece is not tuned.•   User Advisory - See Discrepancy description
4234AdvisoryTune failure: handpiece DC2DC output out of range. Please re-tune. If problem persists, please contact Field Service.US_TUNE_DC2DC_ERR The DC2DC voltage was out of range while tuning the handpiece. The handpiece is not tuned.•   User Advisory - See Discrepancy description
4235AdvisoryTune failure: handpiece removed while tuning. Please connect handpiece and re-tune.US_TUNE_HP_REMOVED_ADV The handpiece was removed while tuning. The handpiece is not tuned.•   User Advisory - See Discrepancy description
4240AdvisoryThe requested Frag continuous power is too high. The Power level will be limited.US_FRAG_POWER_TOO_HIGH_ADV A request for more than 60% frag power was made while not in a pulsed mode. The power will be limited to 60%.•   User Advisory - See Discrepancy description
4400AdvisoryAutoSert* handpiece EEPROM CRC is invalid. Please replace the handpiece.IOL_BAD_CRC_ADV The AutoSert* handpiece EEPROM CRC is invalid. Handpiece needs to be replaced.•   User Advisory - See Discrepancy description
4401AdvisoryAutoSert* handpiece impeded. Retry handpiece operation. If problem persists, please contact Field Service.IOL_MOTOR_NOT_MOVING_ADV The AutoSert* handpiece motor is not moving – a move check timeout has occurred.•   User Advisory - See Discrepancy description
4402AdvisoryAutoSert* handpiece speed out of range. Retry handpiece operation. If problem persists, please contact Field Service.IOL_SPEED_OUT_OF_RANGE_ADV The AutoSert* handpiece motor speed is out of range.•   User Advisory - See Discrepancy description
4403AdvisoryAutoSert* handpiece travel out of range. Reinsert/ replace handpiece. If problem persists, please contact Field Service.IOL_SPEED_OUT_OF_RANGE_ADV The AutoSert* handpiece motor speed is out of range.•   User Advisory - See Discrepancy description
4404AdvisoryAutoSert* handpiece calibration failed. Reinsert/ replace handpiece. If problem persists, please contact Field Service.IOL_CALIBRATION_FAILED_ADV Calibration of the AutoSert* handpiece failed.
Possible causes:
1. The calibration did not complete before the 1 minute calibration timeout
2. The calculated handpiece speed is out of range
3. The handpiece travel is out of range
4. The handpiece motor is not moving
•   User Advisory - See Discrepancy description
4405AdvisoryAutoSert* handpiece must be in fully retracted position prior to autoclave. Reinsert handpiece to correct.IOL_HP_IMPROPER_DISCONNECT_ADV The AutoSert* handpiece was disconnected improperly with the plunger not in the retracted position.•   User Advisory - See Discrepancy description
4406AdvisoryAutoSert* handpiece data corrupted. Replace handpiece. If problem persists, please contact Field Service.IOL_CORRUPT_HP_DATA_ADV The AutoSert* handpiece EEPROM contains corrupted data. Handpiece needs to be replaced.•   User Advisory - See Discrepancy description
4407AdvisoryUnknown AutoSert* handpiece connected. Please connect a known handpiece.IOL_UNKNOWN_HP_ADV The system has detected that an AutoSert* handpiece was connected but it cannot determine its type.•   User Advisory - See Discrepancy description
4408AdvisoryUnexpected AutoSert* handpiece movement detected. Replace handpiece. If problem persists, please contact Field Service.IOL_UNEXPECTED_MOTOR_MOVE_ADV Unexpected AutoSert* handpiece movement \ feedback was detected. Handpiece needs to be replaced.•   User Advisory - See Discrepancy description
4409AdvisoryAutoSert* handpiece motor over current detected. Retry handpiece operation. If problem persists, replace handpiece and\or contact Field Service.IOL_MOTOR_OVERCURRENT_ADV AutoSert* handpiece motor over current condition has been detected.•   User Advisory - See Discrepancy description
4410AdvisoryPreloading of IOL using the treadle is not allowed until the AutoSert* plunger has been fully retracted.IOL_TRDLE_NOT_ALLOW_LPSTART_ADV User pressed the treadle to perform IOL Preloading when a Load Plunger command was previously started but not yet completed (and currently paused).•   User Advisory - See Discrepancy description
4411ErrorSubmodule Failure (1.8 volt). AutoSert* functions will be disabled.IOL_1_8_VOLTS_IS_BAD_ERR IOL Injector CPLD +1.8 Voltage is bad.•   User Advisory - See Discrepancy description
4250ErrorUltrasound failure: ADC calibration. Ultrasound functions will be disabled.US_ADC_CALIBRATE_ERR The ADC feedback reading with power off for DC2DC voltage, handpiece voltage, or handpiece current was too high.•   User Advisory - See Discrepancy description
4300ErrorDiathermy failure: DC2DC output was out of range. Diathermy functions will be disabled.DIA_AMPLIFIER_ERR While powering the handpiece, the DC2DC voltage for handpiece power is out of range.•   User Advisory - See Discrepancy description
4301AdvisoryDiathermy power is too high. Please release the footswitch treadle / button and try again.DIA_EXCESSIVE_POWER_ADV Too much power was being delivered to the diathermy handpiece. Diathermy power is turned off. The operator must release the treadle/switch and the depress the treadle/switch to re-activate power.•   User Advisory - See Discrepancy description
5100ErrorBallast failure (voltage). Illuminator functions will be disabled.ILM_VOLTAGE_SENSOR_ERR - Ballast transducer has reading that is out of specified safety range.• Cable W50
• Illuminator Controller PCB
• Table Top Illuminator Module
5101ErrorBallast failure (current). Illuminator functions will be disabled.ILM_CURRENT_SENSOR_ERR – Ballast transducer has reading that is out of specified safety range.• Cable W50
• Illuminator Control PCB
• Table Top Illuminator Module
5102AdvisoryLamp calibration data is corrupted: lamp needs to be calibrated. Please contact Field Service.ILM_CALIBRATION_DATA_ADV – The calibration data saved in the flash contains incorrect CRC checksum.• Calibration Data corrupt
5103AdvisoryFailure to turn lamp on: lamp needs to be replaced. Please contact Field Service.ILM_LAMP_SOURCE_FAULT_ADV – Lamp status reported by hardware is not the same as software status.• Lamp
• Illuminator Controller PCB
• Table Top Illuminator Module
5104AdvisoryThe lamp has exceeded its rated life: lamp needs to be replaced. Please contact Field Service.ILM_LAMP_REPLACEMENT_ADV – The lamp has exceeded its rated life. The new lamp is expected.• Lamp
5105AdvisoryThe lamp has exceeded its rated maximum life: lamp needs to be replaced. Please contact Field Service.ILM_LAMP_INTENSITY_LOW_ADV – The lamp has exceeded its rated safe life. The new lamp should be installed immediately.• Lamp
5107AdvisoryThe Table-Top Illuminator drawer is ejected. Please close the drawer to continue.ILM_MODULE_DRAWER_OUT_ADV – The module has been pulled out of drawer.• User Advisory - See Discrepancy
description
5109InformationThe calibration data for the Illuminator has changed: due to recalibration or replacement with a new unit.ILM_MODULE_REPLACED_INFO – The calibration data is changed..• User Advisory - See Discrepancy
5113AdvisoryMetrics data is corrupted; Lamp hour is invalid. Please contact Field Service.ILM_METRICS_CORRUPT_ADV – The metrics data for lamp hours is invalid. Either the Flash is corrupted or the checksum is incorrect.• Illuminator Controller PCB
• Table Top Illuminator Module
5200AdvisoryIlluminator optics temperature is high. The lamp will be turned off if the temperature continues to rise.ILM_OPTICS_TEMP_HIGH_ADV – Optics temperature is too high. Lamp is going to be shut down if temperature continues to rise.• Dirty Air Filter - located below Illuminator
• Optics Fan (W29)
• Illuminator Controller PCB
• Table Top Illuminator Module
5201ErrorIlluminator optics temperature has exceeded its limit. Illuminator functions will be disabled.ILM_OPTICS_TEMP_DOWN_ERR – Illuminator is shut down because Optics temperature is too high.• Dirty Air Filter - (located below Illuminator)
• Optics Fan (W29)
• Illuminator Controller PCB
• Table Top Illuminator Module
5202AdvisoryIlluminator optics fan is at full speed. Optics unit may be overheating.ILM_OPTICS_FAN_ALARM_ADV – Warning that the Optics fan is full on.• Dirty Air Filter - (located below Illuminator)
• Optics Fan (W29)
• Illuminator Controller PCB
• Table Top Illuminator Module
5203ErrorIlluminator optics thermo-cut-off has been triggered. Illuminator functions will be disabled.ILM_OPTICS_OVER_TEMP_ERR – Lamp is turned off because thermo cut-off.• Dirty Air Filter - (located below Illuminator)
• Optics Fan (W29)
• Optics Thermal Switch
• Illuminator Controller PCB
• Table Top Illuminator Module
5204AdvisoryCommunication failure with the Illuminator optics fan. The fan may not work properly.ILM_OPTICS_FAN_COMMS_ADV – Communication Error with optics fan.• Communication - Eject and reinsert
Illuminator
• Cable W81
• Illuminator Controller PCB
5300AdvisoryIlluminator ballast temperature is high. The lamp will be turned off if the temperature continues to rise.ILM_BALLAST_TEMP_HIGH_ADV – Ballast temperature is too high. Lamp is going to be shut down if temperature continues to rise.• Dirty Air Filter - (located below Illuminator)
• Ballast Fan (W28)
• Illuminator Controller PCB
• Table Top Illuminator Mod
5301ErrorIlluminator ballast temperature has exceeded its limit. Illuminator functions will be disabled.ILM_BALLAST_TEMP_DOWN_ERR – Illuminator is shut down because Ballast temperature is too high.• Dirty Air Filter - (located below Illuminator)
• Ballast Fan (W28)
• Illuminator Controller PCB
• Table Top Illuminator Module
5302AdvisoryIlluminator ballast fan is at full speed. Ballast unit may be overheating.ILM_BALLAST_FAN_ALARM_ADV – Warning that the Ballast fan is full on.• Dirty Air Filter - (located below Illuminator)
• Ballast Fan (W28)
• Illuminator Controller PCB
• Table Top Illuminator Module
5303ErrorIlluminator ballast thermo-cut-off has been triggered. Illuminator functions will be disabled.ILM_BALLAST_OVER_TEMP_ERR – Lamp is turned off because thermo cut-off.• Dirty Air Filter - (located below Illuminator)
• Ballast Fan (W28)
• Illuminator Controller PCB
• Table Top Illuminator Module
5304AdvisoryCommunication failure with the Illuminator ballast fan. The fan may not work properly.ILM_BALLAST_FAN_COMMS_ADV – Communication Error with ballast fan.• Communication - Eject and reinsert
Illuminator
• Cable W09
• Illuminator Controller PCB
• Interface (Floating) PCB
5400ErrorLamp louver failure: unable to move to home position. Left port will be disabled.ILM_STEP_MOTOR1_HOME_ERR – Step motor at port 1 failed to move to home position.• Home Sensor PCB.
• Home Sensor cable
• Port 1 Stepper Motor assy.
• Interface PCB
• Illuminator Controller PCB
• Table Top Illuminator Module
5401ErrorLamp louver failure: unable to move to home position. Right port will be disabled.ILM_STEP_MOTOR2_HOME_ERR – Step motor at port 2 failed to move to home position.• Home Sensor PCB.
• Home Sensor cable
• Port 1 Stepper Motor assy.
• Interface PCB
• Illuminator Controller PCB
• Table Top Illuminator Module
5402ErrorLamp louver failure: unable to move to specified position. Left port will be disabled.ILM_STEP_MOTOR1_MOVE_ERR – Step motor at port 1 failed to move to specified position.• Port 1 Stepper Motor Assy
• Illuminator Controller PCB
• Table Top Illuminator Module
5403ErrorLamp louver failure: unable to move to specified position. Right port will be disabled.ILM_STEP_MOTOR2_MOVE_ERR – Step motor at port 2 failed to move to specified position.• Port 2 Stepper Motor Assy
• Illuminator Controller PCB
• Table Top Illuminator Module
6101ErrorPressure reading is too high. Cutting and Pneumatics functions will be disabled.PNU_CMN_PRESSURE_TOO_HIGH_ERR – Pneumatics pressure is too high.• Air Distribution Assy
• Pneumatic Controller PCB
• Cable W47
• Pneumatic Module
6103ErrorPneumatics pressure transducers reference voltage is unacceptable. Pneumatics functions are disabled.PNU_CMN_ADC_2048_ERR – Pneumatics ADC 2048 mv. Reference voltage is beyond acceptable range.• Air Distribution Assy
• Pneumatic Controller PCB
• Cable W47
• Pneumatic Module
6104ErrorPneumatics Air Distribution transducers offset voltage is unacceptable. Pneumatics functions are disabled.PNU_CMN_AIR_ADC_OFFSET_ERR – Pneumatics Air Distribution PCB reference offset voltage is beyond acceptable range.• Air Distribution Assy
• Pneumatic Controller PCB
• Cable W47
• Pneumatic Module
6105ErrorPneumatics Main Manifold transducers offset voltage is unacceptable. Pneumatics functions are disabled.PNU_CMN_MAIN_ADC_OFFSET_ERR – Pneumatics Main Manifold PCB reference offset voltage is beyond acceptable range.• Air Distribution Assy
• Pneumatic Controller PCB
• Cable W47
• Pneumatic Module
6201ErrorAir Pressure valves have high transition faults. Cutting and Pneumatics functions will be disabled.PNU_AIR_VALVE_FAULT_HI_ERR - AirPressure valves have high level faults (fail to open).• Air Distribution Assy
• Pneumatic Controller PCB
• Cable W47
• Pneumatic Modul
6202ErrorAir Pressure valves have low transition faults. Cutting and Pneumatics functionals will be disabled.PNU_AIR_VALVE_FAULT_LO_ERR - AirPressure valves have low level faults (fail to close).• Air Distribution Assy
• Pneumatic Controller PCB
• Cable W47
• Pneumatic Mod
6203AdvisoryAir Pressure inlet filter may be dirty and needs to be replaced. Please contact Field Service.PNU_AIR_FILTER_DIRTY_ADV - The unusual high pressure drop cross the Air Filter. This is caused by either the Air Filter is too dirty and need to be replaced or the system may be leaking air.• Filter needs changing
• Air Distribution Assy
• Pneumatic Controller PCB
6204AdvisoryInlet pressure is too low; adjust to between 90-120 psi. System may have reduced performance between 58-90 psi..PNU_AIR_LOW_SWITCH_ON_ADV - The Air source (wall pressure) is too low (below 58 psi) to turn on. On startup, Pneumatics turns on the pressure automatically if the pressure is between 58 psi and 120 psi.• User Advisory - See Discrepancy description
• Air Distribution Assy
• Pneumatic Controller PCB
• Cable W47
• Pneumatic Module
6206AdvisoryInlet pressure is too high; adjust to between 90-120 psi.PNU_AIR_HIGH_SWITCH_ON_ADV - The Air source (wall pressure) is too high (above 120 psi) to turn on. On startup, Pneumatics turns on the pressure automatically if the pressure is between 58 psi and 120 psi.• User Advisory - See Discrepancy description
• Air Distribution Assy
• Pneumatic Controller PCB
• Cable W47
• Pneumatic Module
6209AdvisoryInlet pressure is below 90 psi; system may have reduced performance. Adjust to between 90-120 psi.PNU_AIR_PRESSURE_90PSI_ADV - The Air source pressure is below 90 psi. System performance will be diminished.• User Advisory - See Discrepancy description
6211AdvisoryThe source pressure transducer is out of specification. Please contact Field Service.PNU_AIR_SENSOR_CALIBRATED_ADV - The source pressure transducer is out of specification.• Air Distribution Assy
• Pneumatic Controller PCB
• Cable W47
• Pneumatic Module
6301ErrorCutting errors; valves have high transition faults (failed to open). Cutting functions will be disabled.PNU_CUT_VALVE_FAULT_HI_ERR - Cutters valves have high level faults (fail to open).• Cutter Circuit valves
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6302ErrorCutting errors; valves have low transition faults (failed to close). Cutting functions will be disabled.PNU_CUT_VALVE_FAULT_LO_ERR - Cutters valves have low level faults (fail to close).• Cutter Circuit valves
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6303ErrorCutting error: redundant transducers discrepancy error. Cutting functions will be disabled. Please contact Field Service.PNU_CUT_REDUNDANT_SENSORS_ERR – Cutters redundant transducers discrepancy error.• Cutter Circuit Transducers
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6304ErrorCutting error: redundant transducers calibration. Cutting functions will be disabled. Please contact Field Service.PNU_CUT_SENSOR_CALIBRATION_ERR – Cutters redundant transducers calibration error.• Cutter Circuit Transducers
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6305AdvisoryCutting pressure is oscillating beyond the specified range. Please release the footswitch treadle to reset.PNU_CUT_PRESSURE_OSCIL_ADV – Cutters actual pressure is oscillating beyond the specified tolerance.• Unstable pressure - possible leak
• Proportional Valve
• Cutter Circuit Transducers
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6306AdvisoryCutting pressure is surging beyond the specified tolerance. Please release the footswitch treadle to reset.PNU_CUT_PRESSURE_OSCIL_ADV – Cutters actual pressure is oscillating beyond the specified tolerance.• Unstable pressure - possible leak
• Proportional Valve
• Cutter Circuit Transducers
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6307AdvisoryVitrectomy port transducers are out of tolerance. The Vit Probe is still available. Please contact Field Service.PNU_CUT_VIT_SENSOR_FAULT_ADV – Vit. port transducers discrepancy error. SmartVit function is disabled.• Calibration needed
• Vit Transducers
• Vit Valve
• Pneumatic Controller PCB
• Pneumatic Module
6308AdvisoryPneumatics Flash is in need of Service. The Vit Probe is still available. Please contact Field Service.PNU_CUT_VIT_FLASH_FAULT_ADV – The Pneumatics Flash memory is in need of Service.• Pneumatic Controller PCB
• Pneumatic Module
6401ErrorUtilities valves have high transition faults (failed to open). Pneumatics functions will be disabled.PNU_UTL_VALVE_FAULT_HI_ERR - Utilities valves have high level faults (fail to open).• Utility Circuit valves
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6402ErrorUtilities valves have low transition faults (failed to close). Pneumatics functions will be disabled.PNU_UTL_VALVE_FAULT_LO_ERR - Utilities valves have low level faults (fail to close).• Utility Circuit valves
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6403ErrorUtilities redundant transducers discrepancy error. Pneumatics functions will be disabled. Please contact Field Service.PNU_UTL_REDUNDANT_SENSORS_ERR – Utilities redundant transducers discrepancy error.• Utilities Circuit Transducers
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6404ErrorUtilities redundant transducers calibration error. Pneumatics functions will be disabled. Please contact Field Service.PNU_UTL_SENSOR_CALIBRATION_ERR – Utilities redundant transducers calibration error.• Utilities Circuit Transducers
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6405ErrorVacuum redundant transducers discrepancy error. Pneumatics functions will be disabled. Please contact Field Service.PNU_VAC_REDUNDANT_SENSORS_ERR – Vacuum redundant transducers discrepancy error.• VFC Vacuum Transducers
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6406ErrorVacuum redundant transducers calibration error. Pneumatics functions will be disabled. Please contact Field Service.PNU_VAC_SENSOR_CALIBRATION_ERR – Vacuum redundant transducers calibration error.• VFC Vacuum Transducers
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6407ErrorPressure is unstable. Release the footswitch treadle or restart AGF.PNU_UTL_PRESSURE_OSCIL_ADV – Utilities actual pressure is oscillating beyond the specified tolerance.• Unstable pressure - possible leak
• Proportional Valve
• Cutter Circuit Transducers
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6408AdvisoryUtilities pressure is surging beyond the specified tolerance. Please release the footswitch treadle to reset.PNU_UTL_PRESSURE_SURGE_ADV – Utilities actual pressure is surging beyond the specified tolerance.• Unstable pressure - possible leak
• Proportional Valve
• Cutter Circuit Transducers
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6501ErrorAuto Gas valves have high transition faults (failed to open). Pneumatics functions will be disabled.PNU_AGF_VALVE_FAULT_HI_ERR - Auto Gas valves have high level faults (fail to open).• AGF valves
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6502ErrorAuto Gas valves have low transition faults (failed to close). Pneumatics functions will be disabled.PNU_AGF_VALVE_FAULT_LO_ERR - Auto Gas valves have low level faults (fail to close).• AGF valves
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6503ErrorAuto Gas redundant transducers discrepancy error. Pneumatics functions will be disabled. Please contact Field Service.PNU_AGF_REDUNDANT_SENSORS_ERR – Auto Gas redundant transducers discrepancy error.• AGF Transducers
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6504ErrorAuto Gas redundant transducers calibration error. Pneumatics functions will be disabled. Please remove AGF syringe if connected.PNU_AGF_SENSOR_CALIBRATION_ERR – Auto Gas redundant transducers calibration error.• AGF Transducers
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6505AdvisoryC3F8 bottle may be empty and needs to be replaced. Please press [Replaced] to confirm bottle replacement.PNU_AGF_GAS_1_EMPTY_ADV – Gas 1 bottle may be empty and needs to be filled up.• C3F8 Valve
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6506AdvisorySF6 bottle may be empty and needs to be replaced. Please press [Replaced] to confirm bottle replacement.PNU_AGF_GAS_2_EMPTY_ADV - Gas 2 bottle may be empty and needs to be filled up.• C3F8 Valve
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
6507AdvisoryUnable to run Auto Gas Filling without sufficient source pressure.PNU_AGF_NO_AIR_PRESSURE_ADV – There is not enough air pressure to support Auto Gas Filling functions.• Check Rear panel connection
• AGF Transducers
• Transducer Interface PCB
• Cable W122
• Pneumatic Controller PCB
• Pneumatic Module
7100ErrorBallast failure (voltage). Illuminator functions will be disabled.ILM_VOLTAGE_SENSOR_ERR - Ballast transducer has reading that is out of specified safety range.• Cable W25
• Illuminator Controller PCB
• Auxiliary Illuminator Module
7101ErrorBallast failure (current). Illuminator functions will be disabled.ILM_CURRENT_SENSOR_ERR – Ballast transducer has reading that is out of specified safety range.• Cable W25
• Illuminator Controller PCB
• Auxiliary Illuminator Module
7102AdvisoryLamp calibration data is corrupted: lamp needs to be calibrated. Please contact Field ServiceILM_CALIBRATION_DATA_ADV – The calibration data saved in the flash contains incorrect CRC checksum.• Calibration Data corrupt
7103AdvisoryFailure to turn lamp on: lamp needs to be replaced. Please contact Field Service.ILM_LAMP_SOURCE_FAULT_ADV – Lamp status reported by hardware is not the same as software status.• Lamp
• Illuminator Controller PCB
• Auxillary Illuminator Module
7104AdvisoryThe lamp has exceeded its rated life: lamp needs to be replaced. Please contact Field Service.ILM_LAMP_REPLACEMENT_ADV – The lamp has exceeded its rated life. The new lamp is expected.• Lamp
7105AdvisoryThe lamp has exceeded its rated maximum life: lamp needs to be replaced. Please contact Field Service.ILM_LAMP_INTENSITY_LOW_ADV – The lamp has exceeded its rated safe life. The new lamp should be installed immediately.• Lamp
7113AdvisoryMetrics data is corrupted; Lamp hour is invalid. Please contact Field Service.ILM_METRICS_CORRUPT_ADV – The metrics data for lamp hours is invalid. Either the Flash is corrupted or the checksum is incorrect.• Illuminator Controller PCB
7200AdvisoryIlluminator optics temperature is high. The lamp will be turned off if the temperature continues to rise.ILM_OPTICS_TEMP_HIGH_ADV – Optics temperature is too high. Lamp is going to be shut down if temperature continues to rise.• Dirty air filter. (located below the
Illuminator).
• Optics fan W20.
• Illuminator Controller PCB
• Auxillary Illuminator Module
7201ErrorIlluminator optics temperature has exceeded its limit. Illuminator functions will be disabled.ILM_OPTICS_TEMP_DOWN_ERR – Illuminator is shut down because Optics temperature is too high.• Dirty air filter. (located below the Illuminator).
• Optics fan W20.
• Illuminator Controller PCB
• Auxillary Illuminator Module
7202AdvisoryIlluminator optics fan is at full speed. Optics unit may be overheating.ILM_OPTICS_FAN_ALARM_ADV – Warning that the Optics fan is full on.• Dirty air filter. (located below the Illuminator).
• Optics fan W20.
• Illuminator Controller PCB
• Auxillary Illuminator Module
7203ErrorIlluminator optics thermo-cut-off has been triggered. Illuminator functions will be disabled.ILM_OPTICS_OVER_TEMP_ERR – Lamp is turned off because thermo cut-off.• Dirty air filter. (located below the Illuminator).
• Optics fan W20.
• Optics Thermal Switch
• Illuminator Controller PCB
• Auxillary Illuminator Module
7204AdvisoryCommunication failure with the Illuminator optics fan. The fan may not work properly.ILM_OPTICS_FAN_COMMS_ADV – Communication Error with optics fan.• Communication - Eject and reinsert Illuminator
• Cable W11
• Illuminator Controller PCB
• Auxillary Illuminator Module
7300AdvisoryIlluminator ballast temperature is high. The lamp will be turned off if the temperature continues to rise.ILM_BALLAST_TEMP_HIGH_ADV – Ballast temperature is too high. Lamp is going to be shut down if temperature continues to rise.• Dirty air filter. (located below the Illuminator).
• Ballast Fan (W33)
• Illuminator Controller PCB
• Auxillary Illuminator Module
7301ErrorIlluminator ballast temperature has exceeded its limit. Illuminator functions will be disabled.ILM_BALLAST_TEMP_DOWN_ERR – Illuminator is shut down because Ballast temperature is too high.• Dirty air filter. (located below the Illuminator).
• Ballast Fan (W33)
• Illuminator Controller PCB
• Auxillary Illuminator Module
7302AdvisoryIlluminator ballast fan is at full speed. Ballast unit may be overheating.ILM_BALLAST_FAN_ALARM_ADV – Warning that the Ballast fan is full on.• Dirty air filter. (located below the
Illuminator).
• Ballast Fan (W33)
• Illuminator Controller PCB
• Auxillary Illuminator Module
7303ErrorIlluminator ballast thermo-cut-off has been triggered. Illuminator functions will be disabled.ILM_BALLAST_OVER_TEMP_ERR – Lamp is turned off because thermo cut-off.• Dirty air filter. (located below the
Illuminator).
• Ballast Fan (W)
• Illuminator Controller PCB
• Auxillary Illuminator Module
7304AdvisoryCommunication failure with the Illuminator ballast fan. The fan may not work properly.ILM_BALLAST_FAN_COMMS_ADV – Communication Error with ballast fan.• Communication - Eject and reinsert
Illuminator
• cable W11
• Illuminator Controller PCB
• Auxillary Illuminator Module
7400ErrorLamp louver failure: unable to move to home position. Left port will be disabled.ILM_STEP_MOTOR3_HOME_ERR – Step motor at port 3 failed to move to home position.• Home Sensor PCB.
• Home Sensor cable
• Port 3 Stepper Motor assy.
• Interface PCB
• Illuminator Controller PCB
• Auxiliary Illuminator Module
7401ErrorLamp louver failure: unable to move to home position. Right port will be disabled.ILM_STEP_MOTOR4_HOME_ERR – Step motor at port 4 failed to move to home position.• Home Sensor PCB.
• Home Sensor cable
• Port 4 Stepper Motor assy.
• Interface PCB
• Illuminator Controller PCB
• Auxiliary Illuminator Module
7402ErrorLamp louver failure: unable to move to specified position. Left port will be disabled.ILM_STEP_MOTOR3_MOVE_ERR – Step motor at port 3 failed to move to specified position.• Port Stepper Motor assy.
• Interface PCB
• Illuminator Controller PCB
• Auxiliary Illuminator Module
7403ErrorLamp louver failure: unable to move to specified position. Right port will be disabled.ILM_STEP_MOTOR4_MOVE_ERR – Step motor at port 4 failed to move to specified position.• Port 3 Stepper Motor assy.
• Interface PCB
• Illuminator Controller PCB
• Auxiliary Illuminator Module
8100ErrorLaser controller software error. Laser functions will be disabled.LSR_SOFTWARE_ERR - bad process communication•   Reload system software and recalibrate
8101ErrorLaser controller shutter open for too long error. Laser functions will be disabled.LSR_SHUTR_OPN_TOO_LONG_ERR -
laser beam not properly detected at port
• Photocell Alignment
• Top Sensor PCB
8102ErrorLaser controller shutter unexpected error. Laser functions will be disabled.LSR_SHUTR_UNEXPECTED_ERR - shutter open or position sense problem (at startup)• Shutter position or sense
8103ErrorLaser controller shutter open between firing error. Laser functions will be disabled.LSR_SHUTR_OPN_BTWN_FIRING_ERR - shutter closed or position sense problem (at startup)• Shutter jammed mechanically
• Top Sensor PCB
8104ErrorLaser controller mirror incorrect position error.
Laser functions will be disabled.
LSR_MIRROR_INCORRECT_POS_ERR - Sensor reporting both or no Port Mirror position (at startup)• Port Mirror Switch
• Sensor
• Top Sensor PCB
• Bottom Sensor PCB
8105ErrorLaser controller mirror incorrect position error.
Laser functions will be disabled.
LSR_MIRROR_POS_UNEXPECTED_ERR - Sensor reporting both or no Port Mirror position (at startup)• Port Mirror Switch
• Sensor
• Top Sensor PCB
• Bottom Sensor PCB
8106ErrorLaser controller power over the limit error. Laser
functions will be disabled.
LSR_PWR_OVER_LIMIT_ERR - laser power measure flucuates >20%• PMON calibration
• Power Driver PCB
8107AdvisoryLaser controller maximum current exceeded.LSR_MAX_CURRENT_EXCEEDED_ADV - laser diode current exceeds max limit -• Temperature Calibration
• Laser head;
8108ErrorLaser controller port power detection error. Laser
functions will be disabled.
LSR_PORT_PWR_DETECTION_ERR - laser power incorrectly detected (during normal run time)• Photocell beam alignment
• Photocells
• Top Sensor PCB
8109ErrorLaser controller power detected at wrong port error.
Laser functions will be disabled.
LSR_PWR_AT_WRONG_PORT_ERR - laser detected at wrong port, or not detected at selected port (during normal run time)• Port Mirror Switch jammed mechanically;
• Sensor
• Top Sensor PCB
• Bottom Sensor PCB
8110ErrorLaser controller power reading mismatch error.
Laser functions will be disabled.
LSR_PWR_READING_MISMATCH_ERR - pmons disagree by >20%• PMONs calibration
• Top Sensor PCB;
8111ErrorLaser controller LBO crystal limit error. Laser
functions will be disabled.
LSR_LBO_CRYSTAL_LIMIT_ERR - LBO temp out of limit• LBO temp calibration
• Mid-Extension PCB;
8112ErrorLaser controller pump temperature limit error.
Laser functions will be disabled.
LSR_PUMP_TEMP_LIMIT_ERR - laser diode temp out of limit• Temperature related - Allow unit to reach
normal room temp
• Mid-Extension PCB;
8113ErrorLaser controller power detect digital I/O mismatch
error. Laser functions will be disabled.
LSR_PWR_DETECT_DIO_MISMATCH_ERR - laser photo detect digital input problem• Controller PCB
8114ErrorLaser controller footswitch digital I/O mismatch
error. Laser functions will be disabled.
LSR_FTSW_DIO_MISMATCH_ERR - footswitch digital inputs problem• Controller PCB
8115ErrorLaser controller footswitch no NC (normally closed) error. Laser functions will be disabled.LSR_FTSW_NO_NC_ERR - footswitch inputs show both NO and NC active• Footswitch
• Back Panel PCB
8116ErrorLaser controller system data CRC errorLC_SYSTEM_DATA_CRC_ERR - calculated CRC incorrect (for one of the .dat data files)• Temperatures, PMONs, term efficiencies, etc. Calibration,
8117AdvisoryLaser controller internal parameter error.LSR_INTERNAL_PARAMETER_ADV - Bad parameter call• Wipe memory, reload all software and recalibrate
8118ErrorLaser controller process hunt error. Laser functions will be disabled.LSR_PROCESS_HUNT_ERR - startup process issue• Wipe memory, reload all software and recalibrate
8119ErrorLaser controller process attach error. Laser functions will be disabled.LSR_PROCESS_ATTACH_ERR - startup process issue• Wipe memory, reload all software and recalibrate
8120ErrorLaser controller process attach error. Laser functions will be disabled.LSR_START_UP_TIME_OUT_ERR - state machine error• Wipe memory, reload all software and recalibrate
8121ErrorLaser controller process attach error. Laser functions will be disabled.LSR_JEM_CRC_ERR - crc for JEM incorrect• Wipe memory, reload all software and recalibrate
8122ErrorLaser controller process attach error. Laser functions will be disabled.LSR_KERNEL_CRC_ERR - crc for Kernel incorrect• Wipe memory, reload all software and recalibrate
8123ErrorLaser controller process attach error. Laser functions will be disabled.LSR_FLASH_FILE_ERR - crc for Laser App incorrect• Wipe memory, reload all software and recalibrate
8124AdvisoryLaser LBO diode over temperature errorLSR_PUMP_OVERTEMP_ADV - Advisory Only for temperature setpoints• Recalibrate temperatures
8125AdvisoryLaser LBO diode under temperature error.LSR_PUMP_UNDERTEMP_ADV - Advisory Only for temperature setpoints• Recalibrate temperatures
8126AdvisoryLaser controller crystal over-temperature error.LSR_CRYSTAL_OVERTEMP_ADV - Advisory Only for temperature setpoints• Recalibrate temperatures
8127AdvisoryLaser controller crystal under-temperature error.LSR_CRYSTAL_UNDERTEMP_ADV - Advisory Only for temperature setpoints• Recalibrate temperatures
8129AdvisoryLaser controller probe connection error.LSR_PPP_PROBE_CONNECTION_ADV - Probe Detection; mis-match between sensor pairs• Sensors or sensor position
• Probe Detect PCB
8131ErrorLaser controller power monitor POST error. Laser functions will be disabled.LSR_PMON_POST_FAIL_ERR - pmons fail POST• PMON calibration
• Top Sensor PCB;
8132ErrorLaser controller +2.5 volt out-of-range error. Laser functions will be disabled.LSR_PLUS_2_5_VOLT_ERR - 2.5V on Laser Controller out of range• Controller PCB
8134ErrorLaser controller 12 volt power error. Laser functions will be disabled.LSR_12_VOLT_POWER_ERR - 12V on Power Driver PCB out of range• Power Driver PCB
8135ErrorLaser shutter timer expired error. Laser functions will be disabled.LSR_SHUTTER_TIMER_OFF_ERR - shutter timer timed out• Exposure time calibration
• shutter is jammed mechanically
• shutter sensor
• Top Sensor PCB
8138ErrorLaser controller diode thermal electric cooler error. Laser functions will be disabled.LSR_DIODE_TEC_FAULT_ERR - Main TEC (thermo-electric-cooler) problem• TEC
• Power Driver PCB
8139ErrorLaser probe port process POST error. Laser functions will be disabled.LSR_PPP_POST_FAULT_ERR - unable to switch port mirror or detect position• Port mirror jammed mechanically
• Top Sensor PCB;
8140ErrorLaser controller startup timeout error (laser engine). Laser functions will be disabled.LC_LE_STARTUP_TIMEOUT_ERR - state machine error• Wipe memory, reload all software and recalibrate
8141ErrorLaser controller startup timeout error (shutter control process). Laser functions will be disabled.LC_SCP_STARTUP_TIMEOUT_ERR - state machine error• Wipe memory, reload all software and recalibrate
8142ErrorLaser controller startup timeout error (supervisor process). Laser functions will be disabled.LC_SUP_STARTUP_TIMEOUT_ERR - state machine error• Wipe memory, reload all software and recalibrate
8143ErrorLaser controller startup timeout error (peripheral management process). Laser functions will be disabled.LC_PMP_STARTUP_TIMEOUT_ERR - state machine error• Wipe memory, reload all software and recalibrate
8144ErrorLaser controller startup timeout error (probe port process). Laser functions will be disabled.LC_PPP_STARTUP_TIMEOUT_ERR - state machine error• Wipe memory, reload all software and recalibrate
8145ErrorLaser controller startup timeout error (laser system controller). Laser functions will be disabled.LC_LSC_STARTUP_TIMEOUT_ERR -• Wipe memory, reload all software and recalibrate
8147ErrorLaser controller TMP crystal startup timeout error. Laser functions will be disabled.LC_TMP_CRYSTAL_POST_ERR - LBO temp not stable (on startup)• Recalibrate temperatures
8148ErrorLaser controller TMP diode startup timeout error. Laser functions will be disabled.LC_TMP_DIODE_POST_ERR - laser diode temp not stable (on startup)• Recalibrate temperatures
8149ErrorLaser controller Dr Filter error. Laser functions will be disabled.LC_DRFILTER_ERR - reads both NC and NO contacts active• Dr filter
• Rear Panel PCB
8150ErrorLaser controller invalid maximum power error. Laser functions will be disabled.LC_INVALID_MAX_POWER_ERR - Invalid power request• Wipe memory, reload all software and recalibrate
8153ErrorLaser controller load module CRC error. Laser functions will be disabled.LC_LOADMODULE_CRC_ERR - Invalid software load• Wipe memory, reload all software and recalibrate
8154ErrorLaser controller module requires calibration. Laser functions will be disabled.LC_CALIBRATION_REQUIRED_ERR - A valid calibration flag is missing• Calibration
8155ErrorLaser Hardware and Software version incompatibility error. Laser functions will be disabled.LC_VERSION_INCOMPATIBLE_ERR - Version Incompatibility• Check all Software and Hardware Revision levels,
• Wipe memory, reload all software and recalibrate as needed
8156ErrorLaser controller power monitor saturation error. Laser functions will be disabled.LC_PMON_SATURATION_ERR - PMON Voltage problem• PMON calibration
• PMON assembly
• Top Sensor PCB
8200AdvisoryLaser controller probe removed while firing.LSR_PRBE_REMOVE_WHEN_FIRING_ADV - Laser controller probe removed while firing.User Advisory - See Discrepancy description
8202AdvisoryLaser controller Dr. Filter disengaged while firing.LSR_FLTR_DISENG_WHEN_FIRING_ADV - Laser controller Dr. Filter disengaged while firing.User Advisory - See Discrepancy description
8203AdvisoryLaser controller Dr. Filter disconnected while firing.LSR_FLTR_DISCON_WHEN_FIRING_ADV - Laser controller Dr. Filter disconnected while firing.User Advisory - See Discrepancy description
8204AdvisoryLaser controller interlock opened while firing.LSR_INTRLK_OPN_WHEN_FIRING_ADV - Laser controller interlock opened while firing.User Advisory - See Discrepancy description
8205AdvisoryLaser controller footswitch disconnected while firing.LSR_FTSW_DISCON_WHEN_FIRING_ADV - Laser controller footswitch disconnected while firing.User Advisory - See Discrepancy description
8207AdvisoryLaser controller port changed in Ready State.LSR_PORT_CHANGE_IN_RDYSTE_ADV - Laser controller port changed in Ready State.User Advisory - See Discrepancy description
8208AdvisoryLaser controller Dr. Filter disengaged in Ready State.LSR_FLTR_DISENG_IN_RDYSTE_ADV - Laser controller Dr. Filter disengaged in Ready State.User Advisory - See Discrepancy description
8209AdvisoryLaser controller Dr. Filter disconnected in Ready State.LSR_FLTR_DISCON_IN_RDYSTE_ADV - Laser controller Dr. Filter disconnected in Ready State.User Advisory - See Discrepancy description
8210AdvisoryLaser controller interlock opened in Ready State.LSR_INTRLK_OPN_IN_RDYSTE_ADV - Laser controller interlock opened in Ready State.User Advisory - See Discrepancy description
8211AdvisoryLaser controller footswitch removed in Ready State.LSR_FTSW_REMOVE_IN_RDYSTE_ADV - Laser controller footswitch removed in Ready State.User Advisory - See Discrepancy description
8212AdvisoryLaser controller footswitch engaged when Ready requested.LSR_FTSW_ENGAGE_WHEN_RDYREQ_ADV - Laser controller footswitch engaged when Ready requested.User Advisory - See Discrepancy description
8213AdvisoryLaser controller user data CRC error.LSR_USER_DATA_CRC_ERROR_ADV - Laser controller user data CRC error.User Advisory - See Discrepancy description
8214AdvisoryLaser controller Ready State denied: the filter is disengaged.LcRdyDeniedFilterDisAdv - Laser controller Ready State denied: the filter is disengaged.User Advisory - See Discrepancy description
8215AdvisoryLaser controller Ready State denied: there is no interlock.LcRdyDeniedNoInterlockAdv - Laser controller Ready State denied: there is no interlockUser Advisory - See Discrepancy description
8216AdvisoryLaser controller Ready State denied: there is no valid probe for the active port.LcRdyDeniedNoPortSelAdv - Laser controller Ready State denied: there is no valid probe for the active port.User Advisory - See Discrepancy description
8217AdvisoryLaser controller Ready State denied: the footswitch is depressed.LcRdyDeniedFtswPressAdv - Laser controller Ready State denied: the footswitch is depressed.User Advisory - See Discrepancy description
8218AdvisoryAre all necessary Dr. Filters properly installed and connected?LcRdyDeniedFltrVerAdv - User requests going to Ready Mode when the doctor filters haven’t been verified.User Advisory - See Discrepancy description
8219AdvisoryLaser controller Ready State denied: there is no footswitch present.LC_RDY_DENIED_NO_FTSW_ADV - User presses the Laser Ready button when no footswitch is connected.User Advisory - See Discrepancy description
8220AdvisoryLaser controller Dr. Filter connected while in Transition, Ready or Firing State.LC_DRFILTER_CON_TRF_ADV - User connects an engaged Dr. Filter while in Transition / Ready or Firing mode and and Endo probe is connected to the active port.User Advisory - See Discrepancy description
8221AdvisoryLaser controller detected a port Hardware failure.LC_PORT_HW_FAILURE_ADV - Laser controller detected a port Hardware failure.User Advisory - See Discrepancy description
8222AdvisoryPort can't be selected: the probe type for the port isn't valid.LC_PORT_DENIED_INVLD_PROBE_ADV - The user tries to make a laser port the active port, using one of the Purepoint console port selection buttons, but the currently selected probe type for that port is invalid.User Advisory - See Discrepancy description

MEDiSURG knows the Alcon Infiniti Vision System is being phased out. But Infiniti is not finished yet!

In the world of ophthalmology, procedures are always evolving. Having a group of consultants and specialists on hand can be a huge benefit when it comes time to make adjustments to your units, surgical procedures, or find new equipment.The Alcon Infiniti Vision...

MEDiSURG is offering a 50% discount on the Alcon Centurion Preventative Maintenance Online Course until the end of 2023

Numerous MEDiSURG alumni students and newcomers to the biotech industry have benefited from our best-selling Preventative Maintenance (PM) for the Alcon Centurion Vision System online course. MEDiSURG is currently providing this course at a 50% discount through the...