Constellation
Alumni Page
THE INDUSTRY’S LEADING, INDEPENDENT EXPERTS IN EYE SURGICAL EQUIPMENT.
Training Alumni
Please email or call with any part orders.
Do not hesitate to call or email Medisurg with any tech support!
317-663-3158
medisurg@medisurg.com
“Great course, great instructor. The staff was great to talk to, ask questions to, and very helpful”
Merrell Williams
Aramark
"The information was taught in a very easy to understand way with lots of hands on and with all of the tools and manual to give me full confidence in servicing the Infinity system."
Cory Ladd
UCSF
"Very fun, hands on, and involved course. Knowledgeable staff that are very kind and thoughtful."
Jarrod Huffer
Trimedx
How we can help you?
Lorem ipsum dolor sit amet, consectetuer adipiscing elit.
855.233.4050
Original Part # | |
---|---|
212-0000-XXX | |
212-1083-XXX | Assy, Footpedal, Six Switch |
212-5002-XXX | Assy, Table Top Console |
212-3398-XXX | Assy, Display, LCD 17 |
212-1010-XXX | Assy, Module, Host |
212-1016-XXX | Assy, Power Controller, Tabletop |
212-1023-XXX | Assy, Module, Pneumatic AK 6.103 |
212-1037-XXX | Assy, Module, U/S-Diat Aqua L 6.124 |
212-3564-XXX | Assy, Module, US/Diat IOL |
212-1040-XXX | Assy, Filter, Oil/Air Dryer |
212-1758-501 | Assy, Ballast, Illum Xenon |
212-3245-XXX | Assy, Fluidics |
212-3334-552 | ASSY,PCB,SUPERVISOR MODULE |
212-2981-XXX | Assy, Gas AGF, External |
212-5003-XXX | Assy, Base, Premium |
212-1007-XXX | Assy, Laser |
562-1034-XXX | Assy, Core Module |
562-1564-XXX | Assy, Core Module |
562-1041-XXX | Assy, Engine, Laser |
212-1008-XXX | Assy, Illuminator, Aux |
212-2150-501 | Assy, Ballast, Aux Illuminator |
212-2164-XXX | Assy, Power Dist, Cart |
212-1004-XXX | Assy, Arm, Tray |
212-1191-XXX | Assy, Arm, Lower |
212-1158-XXX | Assy, Tray |
212-1186-XXX | Assy, Arm, Upper |
562-1564-501 | Assy, Module, Core |
562-1510-001 | Assy, Footpedal, Laser |
Constellation CR3
Event Code | Event Type | Message to User | Detail / Discrepancy | ||
---|---|---|---|---|---|
1001 | Fault | Fault - 1001 Call Field Service | The OSE Gateway exited unex- pectedly. | ||
1002 | Fault | Fault - 1002 Call Field Service | The OSE Gateway has reported that the Supervisor process has gone down. | ||
1003 | Fault | Fault - 1003 Call Field Service | Something has failed in the communication between the Host and the Supervisor. | ||
1004 | Fault | Fault - 1004 Call Field Service | The Main Application exited un- expectedly. | ||
1005 | Fault | Fault - 1005 Call Field Service | The Host Controller checksum validation failed to pass all files under its control. | ||
1006 | Fault | Fault - 1006 Call Field Service | The Host has failed to connect to the Supervisor process. | ||
1008 | Fault | Fault - 1008 Call Field Service | The Host Controller failed to launch an executable image during startup. | ||
1009 | Fault | Fault - 1009 Call Field Service | The Host Controller could not communicate with the Power Module over the serial link. | ||
1010 | Fault | Fault - 1010 Call Field Service | Component or Submodule re- ported an incorrect software, hardware, firmware or other. | ||
1011 | Fault | Fault - 1011 Call Field Service | Host Controller communications was lost as the result of a .NET IPC failure. | ||
1012 | Fault | Fault - 1012 Call Field Service | The Host has not received any heartbeats from the Supervisor within the required timeframe. | ||
1013 | Fault | Fault - 1013 Call Field Service | The Host has detected that the User Interface thread has not responded within the required timeframe (i.e. the UI is consid- ered locked up). | ||
1014 | Fault | Fault - 1014 Call Field Service | The system has detected a soft- ware error. | ||
1015 | Fault | Fault - 1015 Call Field Service | The Host has received a mes- sage from the Supervisor that it doesn’t recognize (this is a spe- cific type of software error). | ||
1016 | Fault | Fault - 1016 Call Field Service | The Host application’s display fonts can not be loaded. | ||
1017 | Fault | Fault - 1017 Call Field Service | A failure has occurred during the playback of an audio file. | ||
1103 | Error | Voice Activation not available. | The user tried activate voice ac- tivation, but the voice activation process could not be found. | ||
1107 | Error | Unable to load one or more language packs. | An error occurred while at- tempting to load a non-English language. | ||
1108 | Error | Incompatible version numbers within the Fluidics submodule. Fluidics functions will be dis- abled. | The Fluidics module was shut down due to an incompatible version. The log file shows ex- actly what version information was incompatible. | ||
1109 | Error | Incompatible version numbers within the Pneumatics sub- module. Pneumatics functions will be disabled. | The Pneumatics module was shut down due to an incompat- ible version. The log file shows exactly what version information was incompatible. | ||
1110 | Error | Incompatible version numbers within the Ultrasound submod- ule. Ultrasound and Diathermy functions will be disabled. | The Ultrasound module was shut down due to an incompat- ible version. The log file shows exactly what version information was incompatible. | ||
1111 | Error | Incompatible version numbers within the Auxiliary Illuminator submodule. Auxiliary Illumina- tor functions will be disabled. | The Auxiliary Illuminator mod- ule was shut down due to an incompatible version. The log file shows exactly what version information was incompatible. | ||
1112 | Error | Incompatible version numbers within the Laser submodule. Laser functions will be dis- abled. | The Laser module was shut down due to an incompatible version. The log file shows ex- actly what version information was incompatible. | ||
1200 | Advisory | Function is not allowed when the fluidics subsystem is not functional. | The user tried to invoke a com- mand which is only allowed when the fluidics subsystem is functional. | ||
1202 | Advisory | Function is not allowed when the Laser is in Ready Mode. | The user tried to invoke a com- mand which is not allowed when the laser is in Ready Mode or Firing. | ||
1203 | Advisory | Function is not allowed when the footswitch treadle is down or buttons are pressed. | The user tried to invoke a com- mand which is only allowed when the treadle is up and no buttons pressed. | ||
1204 | Advisory | Flow mode is not available un- til the probe or handpiece has been primed. | Flow mode is unavailable be- cause a probe or handpiece hasn’t been primed. | ||
1205 | Advisory | Flow mode is only avail- able for 20 gauge probes / handpieces. | Flow mode is unavailable be- cause a non 20 gauge probe or handpiece is used. | ||
1206 | Advisory | Please prime the Vit probe. | The user presses the treadle in a Vit Step when the Vit Probe hasn’t been primed. | ||
1207 | Advisory | Please connect a Phaco hand- piece. | The user presses the treadle in a Phaco Step when no Phaco handpiece is connected. | ||
1208 | Advisory | Please tune the Phaco hand- piece. | The user presses the treadle in a Phaco Step when the Phaco handpiece isn’t tuned. | ||
1209 | Advisory | Please connect a Frag hand- piece. | The user presses the treadle in a Frag Step when no Frag handpiece is connected. | ||
1210 | Advisory | Please tune the Frag hand- piece. | The user presses the treadle in a Phaco Step when the Phaco handpiece isn’t tuned. | ||
1211 | Advisory | Function is not allowed when the Ultrasound submodule is not functional. | The user tried to invoke a com- mand which is only allowed when the ultrasound subsystem is functional. | ||
1212 | Advisory | Function is not allowed when the cassette is not ready. | The user tried to invoke a com- mand which is only allowed when the cassette is ready. | ||
1213 | Advisory | Infusion must be on to use VFC Extract. Press [Ignore] to allow VFC. Extract case. without infusion for this case | The user presses the treadle in a VFC Extract Step when infu- sion is off. | ||
1214 | Advisory | Pleas connect a VFC syringe. | The user presses the treadle in a VFC Step when no VFC Sy- ringe is connected. | ||
1215 | Advisory | Please connect Forceps | The user presses the treadle in a Forceps Step when no forceps are connected. | ||
1216 | Advisory | Pleas connect Scissors | The user presses the treadle in a Scissors Step when no Scis- sors are connected. | ||
1217 | Advisory | Pleas connect an AGF syringe | The user presses the “Start” but- ton in the Auto Gas Filling dialog when no AGF syringe has been connected. | ||
1218 | Advisory | Forceps not available during AGF: please try again when AGF is complete. | The user presses the treadle in a Forceps Step while Auto Gas Filling is in progress. | ||
1219 | Advisory | VFC not available during AGF: please try again when AGF is complete. | The user presses the treadle in a VFC Step while Auto Gas Fill- ing is in progress. | ||
1220 | Advisory | Ejecting the cassette is not allowed while the footswitch treadle is pressed. | The user attempts to eject the cassette while the treadle is down. | ||
1221 | Advisory | Ejecting the cassette is not allowed while infusion is on. | The user attempts to eject the cassette while infusion is on. | ||
1222 | Advisory | Ejecting the cassette is not allowed while irrigation is on. | The user attempts to eject the cassette while irrigation is on. | ||
1223 | Advisory | Ejecting the cassette is not allowed while priming, tuning, or testing. | The user attempts to eject the cassette while tests (priming/ tuning/testing) are in progress. | ||
1224 | Advisory | Cleaning the cassette is not allowed while infusion, irrigation, or FAX is on. | The user attempts to start Cas- sette Cleaning when either Infu- sion, Irrigation or FAX is on. | ||
1225 | Advisory | Cleaning the cassette is not allowed without a functional cassette. | The user attempts to start Cas- sette Cleaning when the Cas- sette isn’t available (e.g. not inserted). | ||
1226 | Advisory | Priming the cassette is not allowed without a functional cassette. | User attempt to prime the cas- sette when preconditions aren’t met (e.g. no cassette inserted, cassette not tested or didn’t pass the test). | ||
1227 | Advisory | An error occurred loading the device settings. The system will revert to default values. | A database attempt to load the device settings from the file system has failed. Reverting to defaults. | ||
1228 | Advisory | Command is not allowed while infusion is on. | The user tried to invoke a com- mand which is only allowed when infusion is not on. | ||
1229 | Advisory | Command is not allowed while irrigation is on. | The user tried to invoke a com- mand which is only allowed when irrigation is not on. | ||
1230 | Advisory | The selected step is not sup- ported with the current cas- sette type. | The user attempts to select a Step whose mode/submode isn’t compatible with the current operating mode. | ||
1231 | Advisory | The connected probe is not supported with the current cassette type. | The user connects a probe (with RFID) that’s not compatible with the current operating mode (e.g. an Ultra Vit probe when an Anterior cassette is connected or Ultra Vit Anterior probe when a Posterior cassette is con- nected). | ||
1232 | Advisory | Please connect footswitch. | The Supervisor has reported that no footswitch is connected. | ||
1233 | Advisory | The connected footswitch is not supported. | The Supervisor has detected the connection of an unknown footswitch. | ||
1234 | Advisory | Flow limit is currently not avail- able due to flow mode being unavailabe. | Vacuum mode flow limit de- pends on the availability of flow mode. If flow mode becomes unavailable then the flow limit function can not be performed by the fluidics submodule. | ||
1235 | Advisory | IV Pole pressure mode is cur- rently not available. | User attempts to select IV Pole as the infusion source but the IV Pole is not available (not config- ured, not working or Cassette is of Premium type) | ||
1236 | Advisory | Gravity pressure mode is cur- rently not available. | User attempts to select Gravity as the infusion source when a Premium Cassette is inserted. | ||
1237 | Advisory | IOP Compensation is currently not available. | User attempts to turn IOP Com- pensation on when it’s not avail- able. | ||
1238 | Advisory | An error occurred saving the device settings. | User attempt to save device set- tings fails. | ||
1239 | Advisory | Proportional reflux mode is currently not available. | User attempts to toggle into pro- portional reflux mode when the fluid level in the cassette cham- ber is out of range or the fluid level gets out of range while in proportional reflux. | ||
1240 | Advisory | Extraction flow mode is cur- rently not available. | User attempts to toggle into flow mode when the fluid level in the cassette chamber is out of range or the fluid level gets out of range while in flow mode. | ||
1241 | Advisory | Infusion source is getting low: please check the bottle. | The infusion container fluid level is getting low. | ||
1242 | Advisory | Command is not allowed due to instrument not being avail- able. | The user tried to invoke a com- mand which is not allowed when no instrument is available for the command to work with. | ||
1243 | Advisory | Infusion / Irrigation source is empty: please press [Change] and replace the bottle. | The infusion fluid container is empty but there’s still fluid in the cassette chambers. | ||
1244 | Advisory | Command is not allowed while priming, tuning, or testing. | The user attempts to do some- thing, e.g. change Steps, that’s not allowed when tests are in progress. | ||
1245 | Advisory | Please connect handpiece. | The user attempts to perform an action, e.g. priming, that requires a handpiece but no handpiece is connected. Note: This is only applicable to handpieces the system can de- tect connection for (e.g. Phaco) | ||
1247 | Advisory | Connected handpiece is not supported with the current cassette type. | User connects a handpiece that’s not compatible with the current Cassette type. (e.g., a Frag handpiece is connected when an Anterior Cassette is inserted) | ||
1248 | Advisory | Is the inserted cassette new? | The user has inserted a Cas- sette which the system can’t de- termine whether it’s new or the same Cassette that was previ- ously ejected. | ||
1249 | Advisory | An error occurred parsing the log file. | An error occurred parsing one of more lines of the log file. | ||
1250 | Advisory | Multi-Cut is not available when proportional scissors are selected. Press [Multi-Cut] to indicate that multi-cut scissors are currently connected. | The user has selected scis- sors of type Proportional and attempts to select the Multi- Cut submode or attempt to use momentary cutting in Extrusion Mode. | ||
1251 | Advisory | Press [Multi-Cut] to indicate that multi-cut scissors are currently connected. | The user tried to turn Infusion, Irrigation or FAX on while the cassette is being cleaned. | ||
1252 | Advisory | The scanned barcode is not recognized. | The user has scanned an item that the system doesn’t recog- nize. | ||
1255 | Advisory | Illuminator fiber is not con- nected | User has attempted to turn on an illuminator with no illuminator fiber connected. | ||
1256 | Advisory | Proportional reflux mode is not available when the footswitch treadle is de- pressed. | User attempts to enter Propor- tional Reflux mode when the treadle is down | ||
1257 | Advisory | The report’s header or foot- er has too many rows to fit on the page. | The user has added too many rows to a header or footer table being edited in an End Case report. As a result, the table will not be printed out in its entirety on the various pages compris- ing the report. | ||
1258 | Advisory | The report’s current table has too many rows to fit on the page. | The user has added too many rows to the current table being edited in an End Case report. As a result, the table will not be printed out in its entirety on the various pages comprising the report. | ||
1259 | Advisory | Command is not allowed while proportional diathermy is ac- tive. | The user attempts to enter pro- portional reflux while propor- tional diathermy is active. | ||
1260 | Advisory | Command is not allowed while proportional reflux is active. | The user attempts to enter pro- portional diathermy while pro- portional reflux is active. | ||
1261 | Advisory | Cassette can’t be ejected while being cleaned. | The user attempts to eject the cassette while it’s being cleaned. | ||
1262 | Advisory | Invalid handpiece tip. | The system attempts to select a tip that’s not valid for the specif- ic handpiece. This will only hap- pen if an incorrect tip is speci- fied for a scanned pak, i.e. the user will never be able to select an invalid tip from the UI. | ||
1263 | Advisory | Port can’t be selected: there is no probe connected. | The user tries to make a laser port the active port but there’s no probe connected to that port. | ||
1264 | Advisory | Port can’t be selected: the probe type for the port isn’t valid. | The user tries to make a laser port the active port but the cur- rently selected probe type for that port is invalid. | ||
1266 | Advisory | A laser probe is not connected to the active port. | Shown in any of the following situations: • Laser Step entered and no probe connected to the active port • Probe removed from active port • User tries to go to Ready Mode when no probe is connected to the active port | ||
1267 | Advisory | A valid laser probe is not se- lected for the active port. | Shown in any of the following situations: Laser Step entered and no valid probe type selected for the active port. User tries to go to the Ready Mode when no valid prove type is selected for the active port. | ||
1268 | Advisory | The laser remote interlock is open. | Shown in any of the following situations: • Laser Step entered and the Interlock is open • Interlock opened in a Laser Step • User tries to go to Ready Mode when the Interlock is open | ||
1269 | Advisory | Laser Dr. Filters are not con- nected to the console. Are all necessary Dr. Filters properly installed and con- nected? | Shown in any of the following situations: • Laser Step entered and the Dr. Filters haven’t been verified for the active port • Endo Probe inserted to the active port while in a Laser Step • User tries to go to Ready Mode when Dr. Filters haven’t been verified for the active port. | ||
1270 | Advisory | One Laser Dr. Filter is con- nected to the console. Are all necessary Dr. Filters properly installed and con- nected? | Same as above | ||
1271 | Advisory | Two Laser Dr. Filters are con- nected to the console. Are all necessary Dr. Filters properly installed and con- nected? | Same as above | ||
1273 | Advisory | Laser Dr. Filter 1 is disen- gaged | The user has disengaged the (connected) Dr. Filter 1 while in a Laser Step. | ||
1275 | Advisory | Laser Dr. Filter 2 is disen- gaged | Same as 1273 but for Dr. Filter 2 | ||
1276 | Advisory | Command is not allowed when Laser is firing. | User has attempted to change one of the laser’s settings while the laser is firing. | ||
1277 | Advisory | Port can’t be selected: it’s not functional. | User has attempted to select a laser port that’s not functional. | ||
1278 | Advisory | Cannot go to Laser Ready Mode while the current screen is being displayed. | User has pressed the Ready button from the footswitch when a “conflicting screen” is active. | ||
1279 | Advisory | No laser footswitch is con- nected. | User enters a Laser Step when no laser footswitch is connected or the laser footswitch is discon- nected in a Laser Step. | ||
1280 | Advisory | Unable to write to the report file. | A problem has occurred when trying to write to the specified Report file. | ||
1281 | Advisory | Unable to read from the report file. | A problem has occurred when trying to read from the specified Report file. | ||
1285 | Advisory | The report file and its associ- ated CRC value do not match. | The Report file and its associ- ated CRC value do not match. | ||
1288 | Advisory | Can’t eject the cassette while reflux is active. | The user attempts to eject the cassette while reflux (of any type) is active. For proportional reflux, this ad- visory is displayed when the user attempts cassette ejection while proportional reflux mode is active with the treadle not depressed. Cassette ejection while the treadle is depressed will generate advisory 1220. | ||
1289 | Advisory | Extraction flow mode is not available when FAX is turned on. | FAX was turned on while in Flow Mode or a Posterior Step with Flow Mode preference was selected when FAX was on. The system then changed to Vacu- um Mode. | ||
1290 | Advisory | The surgical function is cur- rently unavailable. | The user has requested surgi- cal functionality that is currently unavailable. This is a default ad- visory that is only displayed for cases in which a more meaning- ful explanation is not available. | ||
1296 | Advisory | The barcode reader failed to initialize. | The bar code reader failed to initialize during startup. | ||
1297 | Advisory | The system is currently low on free disk space. Please backup or remove non-critical files. | During startup, free space was detected at less than 15%. | ||
1298 | Advisory | A printer is not installed. The printer is out of paper. | The user tried to print but no printer is installed. The user sent a print job but the printer is out of, or ran out of paper. | ||
1299 | Advisory | The printer is off line. | The user requested a print job, but the printer is off line. | ||
1302 | Advisory | An unknown error occurred while printing. | Generated by any printer error other the errors above. | ||
1303 | Advisory | The Illuminators are currently turned off from the footswitch and cannot be turned on or off at this time. | The user has attempted to turn on or off an endo illuminator when the footswitch Momentary Endo Illuminators Off function is active. | ||
1304 | Advisory | The Laser is currently unable to deliver the maximum power level of 2 Watts. The maximum Laser power (Watts) currently available is: | Is generated when the power available from the laser drops below the maximum level (2 Watts). | ||
1305 | Advisory | An error occurred loading a surgeon. The system will re- vert to default values. | A database attempt to load a surgeon from the file system has failed. Reverting to defaults. | ||
1306 | Advisory | An error occurred setting the printer port. The system will revert to the previous value. | Attempt to set the printer IP ad- dress failed. | ||
1307 | Advisory | An error occurred loading a procedure. The system will re- vert to default values. | A database attempt to load a procedure from the file system has failed. Reverting to de- faults. | ||
1308 | Advisory | An error occurred loading the system settings. The system will revert to default values. | A database attempt to load the system settings from the file system has failed. Reverting to defaults. | ||
1309 | Advisory | An error occurred saving a surgeon’s settings. Changes will be reverted at the start of the next case. | A database attempt to save a surgeon to the file system has failed. | ||
1310 | Advisory | An error occurred saving a procedure. Changes will be reverted at the start of the next case. | A database attempt to save a procedure from the file system has failed. Reverting to de- faults. | ||
1311 | Advisory | An error occurred saving the system settings. The system will use default values. | A database attempt to save the system settings from the file system has failed. Reverting to defaults. | ||
1312 | Advisory | An error has occurred initializ- ing Video Overlay. Video Over- lay functions will be disabled. | An error happened during Vid- eoOverlay initialization. | ||
1313 | Advisory | An error has occurred in the Video Overlay component. Video Overlay functions will be disabled. | An error happened during Vid- eoOverlay processing. | ||
1314 | Advisory | There was an error saving sur- geon data. | There was an error saving sur- geon data. | ||
1317 | Advisory | There was an error renaming the surgeon. | There was an error renaming the surgeon. | ||
1318 | Advisory | There was an error renaming the procedure. | There was an error renaming the procedure. | ||
1319 | Advisory | There was an error renaming the case data. | There was an error renaming the case data. | ||
1320 | Advisory | There was an error deleting the surgeon. | There was an error deleting the surgeon. | ||
1321 | Advisory | There was an error deleting the procedure. | There was an error deleting the procedure. | ||
1322 | Advisory | There was an error deleting the case data. | There was an error deleting the case data. | ||
1323 | Advisory | There was an error loading the surgeon. | There was an error loading the surgeon. | ||
1324 | Advisory | There was an error loading the procedure. | There was an error loading the procedure. | ||
1325 | Advisory | There was an error loading the case data. | There was an error loading case data. | ||
1326 | Advisory | Unable to load the case report template. | There was an error loading the Case Report Template. | ||
1329 | Advisory | Unable to save the techni- cian’s log. | There was an error saving the Technician’s Log. | ||
1330 | Advisory | Unable to load the video table. | There was an error loading the Video Table. | ||
1331 | Advisory | Unable to play the video. | There was an error loading and\ or playing the Video file. | ||
1332 | Advisory | Unable to load help | There was an error opening the Help pdf file. | ||
1333 | Advisory | Unable to save the case Info. | There was an error saving the Case Info file. | ||
1334 | Advisory | Unable to update the log file | There was an error writing to the Log File. | ||
1335 | Advisory | Unable to read the log file | There was an error reading from the Log File. | ||
1336 | Advisory | Unable to write the incident file. | There was an error writing to the Incident File. | ||
1337 | Advisory | Unable to read the incident file | There was an error reading from the Incident File. | ||
1338 | Advisory | unable to write the system metric file | There was an error writing to the System Metrics File. | ||
1339 | Advisory | Unable to read the system metrics file | There was an error reading from the System Metrics File. | ||
1340 | Advisory | Unable to writ to the removable drive. | There was an error while at- tempting to write to the removable drive. | ||
1341 | Advisory | Further increasing the output level in air can damage fiber tips. Would you like to con- tinue? | The user has exceeded Thresh- old 1 (see SRS-5687.1) illumi- nator value by either turning on the illuminator, changing the setpoint while the illuminator is on, or changing Procedures while the illuminator is on. | ||
1342 | Advisory | Further increasing the output level will reduce exposure time by 35%. Would you like to con- tinue? | The user has exceeded Thresh- old 2 (see SRS-5687.3) illumi- nator value by either turning on the illuminator, changing the setpoint, or changing Proce- dures. | ||
1343 | Advisory | Further increasing the output level in air can damage fiber tips. Also, further increasing the output level will reduce exposure time by 35%. Would you like to continue? | The user has exceeded both Threshold 1 and Threshold 2 (see SRS-5687.1 and SRS- 5687.3) illuminator value by ei- ther turning on the illuminator, changing the setpoint while the illuminator is on, or changing Procedures while the illumina- tor is on. | ||
1344 | Advisory | Only two illuminators can be turned on simultaneously. | User attempts to turn on an illu- minator port when two ports are already on. This is not allowed. | ||
1345 | Advisory | AGF not allowed while For- ceps or VFC is in use. | User attempts to start Auto Gas Filling when either Forceps or VFC is in use (i.e. selected and treadle down). | ||
1346 | Advisory | An error occurred trying to ac- cess the wireless network. | An error occurs when the user attempts to access the wireless network. | ||
1349 | Advisory | A RAID hard drive has failed or is missing | One Disk in a redundant RAID volume is missing or has failed but the volume is still function- al. | ||
1350 | Advisory | The probe in the current laser port is not supported. | The type of probe in the current port is not supported by NGVS but might be supported by NGL (slit lamp for example). | ||
1355 | Advisory | Command is not allowed when the Pneumatics submodule is not functional. | The user tried to invoke a com- mand which is only allowed when the Pneumatics subsys- tem is functional. | ||
1356 | Advisory | Command is not allowed when an Anterior Only Cassette is being used. | The user tried to invoke a com- mand which is not allowed with an Anterior Only Cassette. | ||
1357 | Advisory | Cannot switch to Infusion: the cassette has not been primed. | The user tried to switch to Infu- sion when the cassette was not primed. | ||
1358 | Advisory | Current probe type is unrecog- nized. Please select a valid probe type. | User has attempted to turn on an illuminator when the probe type is “unrecognized”. | ||
1359 | Advisory | Power module communication error. When powering down the system, you will have to press the “Options/Shutdown” button on the screen. | Host controller failed to commu- nicate with power module. | ||
1361 | Advisory | Command is not allowed when Infusion backup pressure is active. | The user tried to invoke a com- mand which is only allowed when Infusion backup pressure is not active. | ||
1362 | Advisory | Command is not allowed when the probe type has been iden- tified by RFID. | The user tried to change the type of a laser probe that has been identified by RFID. | ||
1363 | Advisory | Command is not allowed when the probe type has been pre- viously identified by the user. Disconnect and then recon- nect the probe to change its type. | The user tried to change the type of a laser probe that has been identified by a user selec- tion. Only one selection is allowed after connection. | ||
1364 | Advisory | IOP Compensation is not allowed without a premium cas- sette inserted. Please insert a premium cassette and retry. | The user tried to turn on IOP compensation when a premium cassette is not inserted. | ||
1365 | Advisory | IOP Compensation is not allowed with an uncalibrated cassette. Please calibrate cas- sette and retry. | The user tried to turn on IOP compensation with a cassette that has not been calibrated. | ||
1366 | Advisory | Command is not allowed when FAX is on and the cassette has not been primed. | The user tried to perform a Test Instrument command when FAX is on and the cassette is not primed. | ||
1367 | Advisory | FAX is not allowed when the cassette is being primed. | The user tried to turn on FAX when the cassette is being primed. | ||
1369 | Advisory | Setting values is not allowed when the cassette is in the chamber overflow condition. | The cassette chamber is in an overflow condition and pressure cannot be controlled. | ||
1370 | Advisory | Test instrument is not allowed in this mode. | User has attempted to perform a test instrument command in Setup or End Case. | ||
1371 | Advisory | The remote control battery is low. | User has pressed on a remote control button and the remote’s battery is running out of power. | ||
1372 | Advisory | Infusion pressure will drop to zero during calibration. Con- tinue? | User has pressed the “Cali- brate” button on the advisory popup brought up when there’s been an IV Pole error. Since I/V Pole calibration will change the current Infusion pressure, the user is warned before calibra- tion is started. | ||
1751 | Information | Power recovery is in progress. Please wait until surgical func- tions become available. | AC power has been restored and the recovery process has started. | ||
1752 | Information | Fault - 2100 Call Field Service | A two second heart beat reply from the host was missed. | ||
1753 | Information | Communications failure with the Fluidics submodule. Fluid- ics functions will be disabled. | The supervisor cannot establish communicate with the Fluidics submodule. | ||
2100 | Fault | Communications failure with the Fluidics submodule. Fluid- ics functions will be disabled. | The supervisor lost communi- cate with the Fluidics submod- ule | ||
14300 | Information | Unable to perform this function without a primed cassette. | This advisory is produced when extraction is attempted without a primed cassette. | ||
2201 | Information | Unable to aspirate. Please turn on infusion. | This advisory is produced when extraction is attempted without infusion on. | ||
2202 | Advisory | Unable to aspirate while infu- sion/irrigation is unavailable. | This advisory is produced when extraction is attempted without infusion being functional. | ||
2203 | Advisory | Please wait: draining cassette. | This advisory is produced when extraction is attempted when the extraction chamber is in a overflow condition. | ||
2204 | Advisory | Unable to turn on infusion with- out a primed cassette. | This advisory is produced when infusion is attempted with out a primed cassette. | ||
2205 | Advisory | Unable to turn on infusion with- out a calibrated IV pole. | This advisory is produced when infusion is attempted with out a calibrated IV Pole. | ||
2208 | Advisory | Unable to turn on infusion with- out sufficient source pressure. | This advisory is produced when infusion is attempted with out source air pressure | ||
2209 | Advisory | Unable to turn on irrigation without a primed cassette. | This advisory is produced when irrigation is attempted with out a primed cassette. | ||
2210 | Advisory | Unable to turn on infusion with- out a calibrated IV pole. | This advisory is produced when irrigation is attempted with out a calibrated IV Pole. | ||
2211 | Advisory | Unable to turn on irrigation without sufficient source pres- sure. | This advisory is produced when irrigation is attempted with out suitable source pressure. | ||
2212 | Advisory | Irrigation unavailable: out of fluid. Please change the bot- tle. | This advisory is produced when irrigation is attempted with out sufficient irrigation fluid avail- able | ||
2213 | Advisory | FAX unavailable. Please insert a cassette. | This advisory is produced when F/AX is attempted with out a cassette inserted. | ||
2214 | Error | Extraction setpoint timeout. In- fusion/Irrigation and Extraction functions will be disabled. | This error is produced when the extraction is active and the extraction proxy fails to get a footswitch update within 20 msec | ||
2215 | Advisory | Micro reflux is currently not available. | This error is produced when the the user attempt to activate mi- cro reflux and the fluid level is out of range. | ||
2250 | Error | Communications failure with the Pneumatics submodule. Pneumatics functions will be disabled. | The supervisor cannot com- municate with the Pneumatics submodule. | ||
2251 | Error | Communications failure with the Pneumatics submodule. Pneumatics functions will be disabled. | This error is produced when the Supervisor losses communica- tion with the pneumatics mod- ule. | ||
2252 | Advisory | Unable to turn on cutting with- out sufficient source pressure. | This advisory is produced when cutter or utility operation is requested with out sufficient source pressure. | ||
2253 | Advisory | Unable to turn on cutting with excessive source pressure. | This advisory is produced when cutter or utility operation is re- quested with source pressure that is too high. | ||
2254 | Error | Pneumatics setpoint timeout. Pneumatics functions will be disabled. | This error is produced when the pneumatics is active and the pneumatics proxy fails to get a footswitch update within 20 msec. | ||
2300 | Error | Communications failure with the Ultrasound submodule. Ultrasound functions will be disabled. | The supervisor cannot commu- nicate with the Ultrasound sub- module. | ||
2301 | Error | Communications failure with the Ultrasound submodule. Ultrasound functions will be disabled. | The supervisor lost communi- cation with the Ultrasound sub- module. | ||
2302 | Error | US setpoint timeout. US and Diathermy functions will be disabled. | This error is produced when the US is active and the US proxy fails to get a footswitch update within 20 msec. | ||
2350 | Error | Communications failure with the Laser submodule. Laser functions will be disabled. | The supervisor cannot commu- nicate with the Laser submod- ule. | ||
2351 | Advisory | Communications failure with the Laser submodule. Laser functions will be disabled. | The supervisor lost communica- tion with the Laser submodule. | ||
2400 | Advisory | Please insert the Table Top Il- luminator drawer. | Supervisor detected a non –zero setpoint with the Table Top Illuminator drawer out. | ||
2401 | Advisory | The lamp in the Table Top Illu- minator needs to be replaced. Please contact Field Service. | Supervisor detected a non–zero setpoint with the Table Top Illu- minator Lamp bad. | ||
2500 | Advisory | Communications failure with the Power Control submodule. Please contact Field Service. | Supervisor detected a serial I/O error when trying to talk to the Power Module. | ||
2550 | Error | Footswitch error. Footswitch treadle functions will be dis- abled. | The footswitch has an error. | ||
2600 | Advisory | IV Pole over-current error. IV Pole functions will be dis- abled. | IV Pole motor is drawing too much current. | ||
2700 | Error | Communications failure with the Auxiliary Illuminator sub- module. Auxiliary Illuminator functions will be disabled. | The supervisor cannot commu- nicate with the Auxiliary Illumi- nator submodule. | ||
2701 | Advisory | Communications failure with the Auxiliary Illuminator sub- module. Auxiliary Illuminator functions will be disabled. | The supervisor lost communica- tion with the Auxiliary Illuminator submodule. | ||
2702 | Advisory | The lamp in the Auxiliary Illu- minator needs to be replaced. Please contact Field Service. | - Supervisor detected a non– zero setpoint with the Auxiliary Illuminator Lamp bad. | ||
2750 | Fault | Fault - 2750 Call Field Service | The Supervisor could not assert reset control over all submod- ules. A module that was sup- posed to be reset responded with its module information. | ||
2751 | Fault | Fault - 2751 Call Field Service | The Supervisor got duplicate slot ID for the submodules. | ||
3100 | Advisory | Could not calibrate the fluid level sensors. Please eject and reinsert, or replace the cassette. | Level Sensor calibration failed. The calibration done status bit was not set by the FPGA within the specified timeout limit or the calculated required max/min pixel gain was outside the valid range. | ||
3200 | Error | Cassette latch error. Please remove and reinsert the cas- sette. | Cassette latch hardware feed- back signals indicate a hard- ware problem. | ||
3201 | Advisory | Invalid cassette ID detected. Please eject and insert a valid cassette. | An invalid cassette id was read by the cassette id sensors. | ||
3202 | Advisory | Cassette test failed. Please eject and reinsert, or replace cassette. | Cassette pressure and/or vac- cum tests failed. | ||
3203 | Advisory | The cassette was not properly latched into position. Please remove and reinsert the cas- sette. | The cassette latch optical po- sition sensor indicates that the latch did not reach its locked position. | ||
3204 | Error | Cassette ID sensor error. Infu- sion/irrigation and extraction functions will be disabled. | Cassette ID sensor test failed. The sensor output voltage is not within the expected range. | ||
3205 | Advisory | The Fluidics module fan is not working. | The Fluidics module fan tachom- eter indicates that the Fluidics module fan is not operating. | ||
3300 | Advisory | Draining cassette. Please wait. | The cassette aspiration cham- ber is full of fluid. | ||
3302 | Advisory | N/A | Not enough or too much fluid in the aspiration chamber to al- low aspiration flow mode and reflux. NOTE: This advisory is never explicitly displayed. Instead ad- visory 1239 or 1240 might be displayed if appropriate. See 11239 and 11240 for details. | ||
3304 | Advisory | Leak test failure. Please con- firm the irrigation tubing; as- piration tubing , and the test chamber are properly connect- ed to the handpiece. | A leak in the aspiration or irriga- tion tubing was detected during flow check of a handpiece. | ||
3305 | Advisory | Priming of the aspiration probe was unsuccessful. Please at- tempt to re-prime the probe. | Probe prime was unsuccess- ful. The required volume of fluid was not transferred through the aspiration tubing set within the specified timeout period. | ||
3306 | Advisory | Priming of the aspiration handpiece was unsuccessful. Please attempt to re-prime the handpiece. | Handpiece prime was unsuc- cessful. The required volume of fluid was not transferred through the aspiration tubing set within the specified timeout period. | ||
3307 | Error | Aspiration level sensor prob- lem detected. Infusion/irriga- tion and extraction functions will be disabled. | The level sensor status signals indicate a level sensor hard- ware problem. | ||
3308 | Advisory | Flow check failure: measured flow restriction is too high. Ex- traction and Ultrasound func- tions in Phaco/Frag will be disabled. | The handpiece failed the flow- check. Too high vacuum level was required to achieve the ref- erence flow level. | ||
3309 | Advisory | Flow check failure: aspiration chamber could not be filled with fluid. Extraction and Ultra- sound functions in Phaco/Frag will be disabled. | The handpiece failed the flow- check. Too high vacuum level was required to achieve the ref- erence flow level. | ||
3325 | Advisory | Drain bag is almost full. Please replace bag and press [Done]. | ~50 cc remaining volume in drain bag. | ||
3326 | Advisory | Drain bag is full. Please re- place bag and press [Done]. | ~0 cc remaining volume in drain bag. | ||
3327 | Advisory | Drain bag is critically full. Please replace bag and press [Done]. | Drain bag filled > 50 cc above capacity. | ||
3329 | Error | Drain pump problem detected. Infusion/irrigation and extrac- tion functions will be disabled. | The commanded pump rate does not correspond to the ac- tual pump rate measured by the optical encoder. | ||
3330 | Advisory | The drain pump fan is not working. | The drain pump fan tachometer indicates that the drain pump fan is not operating. | ||
3331 | Advisory | Could not drain the aspiration chamber. Please remove and insert a new cassette. | The aspiration chamber could not be drained within the speci- fied timeout period. | ||
3350 | Error | Extraction pressure transducer offset error. Infusion/irrigation and extraction functions will be disabled. | Extraction pressure transducer 0 offset out of range. | ||
3351 | Error | Extraction pressure transducer discrepancy error. Infusion/irri- gation and extraction functions will be disabled. | A discrepancy between the pri- mary and redundant extraction pressure transducers was de- tected. | ||
3352 | Error | Extraction isolation valve error. Infusion/irrigation and extrac- tion functions will be disabled | Extraction isolation valve hardware error. | ||
3353 | Error | Reflux valve error. Infusion/irri- gation and extraction functions will be disabled. | Extraction reflux valve hardware error. | ||
3354 | Error | Extraction output valve (S1) error. Infusion/irrigation and extraction functions will be dis- abled. | Extraction output pincher valve (S1) error | ||
3355 | Error | Extraction output valve (S11) error. Infusion/irrigation and extraction functions will be dis- abled. | Extraction output port 1 pincher valve error. | ||
3356 | Error | Extraction output valve (S22) error. Infusion/irrigation and extraction functions will be dis- abled. | Extraction output port 2 pincher valve error. | ||
3357 | Error | Extraction cross-connection valve (SC) error. Infusion/irri- gation and extraction functions will be disabled. | Extraction “normally open” cross-connection valve error. | ||
3358 | Error | Extraction cross-connection valve (SC2) error. Infusion/irri- gation and extraction functions will be disabled. | Extraction “normally closed” cross-connection valve error. | ||
3359 | Advisory | Suction pressure surges de- tected. Vacuum will be dis- abled. Please release the footswitch treadle to reset. | Extraction pressure oscillations detected. | ||
3360 | Advisory | Suction flow surges detected. Flow will be disabled. Please release the footswitch treadle to reset. | Extraction flow oscillations detected. | ||
3361 | Advisory | Suction pressure is too high. Vacuum will be disabled. Please release the footswitch treadle to reset. | Extraction pressure overshoot detected. | ||
3362 | Advisory | Aspiration flow too high. Flow will be disabled. Please re- lease the footswitch treadle to reset. | Extraction flow overshoot detected. | ||
3363 | Error | Extraction pressure transducer reference voltage out of range. Infusion/irrigation and extrac- tion functions will be disabled. | Extraction transducer reference voltage error. | ||
3400 | Error | Infusion source container pres- sure transducer offset error. In- fusion/irrigation and extraction functions will be disabled. | Infusion source container pressure transducer 0 offset out of range. | ||
3401 | Error | Infusion source container iso- lation valve error. Infusion/irri- gation and extraction functions will be disabled. | Infusion source container isola- tion valve error. | ||
3402 | Error | Infusion source container transducer discrepancy error. Infusion/irrigation and extrac- tion functions will be disabled. | A dispcrepancy between the source container and LPAS pressure transducers was detected. | ||
3403 | Error | Infusion source container pressure too high. Infusion/irri- gation and extraction functions will be disabled. | Infusion source container pres- sure too high. | ||
3420 | Error | Infusion pressure transducer offset error. Infusion/irrigation and extraction functions will be disabled. | Infusion pressure transducer 0 offset out of range | ||
3421 | Error | Infusion pressure transducer discrepancy error. Infusion/irri- gation and extraction functions will be disabled. | A discrepancy between the primary and redundant infusion pressure transducers was detected. | ||
3422 | Error | Infusion isolation valve error. Infusion/irrigation and extrac- tion functions will be disabled. | Infusion isolation valve error.. | ||
3423 | Error | Infusion FAX valve error. Infu- sion/irrigation and extraction functions will be disabled. | Infusion FAX valve error | ||
3424 | Error | Infusion input valve error. Infu- sion/irrigation and extraction functions will be disabled. | Infusion chamber input pincher valve error. | ||
3425 | Error | Infusion output valve error. In- fusion/irrigation and extraction functions will be disabled. | Infusion chamber output pincher valve error. | ||
3426 | Error | Irrigation output valve error. In- fusion/irrigation and extraction functions will be disabled. | Irrigation output pincher valve error. | ||
3427 | Error | Infusion pressure surges de- tected. Infusion/irrigation and extraction functions will be disabled. | Infusion pressure oscillations detected. | ||
3428 | Error | Infusion pressure too high. In- fusion/irrigation and extraction functions will be disabled. | Infusion high pressure detect- ed. | ||
3429 | Advisory | Low infusion pressure de- tected. Please check infusion connections. Select 30 mmHg backup pressure, or ignore low pressure condition. | Infusion pressure too low. | ||
3431 | Advisory | Low irrigation pressure detected. Please check irrigation connections. | Irrigation pressure too low. | ||
3433 | Error | Infusion chamber isolation valve error. Infusion/irrigation and extraction functions will be disabled. | Infusion chamber isolation valve detected to be stuck closed during power up diagnostics. | ||
3434 | Advisory | Infusion chamber overflow error. Control of infusion pressure has possibly been lost. The current pressure, in mmHg, could be as high as: | Infusion chamber overflow error. The fluid level in the infusion chamber reached above the overflow level. | ||
3436 | Error | FAX valve error. Infusion/irrigation and extraction functions will be disabled. | F/AX valve detected to be stuck closed during power up diagnostics. | ||
3437 | Error | Iinfusion level sensor error. Infusion/irrigation and extraction functions will be disabled. | The level sensor status signals indicate a level sensor hardware problem. | ||
3438 | Advisory | No more infusion fluid available. Press [Change] to change the infusion bottle. | The infusion chamber in the cassette is empty. | ||
3440 | Advisory | No more irrigation fluid available. Press [Change] to change the infusion/irrigation bottle. | The irrigation chamber in the cassette is empty. | ||
3442 | Advisory | Irrigation chamber overflow detected. Control of irrigation pressure has possibly been lost. The current pressure, in mmHg, could be as high as: | Irrigation chamber overflow error. The fluid level in the irrigation chamber reached above the overflow level. | ||
3460 | Error | Infusion backup valve error. Infusion/irrigation, extraction, and Ultrasound functions will be disabled. | Infusion backup valve error. | ||
3461 | Error | Infusion LPAS pump error detected. Infusion/irrigation, extraction, and Ultrasound functions will be disabled. | The infusion LPAS pump tachometer indicate that the LPAS pump is not operating correctly. | ||
3462 | Advisory | Infusion flow sensor communication error. IOP Compensation functions will be disabled. | Infusion flow sensor communication error. | ||
3464 | Advisory | Infusion flow data invalid: IOP Compensation functions will be disabled. Check infusion tubing for air bubbles | Infusion flow sensor readings are not valid. | ||
3466 | Advisory | Infusion flow sensor accuracy error. IOP Compensation functions will be disabled. | Infusion flow sensor accuracy error. Flow sensor and chamber fluid volume measurements are out of range of each other. Detected during flow sensor calibration. | ||
3467 | Advisory | Infusion flow sensor signal amplitude is low. IOP Compensation functions will be disabled. Please eject and reinsert the cassette. | Infusion flow sensor signal amplitude error. Detected during flow sensor calibration | ||
3469 | Advisory | Tubing calibration offset error. IOP Compensation functions will be disabled. Please position the infusion cannula at the height of the center of the cassette and re-prime. | Infusion tubing calibration 0 offset advisory. The infusion cannula was not positioned within the correct vertical range of the cassette during calibration. | ||
3470 | Advisory | Calibration verification error: calculated pressures using the acquired calibration profile are not within the expected range for the selected infusion cannula. IOP Compensation functions will be disabled. Please re-prime. | Infusion tubing calibration check point advisory. The pressure drop calculated using the acquired calibration profile was not within the expected range. | ||
3471 | Advisory | Noisy calibration flow readings. IOP Compensation functions will be disabled. Please re-prime. | Infusion tubing calibration standard deviation advisory. The standard deviation between the acquired flow measurements and the calculated calibration profile was larger than the specified max limit. | ||
3472 | Advisory | Infusion chamber leak detected. Please eject and replace the cassette. | A leak in the infusion chambers was detected during priming. | ||
3473 | Error | Infusion pressure transducer reference voltage out of range. Infusion/irrigation and extraction functions will be disabled. | Infusion transducer reference voltage error. | ||
3474 | Error | Infusion NIFS valve error detected. Infusion/irrigation and extraction functions will be disabled. | Infusion NIFS valve error. | ||
3475 | Advisory | Infusion prime failed. | Infusion prime failed. SEE CR4 3475 Details. | ||
3476 | Advisory | The infusion chambers did not fill with fluid. Please check the infusion bottle and connections or press [Change] to replace the infusion bottle. | Infusion chamber did not fill within the specified timeout period. Source container is out of fluid. | ||
3477 | Error | Infusion LPAS pressure transducer offset error. Infusion/irrigation and extraction functions will be disabled. | Infusion LPAS pressure transducer 0 offset out of range. | ||
3478 | Error | Infusion LPAS pressure error. Infusion/irrigation and extraction functions will be disabled. | Infusion LPAS source pressure too high. | ||
3479 | Advisory | Infusion LPAS pressure is low. Infusion pressure loss is possible. | Infusion LPAS pump output pressure is low. | ||
3481 | Advisory | Infusion flow sensor is disconnected. IOP Compensation functions will be disabled. | Infusion flow sensor is disconnected or the sensor connection has failed. | ||
4100 | Error | U/S voltage failure (+5 analog). Ultrasound and Diathermy functions will be disabled. | US Kernel Analog 5 volt feedback is bad. US Submodule is non-functional. | ||
4101 | Error | U/S voltage failure (+2.5). Ultrasound and Diathermy functions will be disabled. | US Kernel +2.5 volt feedback is bad. US submodule is non-functional. | ||
4102 | Error | U/S voltage failure (-2.5). Ultrasound and Diathermy functions will be disabled. | US Kernel -2.5 volt feedback is bad. US submodule is non-functional. | ||
4103 | Error | U/S voltage failure (+8.5). Ultrasound and Diathermy functions will be disabled. | US Kernel 2.5 volt feedback is bad. US submodule is non-functional. | ||
4111 | Error | US failure: SPI driver write timeout. Ultrasound and Diathermy functions will be disabled. | SPI driver timed out waiting for a write to complete. US submodule is non-functional. | ||
4200 | Advisory | Handpiece EEPROM CRC is invalid. Please replace the handpiece. | The handpiece EEPROM CRC is invalid. Handpiece needs to be replaced. | ||
4201 | Advisory | Only one US handpiece may be connected at a time. Please remove one of the handpieces. | Two US handpieces are connected. One of the handpieces must be removed before the other handpiece can be used. | ||
4202 | Advisory | Handpiece current is too low. Please replace handpiece and re-tune. | U/S handpiece current is too low. A short circuit in the handpiece can cause this. | ||
4203 | Advisory | Handpiece voltage is too low. Please replace handpiece and re-tune. | U/S handpiece voltage is too low. An open circuit in the handpiece can cause this. | ||
4204 | Advisory | Handpiece power is too high. Please replace handpiece and re-tune | U/S handpiece power output is too high. | ||
4206 | Advisory | Handpiece power DC2DC output was out of range. Please release the footswitch treadle to reset. If problem persists, please contact Field Services. | While powering the handpiece, the DC2DC voltage for handpiece power is out of range. | ||
4207 | Advisory | Handpiece was removed while powered. Please reconnect handpiece and re-tune. | User has disconnected the handpiece while it’s being powered. | ||
4208 | Advisory | Data in handpiece EEPROM is out of range. Please replace the handpiece. | Data contained in the handpiece is out of range. | ||
4209 | Advisory | Unknown US handpiece connected. Please connect a known handpiece. | The system has detected that a US handpiece was connected but it cannot determine its type. | ||
4210 | Advisory | Unsupported US handpiece connected. Please connect a supported handpiece. | The system has detected the connection of a recognized US handpiece but that handpiece is not supported by this system. Handpiece is not tuned. | ||
4220 | Advisory | Tune failure – attempted while handpiece was in air. Please re-tune the handpiece. | The handpiece was tuned while in air. Handpiece is not tuned. | ||
4221 | Advisory | Tune failure: handpiece was removed before tuning. Please connect a handpiece and re-tune. | A handpiece tune was requested but no handpiece is connected. | ||
4222 | Advisory | Tune failure: handpiece is an unknown type. Please connect a known handpiece and re-tune. | A handpiece tune was requested but an unknown type of handpiece is connected. | ||
4223 | Advisory | Tune failure: handpiece has a loose tip. Please tighten the tip and re-tune. | The handpiece tip was loose when tuned. Handpiece is not tuned. | ||
4224 | Advisory | Tune failure: handpiece current is low. Please replace handpiece and re-tune. | The handpiece current was too low (open circuit). Handpiece is not tuned. | ||
4225 | Advisory | Tune failure: handpiece voltage is low. Please replace handpiece and re-tune. | The handpiece voltage was too low when tuned (short circuit). Handpiece is not tuned. | ||
4226 | Advisory | Tune failure: handpiece frequency order error. Please re-tune. | The series (low impedance) and parallel (high impedance) frequencies were out of order when the handpiece was tuned. | ||
4228 | Advisory | Tune failure: handpiece series frequency margin error. Please re-tune | The series tune frequency was too close to the tune start frequency while tuning the handpiece. The handpiece is not tuned. | ||
4229 | Advisory | Tune failure: handpiece parallel frequency margin error. Please re-tune. | The parallel tune frequency was too close to the tune end frequency while tuning the handpiece. The handpiece is not tuned. | ||
4231 | Advisory | Tune failure: handpiece frequency bandwidth too low. Please re-tune. | The difference between the series and parallel tune frequencies was too small while tuning the handpiece. The handpiece is not tuned. | ||
4232 | Advisory | Tune failure: handpiece frequency bandwidth too high. Please re-tune | The difference between the series and parallel tune frequencies was too large while tuning the handpiece. The handpiece is not tuned. | ||
4234 | Advisory | Tune failure: handpiece DC2DC output out of range. Please re-tune. If problem persists, please contact Field Service. | The DC2DC voltage was out of range while tuning the handpiece. The handpiece is not tuned. | ||
4235 | Advisory | Tune failure: handpiece removed while tuning. Please connect handpiece and re-tune. | The handpiece was removed while tuning. The handpiece is not tuned. | ||
4240 | Advisory | The requested Frag continuous power is too high. The Power level will be limited. | A request for more than 60% frag power was made while not in a pulsed mode. The power will be limited to 60%. | ||
4250 | Error | Ultrasound failure: ADC calibration. Ultrasound functions will be disabled. | The ADC feedback reading with power off for DC2DC voltage, handpiece voltage, or handpiece current was too high. | ||
4300 | Error | Diathermy failure: DC2DC output was out of range. Diathermy functions will be disabled. | While powering the handpiece, the DC2DC voltage for handpiece power is out of range. | ||
4301 | Advisory | Diathermy power is too high. Please release the footswitch treadle / button and try again. | Too much power was being delivered to the diathermy handpiece. Diathermy power is turned off. The operator must release the treadle/switch and the depress the treadle/switch to re-activate power. | ||
5100 | Error | Ballast failure (voltage). Illuminator functions will be disabled. | Ballast transducer has reading that is out of specified safety range. | ||
5101 | Error | Ballast failure (current). Illuminator functions will be disabled. | Ballast transducer has reading that is out of specified safety range. | ||
5102 | Advisory | Lamp calibration data is corrupted: lamp needs to be calibrated. Please contact Field Service. | The calibration data saved in the flash contains incorrect CRC checksum. | ||
5103 | Advisory | Failure to turn lamp on: lamp needs to be replaced. Please contact Field Service. | Lamp status reported by hardware is not the same as software status. | ||
5104 | Advisory | The lamp has exceeded its rated life. Please replace the lamp. | The lamp has exceeded its rated life. The new lamp is expected. | ||
5105 | Advisory | The lamp has exceeded its rated maximum life. Please install a new lamp immediately. | The lamp has exceeded its rated safe life. The new lamp should be installed immediately. | ||
5107 | Advisory | The Table-Top Illuminator drawer is ejected. Please close the drawer to continue. | The module has been pulled out of drawer. | ||
5109 | Information | The calibration data for the Illuminator has changed: due to recalibration or replacement with a new unit. | The calibration data is changed. | ||
5200 | Advisory | Illuminator optics temperature is high. The lamp will be turned off if the temperature continues to rise. | Optics temperature is too high. Lamp is going to be shut down if temperature continues to rise. | ||
5201 | Error | Illuminator optics temperature has exceeded its limit. Illuminator functions will be disabled. | Illuminator is shut down because Optics temperature is too high. | ||
5202 | Advisory | Illuminator optics fan is at full speed. Optics unit may be overheating. | Warning that the Optics fan is full on. | ||
5203 | Error | Illuminator optics thermo-cut-off has been triggered. Illuminator functions will be disabled. | Lamp is turned off because thermo cut-off. | ||
5204 | Advisory | Communication failure with the Illuminator optics fan. The fan may not work properly. | Communication Error with optics fan. | ||
5300 | Advisory | Illuminator ballast temperature is high. The lamp will be turned off if the temperature continues to rise. | Ballast temperature is too high. Lamp is going to be shut down if temperature continues to rise. | ||
5301 | Error | Illuminator ballast temperature has exceeded its limit. Illuminator functions will be disabled. | Illuminator is shut down because Ballast temperature is too high. | ||
5302 | Advisory | Illuminator ballast fan is at full speed. Ballast unit may be overheating. | Warning that the Ballast fan is full on | ||
5303 | Error | Illuminator ballast thermo-cut-off has been triggered. Illuminator functions will be disabled. | Lamp is turned off because thermo cut-off. | ||
5304 | Advisory | Communication failure with the Illuminator ballast fan. The fan may not work properly. | Communication Error with ballast fan. | ||
5400 | Error | Lamp louver failure at port 1: unable to move to home position. Port 1 will be disabled. | Step motor at port 1 failed to move to home position. | ||
5401 | Error | Lamp Louver failure at port 2: unable to move to home position. Port 2 will be disabled. | Step motor at port 2 failed to move to home position. | ||
5402 | Error | Lamp Louver failure at port 1: unable to move to specified position. Port 1 will be disabled. | Step motor at port 1 failed to move to specified position. | ||
5403 | Error | Lamp Louver failure at port 2: unable to move to specified position. Port 2 will be disabled. | Step motor at port 2 failed to move to specified position. | ||
6101 | Error | Pressure reading is too high. Cutting and Pneumatics functions will be disabled. | Pneumatics pressure is too high. | ||
6201 | Error | Air Pressure valves have high transition faults. Cutting and Pneumatics functions will be disabled. | AirPressure valves have high level faults (fail to open). | ||
6202 | Error | Air Pressure valves have low transition faults. Cutting and Pneumatics functionals will be disabled. | AirPressure valves have low level faults (fail to close). | ||
6203 | Advisory | Air Pressure inlet filter may be dirty and needs to be replaced. Please contact Field Service. | The unusual high pressure drop cross the Air Filter. This is caused by either the Air Filter is too dirty and need to be replaced or the system may be leaking air. | ||
6204 | Advisory | Inlet pressure is too low for the system. Please adjust the inlet pressure to higher than 58 psi. | The Air source (wall pressure) is too low (below 58 psi) to turn on. On startup, Pneumatics turns on the pressure automatically if the pressure is high enough (above 58 psi). | ||
6206 | Advisory | Inlet pressure is too high for the system. Please adjust the inlet pressure to less than 140 psi. | The Air source (wall pressure) is too high (above 120 psi) to turn on. On startup, Pneumatics turns on the pressure automatically if the pressure is high enough (above 58 psi). | ||
6208 | Error | Inlet pressure is unstable. Pneumatics functions will be disabled. Please restart the system. | The Air source pressure (wall pressure) is unstable to cause the system unexpected to turn On/Off pressure. | ||
6301 | Error | Cutting errors; valves have high transition faults (failed to open). Cutting functions will be disabled. | Cutters valves have high level faults (fail to open). | ||
6302 | Error | Cutting errors; valves have low transition faults (failed to close). Cutting functions will be disabled. | Cutters valves have low level faults (fail to close). | ||
6303 | Error | Cutting error: redundant transducers discrepancy error. Cutting functions will be disabled. Please contact Field Service. | Cutters redundant transducers discrepancy error. | ||
6304 | Error | Cutting error: redundant transducers calibration. Cutting functions will be disabled. Please contact Field Service. | Cutters redundant transducers calibration error. | ||
6305 | Advisory | Cutting pressure is oscillating beyond the specified range. Please release the footswitch treadle to reset. | Cutters actual pressure is oscillating beyond the specified tolerance. | ||
6306 | Advisory | Cutting pressure is surging beyond the specified tolerance. Please release the footswitch treadle to reset. | Cutters actual pressure is surging beyond the specified tolerance. | ||
6401 | Error | Utilities valves have high transition faults (failed to open). Pneumatics functions will be disabled. | Utilities valves have high level faults (fail to open). | ||
6402 | Error | Utilities valves have low transition faults (failed to close). Pneumatics functions will be disabled. | Utilities valves have low level faults (fail to close). | ||
6403 | Error | Utilities redundant transducers discrepancy error. Pneumatics functions will be disabled. Please contact Field Service. | Utilities redundant transducers discrepancy error. | ||
6404 | Error | Utilities redundant transducers calibration error. Pneumatics functions will be disabled. Please contact Field Service. | Utilities redundant transducers calibration error. | ||
6405 | Error | Vacuum redundant transducers discrepancy error. Pneumatics functions will be disabled. Please contact Field Service. | Vacuum redundant transducers discrepancy error. | ||
6406 | Error | Vacuum redundant transducers calibration error. Pneumatics functions will be disabled. Please contact Field Service. | Vacuum redundant transducers calibration error. | ||
6407 | Advisory | Utilities pressure is oscillating beyond the specified range. Please release the footswitch treadle to reset. | Utilities actual pressure is oscillating beyond the specified tolerance. | ||
6408 | Advisory | Utilities pressure is surging beyond the specified tolerance. Please release the footswitch treadle to reset. | Utilities actual pressure is surging beyond the specified tolerance. | ||
6501 | Error | Auto Gas valves have high transition faults (failed to open). Pneumatics functions will be disabled. | Auto Gas valves have high level faults (fail to open). | ||
6502 | Error | Auto Gas valves have low transition faults (failed to close). Pneumatics functions will be disabled. | Auto Gas valves have low level faults (fail to close). | ||
6503 | Error | Auto Gas redundant transducers discrepancy error. Pneumatics functions will be disabled. Please contact Field Service. | Auto Gas redundant transducers discrepancy error. | ||
6504 | Error | Auto Gas redundant transducers calibration error. Pneumatics functions will be disabled. Please contact Field Service. | Auto Gas redundant transducers calibration error. | ||
6505 | Advisory | C3F8 bottle may be empty and needs to be replaced. Please press [Replaced] to confirm bottle replacement. | Gas 1 bottle may be empty and needs to be filled up. | ||
6506 | Advisory | SF6 bottle may be empty and needs to be replaced. Please press [Replaced] to confirm bottle replacement. | Gas 2 bottle may be empty and needs to be filled up. | ||
7100 | Error | Ballast failure (voltage). Illuminator functions will be disabled. | Ballast transducer has reading that is out of specified safety range. | ||
7101 | Error | Ballast failure (current). Illuminator functions will be disabled. | Ballast transducer has reading that is out of specified safety range. | ||
7102 | Advisory | Lamp calibration data is corrupted: lamp needs to be calibrated. Please contact Field Service | The calibration data saved in the flash contains incorrect CRC checksum. | ||
7103 | Advisory | Failure to turn lamp on: lamp needs to be replaced. Please contact Field Service. | Lamp status reported by hardware is not the same as software status. | ||
7104 | Advisory | The lamp has exceeded its rated life: lamp needs to be replaced. Please contact Field Service. | The lamp has exceeded its rated life. The new lamp is expected. | ||
7105 | Advisory | The lamp has exceeded its rated maximum life: lamp needs to be replaced. Please contact Field Service. | The lamp has exceeded its rated safe life. The new lamp should be installed immediately. | ||
7200 | Advisory | Illuminator optics temperature is high. The lamp will be turned off if the temperature continues to rise. | Optics temperature is too high. Lamp is going to be shut down if temperature continues to rise. | ||
7201 | Error | Illuminator optics temperature has exceeded its limit. Illuminator functions will be disabled. | Illuminator is shut down because Optics temperature is too high. | ||
7202 | Advisory | Illuminator optics fan is at full speed. Optics unit may be overheating. | Warning that the Optics fan is full on. | ||
7203 | Error | Illuminator optics thermo-cut-off has been triggered. Illuminator functions will be disabled. | Lamp is turned off because thermo cut-off. | ||
7204 | Advisory | Communication failure with the Illuminator optics fan. The fan may not work properly. | Communication Error with optics fan. | ||
7300 | Advisory | Illuminator ballast temperature is high. The lamp will be turned off if the temperature continues to rise. | Ballast temperature is too high. Lamp is going to be shut down if temperature continues to rise. | ||
7301 | Error | Illuminator ballast temperature has exceeded its limit. Illuminator functions will be disabled. | Illuminator is shut down because Ballast temperature is too high. | ||
7302 | Advisory | Illuminator ballast fan is at full speed. Ballast unit may be overheating. | Warning that the Ballast fan is full on. | ||
7303 | Error | Illuminator ballast thermo-cut-off has been triggered. Illuminator functions will be disabled. | Lamp is turned off because thermo cut-off. | ||
7304 | Advisory | Communication failure with the Illuminator ballast fan. The fan may not work properly. | Communication Error with ballast fan. | ||
7400 | Error | Lamp louver failure at port 3: unable to move to home position. Port 3 will be disabled. | Step motor at port 1 failed to move to home position. | ||
7401 | Error | Lamp louver failure at port 4: unable to move to home position. Port 4 will be disabled. | Step motor at port 2 failed to move to home position. | ||
7402 | Error | Lamp louver failure at port 3: unable to move to specified position. Port 3 will be disabled. | Step motor at port 1 failed to move to specified position. | ||
7403 | Error | Lamp louver failure at port 4: unable to move to specified position. Port 4 will be disabled. | Step motor at port 2 failed to move to specified position. | ||
8100 | Error | Laser controller software error. Laser functions will be disabled. | Need description. | ||
8101 | Error | Laser controller shutter open for too long error. Laser functions will be disabled. | Need description. | ||
8102 | Error | Laser controller shutter unexpected error. Laser functions will be disabled. | Need description. | ||
8103 | Error | Laser controller shutter open between firing error. Laser functions will be disabled. | Need description. | ||
8104 | Error | Laser controller mirror incorrect position error. Laser functions will be disabled. | Need description. | ||
8105 | Error | Laser controller mirror position unexpected error. Laser functions will be disabled. | Need description. | ||
8106 | Error | Laser controller power over the limit error. Laser functions will be disabled. | Need description. | ||
8107 | Advisory | Laser controller maximum current exceeded error. Laser functions will be disabled. | Need description. | ||
8108 | Error | Laser controller port power detection error. Laser functions will be disabled. | Need description. | ||
8109 | Error | Laser controller power detected at wrong port error. Laser functions will be disabled. | Need description. | ||
8110 | Error | Laser controller power reading mismatch error. Laser functions will be disabled. | Need description. | ||
8111 | Error | Laser controller LBO crystal limit error. Laser functions will be disabled. | Need description. | ||
8112 | Error | Laser controller pump temperature limit error. Laser functions will be disabled. | Need description. | ||
8113 | Error | Laser controller power detect digital I/O mismatch error. Laser functions will be disabled. | Need description. | ||
8114 | Error | Laser controller footswitch digital I/O mismatch error. Laser functions will be disabled. | Need description. | ||
8115 | Error | Laser controller footswitch no NC (normally closed) error. Laser functions will be disabled. | Need description. | ||
8117 | Advisory | Laser controller internal parameter error. | Need description. | ||
8118 | Error | Laser controller process hunt error. Laser functions will be disabled. | Need description. | ||
8119 | Error | Laser controller process attach error. Laser functions will be disabled. | Need description. | ||
8120 | Error | Laser controller start up time out error. Laser functions will be disabled. | Need description. | ||
8121 | Error | Laser controller JEM CRC error. Laser functions will be disabled. | Need description. | ||
8122 | Error | Laser controller kernel CRC error. Laser functions will be disabled. | Need description. | ||
8123 | Error | Laser controller flash file error. Laser functions will be disabled. | Need description. | ||
8124 | Advisory | Laser LBO diode over temperature error. | Need description. | ||
8125 | Advisory | Laser LBO diode under temperature error. | Need description. | ||
8126 | Advisory | Laser controller crystal over-temperature error. | Need description. | ||
8127 | Advisory | Laser controller crystal under-temperature error. | Need description. | ||
8129 | Advisory | Laser controller probe connection error. | Need Description | ||
8131 | Error | Laser controller power monitor POST error. Laser functions will be disabled. | Need Description | ||
8132 | Error | Laser controller +2.5 volt out-of-range error. Laser functions will be disabled. | Need Description | ||
8134 | Error | Laser controller 12 volt power error. Laser functions will be disabled. | Need Description | ||
8135 | Error | Laser shutter timer expired error. Laser functions will be disabled. | Need Description | ||
8138 | Error | Laser controller diode thermal electric cooler error. Laser functions will be disabled. | Need Description | ||
8139 | Error | Laser probe port process POST error. Laser functions will be disabled. | Need Description | ||
8140 | Error | Laser controller startup timeout error (laser engine). Laser functions will be disabled. | Need Description | ||
8141 | Error | Laser controller startup timeout error (shutter control process). Laser functions will be disabled. | Need Description | ||
8142 | Error | Laser controller startup timeout error (supervisor process). Laser functions will be disabled. | Need Description | ||
8143 | Error | Laser controller startup timeout error (peripheral management process). Laser functions will be disabled. | Need Description | ||
8144 | Error | Laser controller startup timeout error (probe port process). Laser functions will be disabled. | Need Description | ||
8145 | Error | Laser controller startup timeout error (laser system controller). Laser functions will be disabled. | Need Description | ||
8147 | Error | Laser controller TMP crystal startup timeout error. Laser functions will be disabled. | Need Description | ||
8148 | Error | Laser controller TMP diode startup timeout error. Laser functions will be disabled | Need Description | ||
8149 | Error | Laser controller Dr Filter error. Laser functions will be disabled. | Need Description | ||
8150 | Error | Laser controller invalid maximum power error. Laser functions will be disabled. | Need Description | ||
8153 | Error | Laser controller load module CRC error. Laser functions will be disabled. | Need Description | ||
8154 | Error | Laser controller module requires calibration. Laser functions will be disabled. | Need Description | ||
8200 | Advisory | Laser controller probe removed while firing. | Need Description | ||
8202 | Advisory | Laser controller Dr. Filter disengaged while firing. | Need Description | ||
8203 | Advisory | Laser controller Dr. Filter disconnected while firing. | Need Description | ||
8204 | Advisory | Laser controller interlock opened while firing. | Need Description | ||
8205 | Advisory | Laser controller footswitch disconnected while firing. | Need Description | ||
8207 | Advisory | Laser controller port changed in Ready State. | Need Description | ||
8208 | Advisory | Laser controller Dr. Filter disengaged in Ready State. | Need Description | ||
8209 | Advisory | Laser controller Dr. Filter disconnected in Ready State. | Need Description | ||
8210 | Advisory | Laser controller interlock opened in Ready State. | Need Description | ||
8211 | Advisory | Laser controller footswitch removed in Ready State. | Need Description | ||
8212 | Advisory | Laser controller footswitch engaged when Ready requested. | Need Description | ||
8213 | Advisory | Laser controller user data CRC error. | Need Description | ||
8214 | Advisory | Laser controller Ready State denied: the filter is disengaged. | Need Description | ||
8215 | Advisory | Laser controller Ready State denied: there is no interlock. | Need Description | ||
8216 | Advisory | Laser controller Ready State denied: there is no valid probe for the active port. | Need description. | ||
8217 | Advisory | Laser controller Ready State denied: the footswitch is depressed. | Need description. | ||
8218 | Advisory | Are all necessary Dr. Filters properly installed and connected? | User requests going to Ready Mode when the doctor filters haven’t been verified. | ||
8219 | Advisory | Laser controller Ready State denied: there is no footswitch present. | User presses the Laser Ready button when no footswitch is connected. | ||
8220 | Advisory | Laser controller Dr. Filter connected while in Transition, Ready or Firing State. | User connects an engaged Dr. Filter while in Transition / Ready or Firing mode and and Endo probe is connected to the active port. |
Constellation CR4
Event Code | Event Type | Message to User | Detail / Discrepancy | System Action |
---|---|---|---|---|
X001 | Error | Sub module system software error detected. | OSE_ERROR_CODE; Internal OSE RTOS error. | Application Software - Reload; Controller PCB. |
X005 | Error | Sub module failure (+5V). | KRNL_5_VOLTS_IS_BAD; Kernel +5 Voltage is bad. | Controller PCB |
X006 | Error | Sub module failure (+3.3V). | KRNL_3_3_VOLTS_IS_BAD; Kernel +3.3 Voltage is bad | Controller PCB |
X008 | Error | Sub module application failed to load. | KRNL_PRG_LOAD_FAILED; The kernel failed loading a load module component. Attempted to load application code from Host and it failed. | Application Software - Reload; Controller PCB |
X009 | Error | Sub module application failed to start. | KRNL_PRG_START_FAILED; The kernel failed starting a load module component. Software has downloaded, but module failed to run | Controller PCB |
X013 | Error | Sub module software failure (watchdog timeout). | KRNL_WD_TIME_OUT; Watchdog had a task timeout. | Controller PCB |
X015 | Error | Sub module failure (voltage checker). | KRNL_VOLTAGE_CHECKER_UNDERRUN; Kernel Voltage checker ADC underrun. | Controller PCB |
X017 | Error | Sub module failure (+24V). | KRNL_24_VOLTS_IS_BAD; Kernel +24 Voltage is bad. | Power Cable; Controller PCB |
X018 | Error | Sub module failure (+1.5V). | KRNL_1_5_VOLTS_IS_BAD; Kernel +1.5 Voltage is bad. | • Controller PCB |
X024 | Error | Sub module failure (+1.2V). | KRNL_1_2_VOLTS_IS_BAD Kernel +1.2 Voltage is bad. | • Controller PCB |
X025 | Error | Sub module failure (reference voltage A). | KRNL_REF_A_IS_BAD; Kernel Voltage Reference A is bad. | • Controller PCB |
X026 | Error | Sub module failure (reference voltage B). | KRNL_REF_B_IS_BAD; Kernel Voltage Reference B is bad. | • Controller PCB |
X027 | Error | Sub module failure (reference voltage C). | KRNL_REF_C_IS_BAD; Kernel Voltage Reference C is bad. | • Controller PCB |
X028 | Error | Sub module failure (+2.5V). | KRNL_2_5_VOLTS_IS_BAD; Kernel +2.5 Voltage is bad. | • Controller PCB |
X029 | Error | Sub module failure (FPGA). | KRNL_FPGA_NOT_READY; The FPGA ready bit is not set. | • Controller PCB |
X030 | Error | Sub module failure (communication error). | KRNL_SUPERVISOR_COMM_LOSS; The Kernel Status Handler lost communication with the Supervisor module. | Ethernet Cable; Supervisor PCB |
X033 | Error | Sub module failure (watchdog test). | KRNL_WD_INIT_TEST_FAILED; The watchdog startup test failed on power up. | • Controller PCB |
X035 | Error | Sub module failure (ECC memory error). | KRNL_RAM_RUNTIME_FAILED; Bad SDRAM ECC errors were detected. | • Controller PCB |
X037 | Error | Sub module failure (Ethernet receiver). | KRNL_FCC_INTERNAL_RX_ERROR; CRC Error on Ethernet communication. | Ethernet Cable; Controller PCB |
X038 | Error | Sub module failure (Ethernet transmitter). | KRNL_FCC_INTERNAL_TX_ERROR; CRC Error on Ethernet communication. | Ethernet Cable; Controller PCB |
X039 | Error | Sub module failure (+12V). | KRNL_12V_VOLTAGE_IS_BAD; Kernel 12 Voltage is bad error (Pneumatics and Fluidics modules only). | • Controller PCB |
X050 | Error | Sub module failure (software). | CMN_SW_ERR - A sub module software error was detected. Similar to 1014 for the Host system. NOTE: Table Top Only will show 2050 not 5050 because Table Top communicates through Supervisor) | Application Software - Reload; Controller PCB |
X051 | Error | Sub module failure (invalid parameter). | CMN_PARAM_ERR - An invalid parameter value was passed to a function. | • Controller PCB |
X053 | Error | Sub module failure (ADC timeout). | CMN_ADC_TIMEOUT - Timeout waiting for the ADC done bit to get set. | • Controller PCB |
X054 | Error | Sub module failure (DAC timeout). | CMN_DAC_TIMEOUT - Timeout waiting for the DAC busy bit to be cleared. | • Controller PCB |
X055 | Error | Sub module failure (critical data invalid). | CMN_CRITICAL_DATA_ERR - Critical data parameter error. The data and the inverted value do not match. | • Controller PCB |
X056 | Error | Sub module failure (watchdog non- operational). | CMN_WD_ATTACH_ERR - Received an attach signal indicating that the kernel watchdog process died. | • Controller PCB |
X057 | Error | Sub module failure (supervisor timeout) | CMN_SUPERVISOR_TIMEOUT_ERR - Supervisor set point message timeout. The supervisor did not send set point messages at the correct rate when a surgical function was active. | Ethernet cables; Supervisor; Controller PCB; Footswitch Cable |
X059 | Error | Sub module failure (access violation). | CMN_ACCESS_ERR - A process tried to access a resource without sufficient privileges. | • Controller PCB |
X061 | Advisory | Sub module failure (unknown message) | CMN_UNKNOWN_SIG_ERR - A signal not recognized by a process was received. The extra parameter contains a pointer to the unknown signal. | • Controller PCB |
X062 | Error | Submodule failure (triggered to demonstrate recovery functionality). | Error intentionally generated by the Host to test submodule error recovery functionality. Used for user training purposes. | • N/A |
X099 | Error | Sub module failure (no communication with Host) | CMN_SUBSYSTEM_UNAVAILABLE_ERR - The Host did not receive a Sub module information message within approximately 3 1/2 minutes after the splash screen was removed after power up or restart. NOTE: Laser and Auxiliary Illuminators are exempt from this error due to the fact that their power state/ejection state can change at anytime and be reset. | • Controller PCB |
1001 | Fault | Fault - 1001; Call Field Service | OseGatewayDownFlt - The OSE Gateway exited unexpectedly. | • Isolated occurrence - Cycle power; Host; Supervisor |
1002 | Fault | Fault - 1002; Call Field Service | SupervisorProcessDownFlt - The OSE Gateway has reported that the Supervisor process has gone down. | • Isolated occurrence - Cycle power; Supervisor |
1003 | Fault | Fault - 1003; Call Field Service | SupervisorCommErrorFlt - Something has failed in the communication between the Host and the Supervisor. | • Isolated occurrence - Cycle power; Host; Supervisor |
1004 | Fault | Fault - 1004; Call Field Service | MainApplicationDownFlt - The Main Application exited unexpectedly. | • Isolated occurrence - Cycle power; Application Software - Install; SDRAM - Reseat or Replace; Host; Supervisor |
1005 | Fault | Fault - 1005; Call Field Service | CrcChecksumFailureFlt – The Host Controller checksum validation failed to pass all files under its control. | • Isolated occurrence - Cycle power; Host; Supervisor |
1006 | Fault | Fault - 1006; Call Field Service | MainApplicationFailed ToConnectToSupervisorFlt The Host has failed to connect to the Supervisor process. | Isolated occurrence - Cycle power; Signal/Ethernet Cable W17; Isolated occurrence - Cycle power; Host; Supervisor; 24V Supply to Supervisor |
1007 | Fault | Fault - 1007; Call Field Service | HostControllerSoftwareErrorFlt - The Host Controller encountered an unexpected error. | • Isolated occurrence - Cycle power; Host; Supervisor |
1008 | Fault | Fault - 1008; Call Field Service | HostControllerExeLaunchFlt - The Host Controller failed to launch an executable image during startup. | • Isolated occurrence - Cycle power; Reset Power Controller |
1009 | Fault | Fault - 1009; Call Field Service | PowerModuleCommFailureFlt - The Host Controller could not communicate with the Power Module over the serial link | • Isolated occurrence - Cycle power; Host; Power Controller PCB |
1010 | Fault | Fault - 1010; Call Field Service | VersionCheckFailureFlt- Component or Submodule reported an incorrect software, hardware, firmware or other version | Isolated occurrence - Cycle power; Verify correct Software OS and Application; are installed; Reinstall OS or Application Software. |
1011 | Fault | Fault - 1011; Call Field Service | IpcProxyMissingFlt – Host Controller communications was lost as the result of a .NET IPC failure | • Isolated occurrence - Cycle power; Host; Supervisor |
1012 | Fault | Fault - 1012; Call Field Service | MissingSupervisorHeartbeatFlt The Host has not received any heartbeats from the Supervisor within the required timeframe. | Supervisor; Ethernet cable W17; Host |
1013 | Fault | Fault - 1013; Call Field Service | UiThreadUnresponsiveFlt - The Host has detected that the User Interface thread has not responded within the required timeframe (i.e. the UI is considered locked up) | • Isolated occurrence - Cycle power; Host; Supervisor |
1014 | Fault | Fault - 1014; Call Field Service | SoftwareErrorFlt The system has detected a software error. | • Isolated occurrence - Cycle power; Host; Supervisor |
1015 | Fault | Fault - 1015; Call Field Service | UnknownSupervisorMsgFlt The Host has received a message from the Supervisor that it doesn’t recognize (this is a specific type of software error) | • Isolated occurrence - Cycle power; Host; Supervisor |
1016 | Fault | Fault - 1016; Call Field Service | ApplicationFontsFailureFlt The Host application’s display fonts can not be loaded. | • Application Software - Reinstall |
1017 | Fault | Fault - 1017; Call Field Service | AudioPlaybackFlt A failure has occurred during the playback of an audio file. | • Isolated occurrence - Cycle power; Host |
1018 | Fault | Fault - 1018; Call Field Service | The Main Application failed to communicate with the Host Controller. | • Isolated occurrence - Cycle power; Host |
1019 | Fault | Fault - 1019; Call Field Service | The Host Controller failed to communicate with the Main Application. | • Isolated occurrence - Cycle power; Host |
1020 | Fault | Fault - 1020; Call Field Service | The Main Application detected an invalid configuration for the host or submodules. | • Isolated occurrence - Cycle power; Host |
1021 | Fault | Fault - 1021; Call Field Service | Fault intentionally generated by the Host to test the fault recovery functionality. Used for user training purposes. | |
1022 | Fault | Fault - 1022; Call Field Service | Fault intentionally generated by the Host to demonstrate the system behavior when an unrecoverable fault is triggered. Used for user training purposes. | |
1107 | Error | Unable to load one or more language packs. | InvalidLanguageLoadAttemptErr An error occurred while attempting to load a non-English language. | Verify correct version of Language; Translation Tool is being used. |
1108 | Error | Incompatible version numbers within the Fluidics submodule. Fluidics functions will be disabled. | IncompatibleVersionFluidicsErr - The Fluidics module was shut down due to an incompatible version. The log file shows exactly what version information was incompatible. | Software corruption or incompatibility; Fluidics Controller PCB |
1109 | Error | Incompatible version numbers within the Pneumatics submodule. Pneumatics functions will be disabled. | IncompatibleVersionPneumaticsErr The Pneumatics module was shut down due to an incompatible version. The log file shows exactly what version information was incompatible. | Software corruption or incompatibility; Pneumatics Controller PCB |
1110 | Error | Incompatible version numbers within the Ultrasound submodule. Ultrasound and Diathermy functions will be disabled. | IncompatibleVersionUltrasoundErr The Ultrasound module was shut down due to an incompatible version. The log file shows exactly what version information was incompatible. | Software corruption or incompatibility; U/S Controller PCB |
1111 | Error | Incompatible version numbers within the Auxiliary Illuminator submodule. Auxiliary Illuminator functions will be disabled. | IncompatibleVersionAuxIllumErr The Auxiliary Illuminator module was shut down due to an incompatible version. The log file shows exactly what version information was incompatible. | Software corruption or incompatibility; Laser Controller PCB |
1112 | Error | Incompatible version numbers within the Laser submodule. Laser functions will be disabled. | IncompatibleVersionLaserErr The Laser module was shut down due to an incompatible version. The log file shows exactly what version information was incompatible. | Software corruption or incompatibility; Aux Controller PCB |
1200 | Advisory | Function is not allowed when the fluidics subsystem is not functional. | CommandNotAllowedFluidicsNotFunctionalAdv The user tried to select Fluidics function which is only allowed when the fluidics subsystem is functional. | • User Advisory - See Discrepancy description |
1202 | Advisory | Function is not allowed when the Laser is in Ready Mode or firing. | CommandNotAllowedWhenLaserReadyOrFiringAdv The user tried to invoke a command which is not allowed when the laser is in Ready Mode or Firing. | • User Advisory - See Discrepancy description |
1203 | Advisory | Function is not allowed when the footswitch treadle is down or buttons are pressed. | CommandNotAllowedWhenTreadleDownAdv The user tried to invoke a command which is only allowed when the treadle is up and no buttons pressed. | • User Advisory - See Discrepancy description |
1204 | Advisory | Flow mode is not available until the probe or handpiece has been primed. | ExtrFlowModeUnavailProbeHpNotPrimedAdv Flow mode is unavailable because a probe or handpiece hasn’t been primed. | • User Advisory - See Discrepancy description |
1205 | Advisory | Flow mode is only available for 20 gauge probes / handpieces. | ExtrFlowModeUnavailProbeHpNot20GaugeAdv Flow mode is unavailable because a non 20 gauge probe or handpiece is used. | • User Advisory - See Discrepancy description |
1207 | Advisory | Please connect a Phaco handpiece. | NoPhacoHpConnectedAdv The user presses the treadle in a Phaco Step when no Phaco handpiece is connected. | • User Advisory - See Discrepancy description |
1208 | Advisory | Please tune the Phaco handpiece. | PhacoHpNotTunedAdv The user presses the treadle in a Phaco Step when the Phaco handpiece isn’t tuned. | • User Advisory - See Discrepancy description |
1209 | Advisory | Please connect a Frag handpiece. | NoFragHpConnectedAdv The user presses the treadle in a Frag Step when no Frag handpiece is connected. | • User Advisory - See Discrepancy description |
1210 | Advisory | Please tune the Frag handpiece. | FragHpNotTunedAdv The user presses the treadle in a Phaco Step when the Phaco handpiece isn’t tuned. | • User Advisory - See Discrepancy description |
1211 | Advisory | Function is not allowed when the Ultrasound submodule is not functional. | CommandNotAllowedUltrasoundNotFunctionalAdv The user tried to invoke a command which is only allowed when the ultrasound subsystem is functional. | • User Advisory - See Discrepancy description |
1212 | Advisory | Function is not allowed when the cassette is not ready. | CommandNotAllowedCassetteNotReadyAdvThe user tried to invoke a command which is only allowed when the cassette is ready. | • User Advisory - See Discrepancy description |
1213 | Advisory | Infusion must be on to use VFC Extract. Press [Ignore] to allow VFC Extract without infusion for this case. | InfusionNeededInInVfcExtractAdv The user presses the treadle in a VFC Extract Step when infusion is off. | • User Advisory - See Discrepancy description |
1214 | Advisory | System hasn’t detected VFC. If already connected, press “Connected” button. | NoVfcSyringeConnectedAdv The user presses the treadle in a VFC Step when no VFC Syringe is connected. | • User Advisory - See Discrepancy description |
1215 | Advisory | System hasn’t detected Forceps. If already connected, press “Connected” button. | NoForcepsConnectedAdv The user presses the treadle in a Forceps Step when no forceps are connected. | • User Advisory - See Discrepancy description |
1216 | Advisory | System hasn’t detected Scissors. If already connected, press “Connected” button. | NoScissorsConnectedAdv The user presses the treadle in a Scissors Step when no Scissors are connected. | • User Advisory - See Discrepancy description |
1217 | Advisory | Please connect an AGF syringe. | NoAgfSyrinceConnectedAdv - The user presses the “Start” button in the Auto Gas Filling dialog when no AGF syringe has been connected. | • User Advisory - See Discrepancy description |
1218 | Advisory | Forceps not available during AGF: please try again when AGF is complete. | ForcepsNotAvailableDuringAgfAdv The user presses the treadle in a Forceps Step while Auto Gas Filling is in progress. | • User Advisory - See Discrepancy description |
1219 | Advisory | VFC not available during AGF: please try again when AGF is complete. | VfcNotAvailableDuringAgfAdv The user presses the treadle in a VFC Step while Auto Gas Filling is in progress. | • User Advisory - See Discrepancy description |
1220 | Advisory | Ejecting the cassette is not allowed while the footswitch treadle is pressed. | CassetteEjectRejectedTreadleDownAdv The user attempts to eject the cassette while the treadle is down. | • User Advisory - See Discrepancy description |
1221 | Advisory | Ejecting the cassette is not allowed while infusion or FAX is on. | CassetteEjectRejectedInfOnAdv The user attempts to eject the cassette while infusion is on. | • User Advisory - See Discrepancy description |
1222 | Advisory | Ejecting the cassette is not allowed while irrigation is on. | CassetteEjectRejectedIrrOnAdv The user attempts to eject the cassette while irrigation is on. | • User Advisory - See Discrepancy description |
1223 | Advisory | Ejecting the cassette is not allowed while priming, tuning, or testing. | CassetteEjectRejectedTestInProgrAdv The user attempts to eject the cassette while tests (priming/ tuning/testing) are in progress. | • User Advisory - See Discrepancy description |
1224 | Advisory | Cleaning the cassette is not allowed while infusion, irrigation, or FAX is on. | CassetteCleanNotAllowedWhenInfOrIrrIsOnAdv The user attempts to start Cassette Cleaning when either Infusion, Irrigation or FAX is on. | • User Advisory - See Discrepancy description |
1225 | Advisory | Cleaning the cassette is not allowed without a functional cassette. | CassetteCleanNotAllowedCassetteNotAvailableAdv The user attempts to start Cassette Cleaning when the Cassette isn’t available (e.g. not inserted). | • User Advisory - See Discrepancy description |
1226 | Advisory | Priming the cassette is not allowed without a functional cassette. | User attempt to prime the cassette when preconditions aren’t met (e.g. no cassette inserted, cassette not tested or didn’t pass the test) | • User Advisory - See Discrepancy description |
1227 | Advisory | An error occurred loading the device settings. The system will revert to default values. | LoadDeviceSettingsFailureAdv A database attempt to load the device settings from the file system has failed. Reverting to defaults. | • User Advisory - See Discrepancy description |
1228 | Advisory | Command is not allowed while infusion is on. | CommandNotAllowedInfusionIsOnAdv The user tried to invoke a command which is only allowed when infusion is not on. | • User Advisory - See Discrepancy description |
1229 | Advisory | Command is not allowed while irrigation is on. | CommandNotAllowedIrrigationIsOnAdv The user tried to invoke a command which is only allowed when irrigation is not on. | • User Advisory - See Discrepancy description |
1230 | Advisory | The selected step is not supported with the current cassette type. | IncompatibleOperatingModeAdv The user attempts to select a Step whose mode/submode isn’t compatible with the current operating mode. | • User Advisory - See Discrepancy description |
1231 | Advisory | The connected probe is not supported with the current cassette type. | VitProbeNotCompatibleAdv The user connects a probe (with RFID) that’s not compatible with the current operating mode (e.g. an Ultra Vit probe when an Anterior cassette is connected or Ultra Vit Anterior probe when a Posterior cassette is connected) | • User Advisory - See Discrepancy description |
1232 | Advisory | Please connect footswitch. | NoFootswitchAdv The Supervisor has reported that no footswitch is connected. | • User Advisory - See Discrepancy description |
1233 | Advisory | The connected footswitch is not supported. | UnknownFootswitchAdv The Supervisor has detected the connection of an unknown footswitch. | • User Advisory - See Discrepancy description |
1234 | Advisory | Flow limit is currently not available due to flow mode being unavailable. | ExtrVacModeFlowLimitUnavailableAdv Vacuum mode flow limit depends on the availability of flow mode. If flow mode becomes unavailable then the flow limit function can not be performed by the fluidics submodule. | • User Advisory - See Discrepancy description |
1236 | Advisory | Gravity pressure mode is currently not available. | GravityBasedPressureUnavailableAdv User attempts to select Gravity as the infusion source when a Premium Cassette is inserted. | • User Advisory - See Discrepancy description |
1237 | Advisory | IOP Control is currently not available. | IopCompensationUnvailableAdv User attempts to turn IOP Control on when it’s not available. | • User Advisory - See Discrepancy description |
1238 | Advisory | An error occurred saving the device settings. | SaveDeviceSettingsFailureAdv User attempt to save device settings fails. | • User Advisory - See Discrepancy description |
1239 | Advisory | Proportional reflux mode is currently not available. | PropRefluxModeUnavailFluidLvlNotInRangeAdv Displayed in one of the following situations: 1.User attempts to toggle into proportional reflux mode when the fluid level in the cassette chamber is out of range or the fluid 2.Fluid level gets out of range while in proportional reflux. 3.Treadle is depressed while in proportional reflux and the fluid level is out of range. | • User Advisory - See Discrepancy description |
1240 | Advisory | Extraction flow mode is currently not available. | ExtrFlowModeUnavailFluidLvlNotInRangeAdv The fluid level in the cassette chamber is out of range and the treadle is depressed. | • User Advisory - See Discrepancy description |
1241 | Advisory | Infusion source is getting low: please check the fluid container. | InfContainerNearEmptyAdv The infusion container fluid level is getting low. | • User Advisory - See Discrepancy description |
1242 | Advisory | Command is not allowed due to instrument not being available. | Command is not allowed due to instrument not being available. | • User Advisory - See Discrepancy description |
1243 | Advisory | Infusion / Irrigation source is empty: please press [Change] and replace the fluid container. | InfContainerOnReserveAdv The infusion fluid container is empty but there’s still fluid in the cassette chambers. | • User Advisory - See Discrepancy description |
1244 | Advisory | Command is not allowed while priming, tuning, or testing. | CommandNotAllowedWhenTestInProgressAdv The user attempts to do something, e.g. change Steps, that’s not allowed when tests are in progress. | • User Advisory - See Discrepancy description |
1245 | Advisory | Please connect handpiece. | NoHpConnectedAdv The user attempts to perform an action, e.g. priming, that requires a handpiece but no handpiece is connected. Note: This is only applicable to handpieces the system can detect connection for (e.g. Phaco) | • User Advisory - See Discrepancy description |
1247 | Advisory | Connected handpiece is not supported with the current cassette type. | HpNotCompatibleWithOperatingModeAdv User connects a handpiece that’s not compatible with the current Cassette type. (e.g., a Frag handpiece is connected when an Anterior Cassette is inserted) | • User Advisory - See Discrepancy description |
1248 | Advisory | Is the inserted cassette new? | IsInsertedCassetteNewAdv The user has inserted a Cassette which the system can’t determine whether it’s new or the same Cassette that was previously ejected. | • User Advisory - See Discrepancy description |
1249 | Advisory | An error occurred parsing the log file. | LogFileParseFailureAdv An error occurred parsing one of more lines of the log file. | • User Advisory - See Discrepancy description |
1250 | Advisory | Multi-Cut is not available when proportional scissors are selected. Press [Multi-Cut] to indicate that multi-cut scissors are currently connected. | ScissorsMultiCutUnavailWhenPropScissSelAdv The user has selected scissors of type Proportional and attempts to select the Multi-Cut submode or attempt to use momentary cutting in Extrusion Mode. The advisory lets the user change both the Scissors type and the Scissors submode. | • User Advisory - See Discrepancy description |
1251 | Advisory | Command is not allowed while the cassette is being cleaned. | The user tried to turn Infusion, Irrigation or FAX on while the cassette is being cleaned. | • User Advisory - See Discrepancy description |
1252 | Advisory | The scanned barcode is not recognized. | UnknownBarcodeScannedAdv The user has scanned an item that the system doesn’t recognize. | • User Advisory - See Discrepancy description |
1255 | Advisory | Illuminator fiber is not connected. | NoIlluminatorConnectedAdv User has attempted to turn on an illuminator with no illuminator fiber connected. | • Probe not fully inserted • Fiber Detection Cable mounting screw too Short • Fiber Detection Cable assembly • RFID PCB |
1256 | Advisory | Proportional reflux mode is not available when the footswitch treadle is depressed. | PropRefluxUnavailableTreadleDownAdv User attempts to enter Proportional Reflux mode when the treadle is down | • User Advisory - See Discrepancy description |
1257 | Advisory | The report’s header or footer has too many rows to fit on the page. | ReportExcessiveHeaderFooterSizeAdv The user has added too many rows to a header or footer table being edited in an End Case report. As a result, the table will not be printed out in its entirety on the various pages comprising the report. | • User Advisory - See Discrepancy description |
1258 | Advisory | The report’s current table has too many rows to fit on the page. | ReportExcessiveTableSizeAdv The user has added too many rows to the current table being edited in an End Case report. As a result, the table will not be printed out in its entirety on the various pages comprising the report. | • User Advisory - See Discrepancy description |
1259 | Advisory | Command is not allowed while proportional diathermy is active. | CommandNotAllowedInProportionalDiathermyAdv The user attempts to enter proportional reflux while proportional diathermy is active. | • User Advisory - See Discrepancy description |
1260 | Advisory | Command is not allowed while proportional reflux is active. | CommandNotAllowedInProportionalRefluxAdv The user attempts to enter proportional diathermy while proportional reflux is active. | • User Advisory - See Discrepancy description |
1261 | Advisory | Cassette can’t be ejected while being cleaned. | CassetteEjectRejectedCassetteIsCleaningAdv The user attempts to eject the cassette while it’s being cleaned. | • User Advisory - See Discrepancy description |
1263 | Advisory | Port can’t be selected: there is no probe connected. | LsrActivePortSelRejectedNoProbeConnAdv The user tries to make a laser port the active port but there’s no probe connected to that port. | • User Advisory - See Discrepancy description |
1264 | Advisory | Port can’t be selected: the probe type for the port isn’t valid. | LsrActivePortSelRejectedNoValidProbeSelAdv The user tries to make a laser port the active port but the currently selected probe type for that port is invalid. | • User Advisory - See Discrepancy description |
1266 | Advisory | A laser probe is not connected to the active port. | LsrNoProbeConnAdv Shown in any of the following situations: · Laser Step entered and no probe connected to the active port · Probe removed from active port · User tries to go to Ready Mode when no probe is connected to the active port | • User Advisory - See Discrepancy description |
1267 | Advisory | A valid laser probe is not selected for the active port. | LsrNoValidProbeSelAdv Shown in any of the following situations: · Laser Step entered and no valid probe type selected for the active port · User tries to go to Ready Mode when no valid probe type is selected for the active port | • User Advisory - See Discrepancy description |
1268 | Advisory | The laser remote interlock is open. | LsrInterlockOpenAdv Shown in any of the following situations: · Laser Step entered and the Interlock is open · Interlock opened in a Laser Step · User tries to go to Ready Mode when the Interlock is open | • User Advisory - See Discrepancy description |
1269 | Advisory | Laser Dr. Filters are not connected to the console. Are all necessary Dr. Filters properly installed and connected? | LsrDoctorFilterUnverified0ConnAdv Shown in any of the following situations: Laser Step entered and the Dr. Filters haven’t been verified for the active port · Endo Probe inserted to the active port while in a Laser Step. User tries to go to Ready Mode when Dr. Filters haven’t been verified for the active port. | • User Advisory - See Discrepancy description |
1270 | Advisory | One Laser Dr. Filter is connected to the console. Are all necessary Dr. Filters properly installed and connected? | LsrDoctorFilterUnverified1ConnAdv Same as above | • User Advisory - See Discrepancy description |
1271 | Advisory | Two Laser Dr. Filters are connected to the console. Are all necessary Dr. Filters properly installed and connected? | LsrDoctorFilterUnverified2ConnAdv Same as above | • User Advisory - See Discrepancy description |
1273 | Advisory | Laser Dr. Filter 1 is disengaged | LsrDoctorFilter1DisengagedAdv The user has disengaged the (connected) Dr. Filter 1 while in a Laser Step. | • User Advisory - See Discrepancy description |
1275 | Advisory | Laser Dr. Filter 2 is disengaged | LsrDoctorFilter2DisengagedAdv Same as 1273 but for Dr. Filter 2 | • User Advisory - See Discrepancy description |
1276 | Advisory | Command is not allowed when Laser is firing. | LsrCommandNotAllowedWhenFiringAdv User has attempted to change one of the laser’s settings while the laser is firing. | • User Advisory - See Discrepancy description |
1277 | Advisory | Port can’t be selected: it’s not functional. | LsrActivePortSelRejectedPortFailureAdv User has attempted to select a laser port that’s not functional. | • User Advisory - See Discrepancy description |
1278 | Advisory | Cannot go to Laser Ready Mode while the current screen is being displayed. | LsrReadyModeRejectedScreenActiveAdv User has pressed the Ready button from the footswitch when a “conflicting screen” is active. | • User Advisory - See Discrepancy description |
1279 | Advisory | No laser footswitch is connected. | LsrNoFootswitchConnectedAdv User enters a Laser Step when no laser footswitch is connected or the laser footswitch is disconnected in a Laser Step. | • User Advisory - See Discrepancy description |
1280 | Advisory | Unable to write to the report file. | ReportFileWriteErrorAdv A problem has occurred when trying to write to the specified Report file. | • User Advisory - See Discrepancy description |
1281 | Advisory | Unable to read from the report file. | ReportFileReadErrorAdv A problem has occurred when trying to read from the specified Report file. | • User Advisory - See Discrepancy description |
1285 | Advisory | The report file and its associated CRC value do not match. | ReportFileCrcMismatchErrorAdv The Report file and its associated CRC value do not match. | • User Advisory - See Discrepancy description |
1288 | Advisory | Can’t eject the cassette while reflux is active. | CassetteEjectRejectedRefluxInUseAdv The user attempts to eject the cassette while reflux (of any type) is active. For proportional reflux, this advisory is displayed when the user attempts cassette ejection while proportional reflux mode is active with the treadle not depressed. Cassette ejection while the treadle is depressed will generate advisory 1220. | • User Advisory - See Discrepancy description |
1289 | Advisory | Extraction flow mode is not available when FAX is turned on. | ExtrFlowModeUnavailFaxInUseAdv FAX was turned on while in Flow Mode or a Posterior Step with Flow Mode preference was selected when FAX was on. | • User Advisory - See Discrepancy description |
1290 | Advisory | The surgical function is currently unavailable. | SurgicalFunctionProxyUnavailableAdv The user has requested surgical functionality that is currently unavailable. This is a default advisory that is only displayed for cases in which a more meaningful explanation is not available. | • User Advisory - See Discrepancy description |
1292 | Advisory | An error occurred accessing the video recorder. Please verify it is on and connected. | VideoRecorderNotCommunicatingAdv An error occurred accessing the video recorder. | • Check connection to Video Recorder. |
1296 | Advisory | The barcode reader failed to initialize. | BarcodeReaderInitFailureAdv The bar code reader failed to initialize during startup. | • User Advisory - See Discrepancy description |
1297 | Advisory | The system is currently low on free disk space. Please backup or remove non-critical files. | LowFreeSpace During startup, free space was detected at less than the desired minimum %. | • User Advisory - See Discrepancy description |
1298 | Advisory | A printer must be configured in System Settings before printing. | PrinterNotConfiguredAdv The user tried to print but has not configured the printer in system settings | • User Advisory - See Discrepancy description |
1299 | Advisory | The current printer configuration could not be saved. | DefaultPrintSettingsErrorAdv An error occurred attempting to save the default printer with the current configuration | • User Advisory - See Discrepancy description |
1302 | Advisory | An unknown error occurred while printing. | PrinterGeneralErrorAdv Generated by any printer error other the errors above. | • User Advisory - See Discrepancy description |
1303 | Advisory | The Illuminators are currently turned off from the footswitch and cannot be turned on or off at this time. | IlluminatorMasterSwitchOffAdv The user has attempted to turn on or off an endo illuminator when the footswitch Momentary Endo Illuminators Off function is active. | • User Advisory - See Discrepancy description |
1304 | Advisory | The Laser is currently unable to deliver the maximum power level of 2 Watts. The maximum Laser power (Watts) currently available is: | LsrMaxPowerAvailableAdv Is generated when the power available from the laser drops below the maximum level (2 Watts). | • User Advisory - See Discrepancy description |
1305 | Advisory | An error occurred loading a surgeon. The system will revert to default values. | LoadSurgeonFailureAdv A database attempt to load a surgeon from the file system has failed. Reverting to defaults. | • User Advisory - See Discrepancy description |
1306 | Advisory | An error occurred setting the printer port. The system will revert to the previous value. | PrinterPortConfigErrorAdv Attempt to set the printer IP address failed. | • User Advisory - See Discrepancy description |
1307 | Advisory | An error occurred loading a procedure. The system will revert to default values. | LoadProcedureFailureAdv A database attempt to load a procedure from the file system has failed. Reverting to defaults. | • User Advisory - See Discrepancy description |
1308 | Advisory | An error occurred loading the system settings. The system will revert to default values. | LoadSystemSettingsFailureAdv A database attempt to load the system settings from the file system has failed. Reverting to defaults. | • User Advisory - See Discrepancy description |
1309 | Advisory | An error occurred saving a surgeon’s settings. Changes will be reverted at the start of the next case. | SaveSurgeonFailureAdv A database attempt to save a surgeon to the file system has failed. | • User Advisory - See Discrepancy description |
1310 | Advisory | An error occurred saving a procedure. Changes will be reverted at the start of the next case. | SaveProcedureFailure A database attempt to save a procedure from the file system has failed. Reverting to defaults. | • User Advisory - See Discrepancy description |
1311 | Advisory | An error occurred saving the system settings. The system will use default values. | SaveSystemSettingsFailureAdv A database attempt to save the system settings from the file system has failed. Reverting to defaults. | • User Advisory - See Discrepancy description |
1312 | Advisory | An error has occurred initializing Video Overlay. Video Overlay functions will be disabled. | VideoOverlayInitializationErrorAdv An error happened during VideoOverlay initialization. | • User Advisory - See Discrepancy description |
1313 | Advisory | An error has occurred in the Video Overlay component. Video Overlay functions will be disabled. | VideoOverlayErrorAdv An error happened during VideoOverlay processing. | • User Advisory - See Discrepancy description |
1314 | Advisory | There was an error saving surgeon data. | VcdSaveSurgeonAdv There was an error saving surgeon data. | • User Advisory - See Discrepancy description |
1317 | Advisory | There was an error renaming the surgeon. | VcdRenameSurgeonAdv There was an error renaming the surgeon. | • User Advisory - See Discrepancy description |
1318 | Advisory | There was an error renaming the procedure. | VcdRenameProcedureAdv There was an error renaming the procedure. | • User Advisory - See Discrepancy description |
1319 | Advisory | There was an error renaming the case data. | VcdRenameCaseAdv There was an error renaming the case data. | • User Advisory - See Discrepancy description |
1320 | Advisory | There was an error deleting the surgeon. | VcdDeleteSurgeonAdv There was an error deleting the surgeon. | • User Advisory - See Discrepancy description |
1321 | Advisory | There was an error deleting the procedure | VcdDeleteProcedureAdv There was an error deleting the procedure | • User Advisory - See Discrepancy description |
1322 | Advisory | There was an error deleting the case data. | VcdDeleteCaseAdv There was an error deleting the case data | • User Advisory - See Discrepancy description |
1323 | Advisory | There was an error loading the surgeon. | VcdLoadSurgeonAdv There was an error loading the surgeon. | • User Advisory - See Discrepancy description |
1325 | Advisory | There was an error loading the case data. | VcdLoadCaseAdv There was an error loading the case data. | • User Advisory - See Discrepancy description |
1326 | Advisory | Unable to load the case report template. | VcdLoadCaseReportTemplateAdv There was an error loading the Case Report Template. | • User Advisory - See Discrepancy description |
1329 | Advisory | Unable to save the technician’s log. | VcdSaveTechniciansLogAdv There was an error saving the Technicians Log. | • User Advisory - See Discrepancy description |
1330 | Advisory | Unable to load the video table. | VideoInfoTableReadFailureAdv There was an error loading the Video Table. | • User Advisory - See Discrepancy description |
1331 | Advisory | Unable to play the video. | VideoDfuFailureAdv There was an error loading and\or playing the Video file. | • User Advisory - See Discrepancy description |
1332 | Advisory | Unable to load Help. | HelpFileReadFailureAdv There was an error opening the Help pdf file. | • User Advisory - See Discrepancy description |
1333 | Advisory | Unable to save the case info. | SaveCaseInfoFailureAdv There was an error saving the Case Info file. | • User Advisory - See Discrepancy description |
1334 | Advisory | Unable to update the log file. | LogFileWriteFailureAdv There was an error writing to the Log File. | • User Advisory - See Discrepancy description |
1335 | Advisory | Unable to read the log file. | LogFileReadFailureAdv There was an error reading from the Log File. | • User Advisory - See Discrepancy description |
1336 | Advisory | Unable to write the incident file. | IncidentFileWriteFailureAdv There was an error writing to the Incident File. | • User Advisory - See Discrepancy description |
1337 | Advisory | Unable to read the incident file. | IncidentFileReadFailureAdv There was an error reading from the Incident File. | • User Advisory - See Discrepancy description |
1338 | Advisory | Unable to write the system metrics file. | SystemMetricFileWriteFailureAdv There was an error writing to the System Metrics File. | • User Advisory - See Discrepancy description |
1339 | Advisory | Unable to read the system metrics file. | SystemMetricFileReadFailureAdv There was an error reading from the System Metrics File. | • User Advisory - See Discrepancy description |
1340 | Advisory | Unable to write to the removable drive. | ErrorAccessingRemovableDriveAdv There was an error while attempting to write to the removable drive. | • User Advisory - See Discrepancy description |
1341 | Advisory | Further increasing the output level in air can damage fiber tips. Would you like to continue? | IllumSafeThreshold1ExceededAdv The user has exceeded Threshold 1 illuminator value by either turning on the illuminator, changing the setpoint while the illuminator is on, or changing Procedures while the illuminator is on. | • User Advisory - See Discrepancy description |
1342 | Advisory | Further increasing the output level will reduce exposure time by 35%. Would you like to continue? | IllumSafeThreshold2ExceededAdv The user has exceeded Threshold 2 illuminator value by either turning on the illuminator, changing the setpoint, or changing Procedures. | • User Advisory - See Discrepancy description |
1343 | Advisory | Further increasing the output level in air can damage fiber tips. Also, further increasing the output level will reduce exposure time by 35%. Would you like to continue? | IllumSafeThreshold1And2ExceededAdv The user has exceeded both Threshold 1 and Threshold 2 illuminator value by either turning on the illuminator, changing the setpoint while the illuminator is on, or changing Procedures while the illuminator is on. | • User Advisory - See Discrepancy description |
1344 | Advisory | Only two illuminators can be turned on simultaneously. | IllumMaxTwoIlluminatorsTurnedOnAdv User attempts to turn on an illuminator port when two ports are already on. This is not allowed. | • User Advisory - See Discrepancy description |
1345 | Advisory | AGF not allowed while Forceps or VFC is in use. | AgfNotAllowedForcepsOrVfcInUseAdv User attempts to start Auto Gas Filling when either Forceps or VFC is in use (i.e. selected and treadle down). | • User Advisory - See Discrepancy description |
1346 | Advisory | An error occurred trying to access the wireless network. | WirelessNetworkAccessAdv An error occurs when the user attempts to access the wireless network. | • User Advisory - See Discrepancy description |
1349 | Advisory | A RAID hard drive has failed or is missing | RaidDriveFailedOrMissingAdv One Disk in a redundant RAID volume is missing or has failed but the volume is still functional. | • User Advisory - See Discrepancy description |
1350 | Advisory | The probe in the current laser port is not supported. | LsrReadyModeRejectedInvalidProbeTypeAdv The type of probe in the current port is not supported by CONSTELLATION* VISION SYSTEM EH1 but might be supported by PurePoint* Laser (slit lamp for example). | • User Advisory - See Discrepancy description |
1355 | Advisory | Command is not allowed when the Pneumatics submodule is not functional. | CommandNotAllowedPneumaticsNotFunctionalAdv The user tried to invoke a command which is only allowed when the Pneumatics subsystem is functional. | • User Advisory - See Discrepancy description |
1356 | Advisory | Command is not allowed when an Anterior Only Cassette is being used. | CommandNotAllowedWithAnteriorOnlyCassetteAdv The user tried to invoke a command which is not allowed with an Anterior Only Cassette. | • User Advisory - See Discrepancy description |
1357 | Advisory | Cannot switch to Infusion: the cassette has not been primed. | CantSwitchToInfusionCassetteNotPrimedAdv The user tried to switch to Infusion when the cassette was not primed. | • User Advisory - See Discrepancy description |
1358 | Advisory | Current probe type is unrecognized. Please select a valid probe type. | IllumTurnOnRejectedUnrecognizedTypeAdv User has attempted to turn on an illuminator when the probe type is “unrecognized” | • User Advisory - See Discrepancy description |
1359 | Advisory | Power module communication error. When powering down the system, you will have to press the “Options/Shutdown” button on the screen. | PowerModuleCommFailureAdv Host controller failed to communicate with power module. | • User Advisory - See Discrepancy description |
1361 | Advisory | Command is not allowed when Infusion backup pressure is active. | The user tried to invoke a command which is only allowed when Infusion backup pressure is not active. | • User Advisory - See Discrepancy description |
1362 | Advisory | Command is not allowed when the probe type has been identified by RFID. | The user tried to change the type of a laser probe that has been identified by RFID. | • User Advisory - See Discrepancy description |
1363 | Advisory | Command is not allowed when the probe type has been previously identified by the user. Disconnect and then reconnect the probe to change its type. | The user tried to change the type of a laser probe that has been identified by a user selection. Only one selection is allowed after connection. | • User Advisory - See Discrepancy description |
1364 | Advisory | IOP Control is not allowed without a premium cassette inserted. Please insert a premium cassette and retry. | IopNotAllowedWhenUsingNonPremiumCassetteAdv The user tried to turn on IOP Control when a premium cassette is not inserted. | • User Advisory - See Discrepancy description |
1365 | Advisory | IOP Control is not allowed with an uncalibrated cassette. Please calibrate cassette and retry. | IopNotAllowedWhenCassetteNotCalibratedAdv The user tried to turn on IOP Control with a cassette that has not been calibrated. | • User Advisory - See Discrepancy description |
1366 | Advisory | Command is not allowed when FAX is on and the cassette has not been primed. | The user tried to perform a Test Instrument command when FAX is on and the cassette is not primed. | • User Advisory - See Discrepancy description |
1367 | Advisory | FAX is not allowed when the cassette is being primed. | FaxNotAllowedWhenCassetteIsPrimingAdv The user tried to turn on FAX when the cassette is being primed. | • User Advisory - See Discrepancy description |
1369 | Advisory | Setting values is not allowed when the cassette is in the chamber overflow condition. | CommandNotAllowedDuringChamberOverflowAdv The cassette chamber is in an overflow condition and pressure cannot be controlled. | • User Advisory - See Discrepancy description |
1370 | Advisory | Test instrument is not allowed in this mode. | TestInstrumentNotAllowedInThisModeAdv User has attempted to perform a test instrument command in Setup or End Case. | • User Advisory - See Discrepancy description |
1371 | Advisory | Update is only available from End Case. | UpdateOnlyAvailableInEndCaseAdv User has attempted to perform a update command in Setup or Surgery. | • User Advisory - See Discrepancy description |
1372 | Advisory | Cleaning the cassette is not allowed when the inlet pressure is out of range. Please adjust the inlet pressure between 58 psi and 120 psi. | CassetteCleanNotAllowedAirPressureOutOfRange The user attempts to start Cassette Cleaning when the Air source (wall pressure) is either below 58 psi or above 120 psi. | • User Advisory - See Discrepancy description |
1751 | Information | The remote control battery is low. | RemoteControlBatteryLow User has pressed on a remote control button and the remote’s battery is running out of power | • User Advisory - See Discrepancy description |
1752 | Information | Infusion pressure will drop to zero during calibration. Continue? | CalibrationBringsInfusionToZeroInfo User has pressed the “Calibrate” button on the advisory popup brought up when there’s been an IV Pole error. Since I/V Pole calibration will change the current Infusion pressure, the user is warned before calibration is started. | • User Advisory - See Discrepancy description |
1753 | Information | Power recovery in progress. Please wait until surgical functionality becomes available. | PowerRecoveryInProgressInfo AC power has been restored and the recovery process has started. | • User Advisory - See Discrepancy description |
1373 | Advisory | The Default Doctor can not be modified. | CommandNotAllowedDefaultDoctorAdv - The user attempts to save, modify or add a procedure when the default doctor is selected. | • User Advisory - See Discrepancy description |
1374 | Advisory | The Accurus* Classic procedure can not be modified. | CommandNotAllowedAccurusClassicAdv - The user attempts to modify an Accurus Classic procedure. | • User Advisory - See Discrepancy description |
1375 | Advisory | This command is not available in End Case. | CommandNotAllowedInEndCaseAdv - The user attempts to change either the doctor or procedure in End Case. | • User Advisory - See Discrepancy description |
1376 | Advisory | This command is not available when the laser is firing. | CommandNotAllowedLaserFiringAdv The user attempt to reset the shot count and energy metrics when the Laser is firing. | • User Advisory - See Discrepancy description |
1377 | Advisory | No changes have been made to the current procedure. | CommandNotAllowedNoProcChanges The user attempts to save a Procedure when no procedure changes have been made. | • User Advisory - See Discrepancy description |
1378 | Advisory | This command is not available when system is in the surgery screen. | CommandNotAllowedInSurgeryMode The command is not available when system is in the Surgery screen. | • User Advisory - See Discrepancy description |
1379 | Advisory | Connected vitreous probe is not valid. The probe will be used as a low-speed probe. | VitProbeInvalidAdv A connected vitreous probe has valid checksum but one or more invalid parameters. | • User Advisory - See Discrepancy description |
1380 | Advisory | The CPU battery is bad. Please contact Field Service. | CPUBatteryBadAdv | • User Advisory - See Discrepancy description |
1381 | Advisory | Unable to copy to the removable drive. | RemovableDriveCopyErrorAdv | • User Advisory - See Discrepancy description |
1382 | Advisory | Unable to delete from the removable drive. | RemovableDriveDeleteErrorAdv | • User Advisory - See Discrepancy description |
1383 | Advisory | Command is not allowed when the inlet pressure is out of range. Please adjust the inlet pressure between 58 psi and 120 psi. | CommandNotAllowedAirPressureOutOfRange The user tried to invoke a command when the Air source (wall pressure) is either below 58 psi or above 120 psi. | • User Advisory - See Discrepancy description |
1384 | Advisory | Ensure proper scissors tip attachment to pneumatic handpiece. Prior to use in the eye, depress foot pedal\button to ensure proper tip function. | TestScissorsTipAttachmentAdv User has chosen to use scissors and is reminded to confirm attachment and functionality before use. | • User Advisory - See Discrepancy description |
1385 | Advisory | Ensure proper forceps tip attachment to pneumatic handpiece. Prior to use in the eye, depress foot pedal to ensure proper tip function. | TestForcepsTipAttachmentAdv User has chosen to use forceps and is reminded to confirm attachment and functionality before use. | • User Advisory - See Discrepancy description |
1386 | Advisory | Fluidics submodule recovery unsuccessful. Fluidics functions will remain disabled. | Fluidics submodule recovery unsuccessful. Fluidics functions will remain disabled. | • User Advisory - See Discrepancy description |
1387 | Advisory | US Diathermy submodule recovery unsuccessful. US Diathermy functions will remain disabled. | US Diathermy submodule recovery unsuccessful. US Diathermy functions will remain disabled. | • User Advisory - See Discrepancy description |
1388 | Advisory | TableTop Illuminator submodule recovery unsuccessful. TableTop Illuminator functions will remain disabled. | Displayed after the Table Top Illuminator submodule has unsuccessfully recovered from an error. | • User Advisory - See Discrepancy description |
1389 | Advisory | Pneumatics submodule recovery unsuccessful. Pneumatics functions will remain disabled. | Displayed after the Pneumatics submodule has unsuccessfully recovered from an error. | • User Advisory - See Discrepancy description |
1390 | Advisory | Auxiliary Illuminator submodule recovery unsuccessful. Auxiliary Illuminator functions will remain disabled. | Displayed after the Auxiliary Illuminator submodule has unsuccessfully recovered from an error. | • User Advisory - See Discrepancy description |
1391 | Advisory | Laser submodule recovery unsuccessful. Laser functions will remain disabled. | LaserSubmoduleErrorRecoveryUnsuccessfulAdv Displayed after the Laser submodule has unsuccessfully recovered from an error. | • User Advisory - See Discrepancy description |
1392 | Advisory | Command not allowed without a calibrated AutoSert* handpiece. | IolCmdNotAllowedHandpieceNotReadyAdv The user pressed the treadle in an AutoSert* Step when no AutoSert* handpiece was connected or the handpiece was connected but not calibrated. | • User Advisory - See Discrepancy description |
1394 | Advisory | The AutoSert* handpiece is already prepared. The command is not allowed. | IolCmdNotAllowedHandpiecePreparedAdv User pressed the “Load Plunger” or “Preload IOL” button when the AutoSert* handpiece was “Prepared” | • User Advisory - See Discrepancy description |
1395 | Advisory | The current AutoSert* operation was cancelled. | IolCmdCancelledEnteringEndCaseAdv User selected End Case state when one of the following AutoSert* setup commands were in progress: • Load Plunger • Preload IOL | • User Advisory - See Discrepancy description |
1396 | Advisory | Preload IOL command not allowed until the AutoSert* plunger has been fully retracted. | IolCmdNotAllowedLoadPlungerStartedAdv User pressed the [Preload IOL] button when a Load Plunger command was previously started but not yet completed (and currently paused). | • User Advisory - See Discrepancy description |
1397 | Advisory | Load Plunger command not allowed until the AutoSert* plunger has been fully retracted. | IolCmdNotAllowedPreloadIolStartedAdv User pressed the [Load Plunger] button when a Preload IOL command was previously started but not yet completed (and currently paused). | • User Advisory - See Discrepancy description |
1399 | Advisory | Function is not allowed when the IOL Injection submodule is not functional. | CommandNotAllowedIOLInjectionNotFunctionalAdv The user tried to invoke a command which is only allowed when the IOL Injection subsystem is functional. | • User Advisory - See Discrepancy description |
1401 | Advisory | 20 cc of undiluted gas has been dispensed. After removing the syringe from the system, adjust the plunger to obtain the desired gas concentration before injection. | AgfUndilutedGasAdv Warns user that there are 20 cc of undiluted gas in the syringe. | • User Advisory - See Discrepancy description |
1402 | Advisory | Fluidics recovery complete. Cassette status not fully restored. | FluidicsSubmoduleRecoveryPartialSuccessAdv The Fluidics submodule recovered, but the system could not fully restore the state due to missing instruments (no valid cassette reported) | • User Advisory - See Discrepancy description |
1403 | Advisory | Table Top Illuminator recovery complete. Illuminator probe status not fully restored. | The Table Top Illuminator load module recovered, but the system could not fully restore the state due to missing instruments (illuminator probe(s) not detected) | • User Advisory - See Discrepancy description |
1404 | Advisory | Pneumatics recovery complete. Vitreous probe status not fully restored. | PneumaticsSubmoduleRecoveryPartialSuccessAdv The Pneumatics submodule recovered, but the system could not fully restore the state due to missing instruments (vitreous probe not detected) | • User Advisory - See Discrepancy description |
1405 | Advisory | Aux Illuminator recovery complete. Illuminator probe status not fully restored. | AuxIllumSubmoduleRecoveryPartialSuccessAdv The Auxiliary Illuminator submodule recovered, but the system could not fully restore the state due to missing instruments (illuminator probe(s) not detected) | • User Advisory - See Discrepancy description |
1406 | Advisory | Laser recovery complete. Laser probe status not fully restored. | LaserSubmoduleRecoveryPartialSuccessAdv The Laser submodule recovered, but the system could not fully restore the state due to missing instruments (laser probe(s) not detected). | • User Advisory - See Discrepancy description |
1407 | Advisory | Fault recovery complete. AutoSert* Handpiece status not fully restored. | IolFunctionalityNotRestoredAfterFaultRecoveryAdv An IOL handpiece was inserted when a Fault occurred. After recovery the system will not restore the handpiece state and the user is notified via this advisory. | • User Advisory - See Discrepancy description |
2100 | Fault | Fault - 2100 Call Field Service | SUP_HOST_HEARTBEAT_TIMEOUT – A two second heart beat reply from the host was missed. | • Cable W17 • Host Module • Supervisor |
2200 | Error | Communications failure with the Fluidics submodule. Fluidics functions will be disabled. | SUP_FLUIDICS_HUNT_FAIL_ERR - The supervisor cannot establish communicate with the Fluidics submodule. | • Cable W24 • Supervisor • Fluidics Controller PCB |
2201 | Error | Communications failure with the Fluidics submodule. Fluidics functions will be disabled. | SUP_FLUIDICS_COMM_FAIL_ERR The supervisor lost communicate with the Fluidics submodule | • Cable W24 • Supervisor • Fluidics Controller PCB |
2202 | Advisory | Unable to perform this function without a primed cassette. | SUP_CASSETTE_NOT_READY_ADV - This advisory is produced when extraction is attempted without a primed cassette. | • User Advisory - See Discrepancy description |
2203 | Advisory | Unable to aspirate. Please turn on infusion. | SUP_INFUSION_IRRIGATION_OFF_ADV - This advisory is produced when extraction is attempted without infusion on. | • User Advisory - See Discrepancy description |
2204 | Advisory | Unable to aspirate while infusion/irrigation is unavailable. | SUP_INFUSION_UNAVAILABLE_ADV - This advisory is produced when extraction is attempted without infusion being functional. | • User Advisory - See Discrepancy description |
2205 | Advisory | Please wait: draining cassette. | SUP_EXT_CHMBR_OVER_ADV - This advisory is produced when extraction is attempted when the extraction chamber is in a overflow condition. | • User Advisory - See Discrepancy description |
2206 | Advisory | Unable to turn on infusion without a primed cassette. | SUP_INF_CASSETTE_NOT_READY_ADV - This advisory is produced when infusion is attempted with out a primed cassette. | • User Advisory - See Discrepancy description |
2207 | Advisory | Unable to turn on infusion without a calibrated IV pole. | SUP_INF_IV_POLE_NOT_CALIB_ADV - This advisory is produced when infusion is attempted with out a calibrated IV Pole. | • User Advisory - See Discrepancy description |
2208 | Advisory | Unable to turn on infusion without sufficient source pressure. | SUP_INF_NO_AIR_PRESSURE_ADV - This advisory is produced when infusion is attempted with out source air pressure | • User Advisory - See Discrepancy description |
2209 | Advisory | Unable to turn on irrigation without a primed cassette. | SUP_IRR_CASSETTE_NOT_READY_ADV - This advisory is produced when irrigation is attempted with out a primed cassette. | • User Advisory - See Discrepancy description |
2211 | Advisory | Unable to turn on irrigation without sufficient source pressure. | SUP_IRR_NO_AIR_PRESSURE_ADV - This advisory is produced when irrigation is attempted with out suitable source pressure. | • User Advisory - See Discrepancy description |
2212 | Advisory | Irrigation unavailable: out of fluid. Please change the bottle. | SUP_IRR_NO_FLUID_ADV - This advisory is produced when irrigation is attempted with out sufficient irrigation fluid available | • User Advisory - See Discrepancy description |
2213 | Advisory | FAX unavailable. Please insert a cassette. | SUP_FAX_NO_CASSETTE_ADV - This advisory is produced when FAX is attempted with out a cassette inserted. | • User Advisory - See Discrepancy description |
2214 | Error | Extraction setpoint timeout. Infusion/Irrigation and Extraction functions will be disabled. | SUP_EXT_NO_ACTIVE_UPDATE_ERR – This error is produced when the extraction is active and the extraction proxy fails to get a footswitch update within the required timeframe. | • Supervisor • Footswitch • Fluidics Controller PCB |
2215 | Advisory | Micro reflux is currently not available. | SUP_EXT_NO_REFLUX_FLUID_ADV – This error is produced when the user attempts to activate micro reflux and the fluid level is out of range. | • User Advisory - See Discrepancy description |
2216 | Advisory | Please change the Bottle to regain extraction. | SUP_EXT_NO_INFUSE_FLUID_ADV – This error is generated when aspiration is attempted when the infusion in the cassette is empty | • User Advisory - See Discrepancy description |
2217 | Advisory | Extraction is not allowed while in 30 mmHg backup mode. | SUP_EXT_INFUSE_IN_BACKUP_ADV – This advisory occurs when Extraction is attempted while in manual backup. | • User Advisory - See Discrepancy description |
2218 | Advisory | Irrigation is not allowed while in 30 mmHg backup mode. | SUP_IRR_INFUSE_IN_BACKUP_ADV – This advisory occurs when Irrigation is attempted while in manual backup. | • User Advisory - See Discrepancy description |
2250 | Error | Communications failure with the Pneumatics submodule. Pneumatics functions will be disabled. | SUP_PNEUMATICS_HUNT_FAIL_ERR – The supervisor cannot communicate with the Pneumatics submodule. | • Cable W21 • Supervisor • Pneumatic Controller |
2251 | Error | Communications failure with the Pneumatics submodule. Pneumatics functions will be disabled. | SUP_PNEUMATICS_COMM_FAIL_ERR – This error is produced when the Supervisor losses communication with the pneumatics module | • Cable W21 • Supervisor • Pneumatic Controller |
2252 | Advisory | Unable to enable function without sufficient source pressure. | SUP_PNEU_NO_AIR_PRESSURE_ADV - This advisory is produced when cutter or utility operation is requested with out sufficient source pressure. | • User Advisory - See Discrepancy description |
2253 | Advisory | Unable to enable function with excessive source pressure. | SUP_PNEU_HIGH_AIR_PRESSURE_ADV - This advisory is produced when cutter or utility operation is requested with source pressure that is too high. | • User Advisory - See Discrepancy description |
2254 | Error | Pneumatics setpoint timeout. Pneumatics functions will be disabled. | SUP_PNEU_NO_ACTIVE_UPDATE_ERR - This error is produced when the pneumatics is active and the pneumatics proxy fails to get a footswitch update within the required timeframe. | • Supervisor • Footswitch • Pneumatics Controller PCB |
2300 | Error | Communications failure with the Ultrasound submodule. Ultrasound functions will be disabled. | SUP_ULTRASOUND_HUNT_FAIL_ERR – Thesupervisor cannot communicate with the Ultrasound submodule. | • Cable W20 • Supervisor • U/S Controller PCB |
2301 | Error | Communications failure with the Ultrasound submodule. Ultrasound functions will be disabled. | SUP_ULTRASOUND_COMM_FAIL_ERR - The supervisor lost communication with the Ultrasound submodule | • Cable W20 • Supervisor • U/S Controller PCB |
2302 | Error | US setpoint timeout. US, IOL injector and Diathermy functions will be disabled. | SUP_US_NO_ACTIVE_UPDATE_ERR - This error is produced when the US is active and the US proxy fails to get a footswitch update within required timeframe. | • Supervisor • Footswitch • Fluidics Controller PCB |
2350 | Error | Communications failure with the Laser submodule. Laser functions will be disabled. | SUP_LASER_HUNT_FAIL_ERR - The supervisor cannot communicate with the Laser submodule. | • Communication Cable • Supervisor • Laser Core Module |
2351 | Advisory | Communications failure with the Laser submodule. Laser functions will be disabled. | SUP_LASER_COMM_FAIL_ADV - The supervisor lost communication with the Laser submodule | • This message is normally displayed whenever the key is turned off. If displayed any other time check: • Communication Cable • Supervisor • Laser Core Module |
2400 | Advisory | Please insert the Table Top Illuminator drawer. | SUP_ILLUM_DRAWER_OUT_ADV – Supervisor detected a non –zero setpoint with the Table Top Illuminator drawer out | • User Advisory - See Discrepancy description |
2401 | Advisory | The lamp in the Table Top Illuminator needs to be replaced. Please contact Field Service. | SUP_TT_ILLUM_LAMP_BAD_ADV - Supervisor detected a non–zero setpoint with the Table Top Illuminator Lamp bad. | • User Advisory - See Discrepancy description |
2500 | Advisory | Communications failure with the Power Control submodule. Please contact Field Service. | SUP_POWER_MONITOR_COMM_FAIL_ERR – Supervisor detected a serial I/O error when trying to talk to the Power Module | • Reset needed for Power Controller - Use PCB reset switch or press and hold in Standby switch to power down • W48 Cable Assy • Power Controller PCB • Supervisor |
2550 | Error | Footswitch error. Footswitch treadle functions will be disabled. | SUP_FOOTSWITCH_ERR – The footswitch has an error. | • Footswitch • Supervisor • Cable W22 • Cable W01 (Base Footswitch Cable) |
2700 | Error | Communications failure with the Auxiliary Illuminator submodule. Auxiliary Illuminator functions will be disabled. | SUP_AUX_ILLUMINATOR_HUNT_FAIL_ERR – The supervisor cannot communicate with the Auxiliary Illuminator submodule. | • Supervisor • Aux illuminator PCB |
2701 | Advisory | Communications failure with the Auxiliary Illuminator submodule. Auxiliary Illuminator functions will be disabled. | SUP_AUX_ILLUM_COMM_FAIL_ADV - The supervisor lost communication with the Auxiliary Illuminator submodule | • Supervisor • Aux illuminator PCB |
2702 | Advisory | The lamp in the Auxiliary Illuminator needs to be replaced. Please contact Field Service. | SUP_AUX_ILLUM_LAMP_BAD_ADV - Supervisor detected a non–zero setpoint with the Auxiliary Illuminator Lamp bad. | • User Advisory - See Discrepancy description |
2750 | Fault | Fault - 2750 Call Field Service | SUP_RESET_FAIL_ERR – The Supervisor could not assert reset control over all submodules. A module that was supposed to be reset responded with its module information. | • Supervisor • Module not responding |
2751 | Fault | Fault - 2751 Call Field Service | SUP_SLOT_ID_ERR – The Supervisor got duplicate slot ID for the submodules. | • Supervisor • Module not responding |
3100 | Advisory | Could not calibrate the fluid level sensors. Please eject and reinsert, or replace the cassette. | FLD_CMN_LEVEL_SENSOR_CAL_ERR - Level Sensor calibration failed. Possible reasons: 1) the calibration done status bit was not set by the FPGA within the specified timeout limit 2) the calculated required max/min pixel gain was outside the valid range 3) the brightness level of the level sensor LEDs could not be adjusted. | • Cassette • Level Sensor • Fluidics Controller PCB • Fluidics Module |
3200 | Error | Cassette latch error. Infusion/irrigation and extraction functions will be disabled. | CASSETTE_LATCH_ERR - Cassette latch hardware feedback signals indicate a hardware problem. | • Cassette • Mechanical interference with latch mechanism • Latch Cylinder • Fluidics Controller PCB • Fluidics Module |
3201 | Advisory | Invalid cassette ID detected. Please eject and insert a valid cassette. | CASSETTE_ID_INVALID_ERR - An invalid cassette id was read by the cassette id sensors. | • Cassette • Cassette ID Sensors • Suction PCB • Fluidics Controller PCB • Fluidics Module |
3202 | Advisory | Cassette test failed. Please eject and reinsert, or replace cassette. | CASSETTE_TEST_FAILED_ERR Cassette pressure and/or vacuum tests failed. | • Cassette • Vacuum or Pressure leak • Fluidics Controller • FluidIcs Module |
3203 | Advisory | The cassette was not properly latched into position. Please remove and reinsert the cassette. | CASSETTE_NOT_LATCHED_ADV The cassette latch optical position sensor indicates that the latch did not reach its locked position. | • Cassette • Mechanical interference with latch mechanism • Latch Cylinder • Fluidics Controller PCB • Fluidics Module |
3204 | Error | Cassette ID sensor error. Infusion/irrigation and extraction functions will be disabled. | CASSETTE_ID_SENSOR_ERR Cassette ID sensor test failed. The sensor output voltage is not within the expected range. | • Cassette • Cassette ID Sensors • Suction PCB • Fluidics Controller PCB • Fluidics Module |
3205 | Advisory | The Fluidics module fan is not working. | FLD_MODULE_FAN_TAC_ADV The Fluidics module fan tachometer indicates that the Fluidics module fan is not operating. | • Fluidics Controller PCB • Fan • Fluidics Module |
3207 | Information | Cassette cannot be loaded. Please adjust the inlet pressure between 58 psi and 120 psi. | CASS_NO_AIR_PRSR_INFO A cassette was inserted in the receiver mechanism while the wall pressure is outside its valid operating range. | • User Information - See Discrepancy description |
3300 | Advisory | Draining cassette. Please wait. | EXT_CHAMBER_OVERFLOW_ADV The cassette aspiration chamber is full of fl | • Cassette not draining to drain bag • Fluid Level Sensors • Fluidics Controller PCB • Fluidics Module |
3302 | Advisory | Fault - 3302 Call Field Service | EXT_NOT_IN_OP_RANGE_ADV Not enough or too much fluid in the aspiration chamber to allow aspiration flow mode and reflux. NOTE: This advisory is never explicitly displayed. Instead advisory 1239 or 1240 might be displayed if appropriate. See 1239 and 1240 for details. | • User Advisory - See Discrepancy description |
3304 | Advisory | Leak test failure. Please confirm the irrigation tubing, aspiration tubing, and the test chamber are properly connected to the handpiece. | EXT_HP_TUBING_LEAK_ADV A leak in the aspiration or irrigation tubing was detected during flow check of a handpiece. | • User Advisory - See Discrepancy description |
3305 | Advisory | Priming of the aspiration probe was unsuccessful. Please attempt to re-prime the probe. | EXT_PROBE_PRIME_FAILED_ADV Probe prime was unsuccessful. The required volume of fluid was not transferred through the aspiration tubing set within the specified timeout period. | • User Advisory - See Discrepancy description |
3306 | Advisory | Priming of the aspiration handpiece was unsuccessful. Please attempt to re-prime the handpiece. | EXT_HP_PRIME_FAILED_ADV Handpiece prime was unsuccessful. The required volume of fluid was not transferred through the aspiration tubing set within the specified timeout period. | • User Advisory - See Discrepancy description |
3307 | Error | Aspiration level sensor problem detected. Infusion/ irrigation and extraction functions will be disabled. | EXT_LVL_STATUS_ERR The level sensor status signals indicate a level sensor hardware problem. | • Aspiration Level Sensor • Cable W93 • Fluidics Controller PCB • Fluidics Module |
3308 | Advisory | Flow check failure: measured flow restriction is too high. Extraction and Ultrasound functions in Phaco/Frag will be disabled. | EXT_FLOW_CHECK_FAILED_ADV The handpiece failed the flowcheck. Too high vacuum level was required to achieve the reference flow level. | • User Advisory - See Discrepancy description |
3309 | Advisory | Flow check failure: aspiration chamber could not be filled with fluid. Extraction and Ultrasound functions in Phaco/Frag will be disabled. | EXT_FILL_TIMEOUT_ADV The handpiece failed the flowcheck. Too high vacuum level was required to achieve the reference flow level. | • User Advisory - See Discrepancy description |
3325 | Advisory | Drain bag is almost full. Please replace bag and press [Done]. | EXT_BAG_NEAR_FULL_ADV ~400 cc of fluid has been pumped into the drain bag. | • User Advisory - See Discrepancy description |
3326 | Advisory | Drain bag is full. Please replace bag and press [Done]. | EXT_BAG_FULL_ADV - ~450 cc of fluid has been pumped into the drain bag. | • User Advisory - See Discrepancy description |
3327 | Advisory | Drain bag is critically full. Please replace bag and press [Done]. | EXT_BAG_CRITICALLY_FULL_ADV ~500 cc of | • User Advisory - See Discrepancy description |
fluid has been pumped into the drain bag. | ||||
3329 | Error | Drain pump problem detected. Infusion/irrigation and extraction functions will be disabled. | EXT_PUMP_ROTATION_ERR The commanded pump rate does not correspond to the actual pump rate measured by the optical encoder. | • Drain Pump • Fluidics Controller PCB • Fluidics Module |
3330 | Advisory | The drain pump fan is not working. | EXT_PUMP_FAN_STATUS_ERR The drain pump fan tachometer indicates that the drain pump fan is not operating. | • Fan • Fluidics Controller PCB • Fluidics Module |
3331 | Advisory | Could not drain the aspiration chamber. Please remove and insert a new cassette. | EXT_CHAMBER_DRAIN_TIMEOUT_ADV The | • Cassette • Drain Pump • Aspiration Level Sensor • Fluidics Controller PCB • Fluidics Modu |
3350 | Error | Extraction pressure transducer offset error. Infusion/irrigation and extraction functions will be disabled. | EXT_PT_OFFS_ERR Extraction pressure transducer 0 offset out of range. | • Cable W86 • Fluidics Controller PCB • Suction PCB • Transducer MT7 and/or MT8 • Fluidics Module |
3351 | Error | Extraction pressure transducer discrepancy error. Infusion/irrigation and extraction functions will be disabled. | EXT_PT_DISCR_ERR - A discrepancy between the primary and redundant extraction pressure transducers was detected. | • Transducer MT7 and/or MT8 • Fluidics Controller PCB • Suction PCB • Cable W86 • Fluidics Module |
3352 | Error | Extraction isolation valve error. Infusion/irrigation and extraction functions will be disabled. | EXT_ISO_VLV_ERR Extraction isolation valve hardware error. | • L14 • Fluidics Controller PCB • Suction PCB • Cable W86 • Fluidics Module |
3353 | Error | Reflux valve error. Infusion/irrigation and extraction functions will be disabled. | EXT_REFLUX_VLV_ERR Extraction reflux valve hardware error. | • L15 • Fluidics Controller PCB • Suction PCB • Cable W86 • Fluidics Module |
3354 | Error | Extraction output valve (S1) error. Infusion/ irrigation and extraction functions will be disabled. | EXT_OUTPUT_VLV_ERR Extraction output pincher valve (S1) error | • 5 Valve PCB, Yellow Lower • Cable W90 • Fluidics Controller PCB • Fluidics Module |
3355 | Error | Extraction output valve (S11) error. Infusion/ irrigation and extraction functions will be disabled. | EXT_PORT1_VLV_ERR Extraction output port 1 pincher valve error. | • 3 Valve PCB, Green • Cable W90 • Fluidics Controller PCB • Fluidics Module |
3356 | Error | Extraction output valve (S22) error. Infusion/ irrigation and extraction functions will be disabled. | EXT_PORT2_VLV_ERR Extraction output port 2 pincher valve error. | • 3 Valve PCB, Green • Cable W90 • Fluidics Controller PCB • Fluidics Module |
3357 | Error | Extraction cross-connection valve (SC) error. Infusion/irrigation and extraction functions will be disabled. | EXT_CROSS_NO_VLV_ERR Extraction “normally open” cross-connection valve error. | • 5 Valve PCB, Red Upper • Cable W90 • Fluidics Controller PCB • Fluidics Module |
3358 | Error | Extraction cross-connection valve (SC2) error. Infusion/irrigation and extraction functions will be disabled. | EXT_CROSS_NC_VLV_ERR Extraction “normally closed” cross-connection valve error. | • 5 Valve PCB, Red Upper • Cable W90 • Fluidics Controller PCB • Fluidics Module |
3359 | Advisory | Suction pressure surges detected. Vacuum will be disabled. Please release the footswitch treadle to reset. | EXT_PRSR_OSCILLATION_ADV Extraction pressure oscillations detected. | • Bubbles in tubing - Check Cassette connections and/or reprime • Transducers MT7 & MT8 • Fluidics Controller PCB • Suction PCB • Cable W86 • Fluidics Module |
3360 | Advisory | Suction flow surges detected. Flow will be disabled. Please release the footswitch treadle to reset. | EXT_FLOW_OSCILLATION_ADV Extraction flow oscillations detected. | • Bubbles in tubing - Check Cassette connections and/or reprime • Transducers MT7 & MT8 • Fluidics Controller PCB • Suction PCB • Cable W86 • Fluidics Module |
3361 | Advisory | Suction pressure is too high. Vacuum will be disabled. Please release the footswitch treadle to reset. | EXT_PRSR_OVERSHOOT_ADV Extraction pressure overshoot detected. | • Bubbles in tubing - Check Cassette connections and/or reprime • Transducers MT7 & MT8 • Fluidics Controller PCB • Suction PCB • Cable W86 • Fluidics Module |
3362 | Advisory | Aspiration flow too high. Flow will be disabled. Please release the footswitch treadle to reset. | EXT_FLOW_OVERSHOOT_ADV Extraction flow overshoot detected. | • Bubbles in tubing - Check Cassette connections and/or reprime • Fluidics Controller PCB • Suction PCB • Cable W86 • Fluidics Module |
3363 | Error | Extraction pressure transducer reference voltage out of range. Infusion/irrigation and extraction functions will be disabled. | EXT_TRANSDUCER_REFERENCE_ERR Extraction transducer reference voltage error. | • Transducers MT7 & MT8 • Fluidics Controller PCB • Fluidics Module |
3364 | Advisory | Aspiration manifold input pressure is too high. | EXT_75_PSI_REGULATOR_ADV Extraction 75 psi pressure regulator advisory. The aspiration manifold 75 psi regulator is out of regulation. A pressure higher than 220 mmHg was measured during the regulator pressure test which is part of the cassette test. | • Relief Valve • Transducers MT7 & MT8 • Fluidics Controller PCB • Fluidics Module |
3400 | Error | Infusion source container pressure transducer offset error. Infusion/irrigation and extraction functions will be disabled. | PRSR_SRC_PT_OFFS_ERR Infusion source container pressure transducer 0 offset out of range. | • Fluidics Infusion PCB • Cable W88 • Fluidics Controller PCB • Fluidics Module |
3401 | Error | Infusion source container isolation valve error. Infusion/irrigation and extraction functions will be disabled. | PRSR_SRC_ISO_VLV_ERR Infusion source container isolation valve error. | • Infusion 5 Valve PCB • Fluidics Infusion PCB • Cable W88 • Fluidics Controller PCB • Fluidics Module |
3402 | Error | Infusion source container transducer discrepancy error. Infusion/irrigation and extraction functions will be disabled. | PRSR_SRC_PT_DISCR_ERR A dispcrepancy between the source container and LPAS pressure transducers was detected. | • Fluidics Infusion PCB • Cable W88 • Fluidics Controller PCB • Fluidics Module |
3403 | Error | Infusion source container pressure too high. Infusion/irrigation and extraction functions will be disabled. | PRSR_SRC_HI_ERR Infusion source container pressure too high. | • Proportional Valve - PV3 or PV6 • Fluidics Infusion PCB • Cable W88 • Fluidics Controller PCB • Fluidics Module |
3420 | Error | Infusion pressure transducer offset error. Infusion/ irrigation and extraction functions will be disabled. | INF_PT_OFFS_ERR Infusion pressure transducer 0 offset out of range | • Cable W88 • Fluidics Infusion PCB • Fluidics Controller PCB • Fluidics Module |
3421 | Error | Infusion pressure transducer discrepancy error. Infusion/irrigation and extraction functions will be disabled. | INF_PT_DISCR_ERR A dispcrepancy between the primary and redundant infusion pressure transducers was detected. | • Cable W88 • Fluidics Infusion PCB • Fluidics Controller PCB • Fluidics Module |
3422 | Error | Infusion isolation valve error. Infusion/irrigation and extraction functions will be disabled. | INF_ISO_VLV_ERR Infusion isolation valve error. | • Infusion 5 Valve PCB • Cable W88 • Fluidics Infusion PCB • Fluidics Controller PCB • Fluidics Modu |
3423 | Error | Infusion FAX valve error. Infusion/irrigation and extraction functions will be disabled. | INF_FAX_VLV_ERR Infusion FAX valve error. | • Infusion 5 Valve PCB • Cable W88 • Fluidics Infusion PCB • Fluidics Controller PCB • Fluidics Modu |
3424 | Error | Infusion input valve error. Infusion/irrigation and extraction functions will be disabled. | INF_INPUT_VLV_ERR Infusion chamber input pincher valve error. | • 5 Valve PCB, Red Upper • Cable W90 • Fluidics Controller PCB • Fluidics Module |
3425 | Error | Infusion output valve error. Infusion/irrigation and extraction functions will be disabled. | INF_OUTPUT_VLV_ERR Infusion chamber output pincher valve error. | • 5 Valve PCB, Red Upper • Cable W90 • Fluidics Controller PCB • Fluidics Module |
3426 | Error | Irrigation output valve error. Infusion/irrigation and extraction functions will be disabled. | IRR_OUTPUT_VLV_ERR Irrigation output pincher valve error. | • Cassette (fluid can interfere with sensor • detection and cylinder movement) • CY4-CO Cylinder • Optical position sensor • 3 Valve PCB, Green • Cable W90 • Fluidics Controller PCB • Low pressure or pressure leak. • Fluidics Module |
3427 | Error | Infusion pressure surges detected. Infusion/ irrigation and extraction functions will be disabled. | INF_OSCILLATION_ERR Infusion pressure oscillations detected. | • Infusion Proportional Valve • Fluidics Infusion PCB • Cable W88 • Fluidics Controller PCB • Fluidics Module |
3428 | Error | Infusion pressure too high. Infusion/irrigation and extraction functions will be disabled. | INF_HI_PRSR_ERR Infusion high pressure detected. | • Infusion Proportional Valve • Fluidics Infusion PCB • Cable W88 • Fluidics Controller PCB • Fluidics Module |
3429 | Advisory | Low infusion pressure detected. Please check infusion connections. Select 30 mmHg liquid backup pressure, or ignore low pressure condition. | INF_LO_PRSR_ADV Infusion pressure too low. | • User Advisory - See Discrepancy description |
3431 | Advisory | Low irrigation pressure detected. Please check irrigation connections. | IRR_LO_PRSR_ADV Irrigation pressure too low. | • User Advisory - See Discrepancy description |
3433 | Error | Infusion chamber isolation valve error. Infusion/ irrigation and extraction functions will be disabled. | INF_ISO_VLV_STUCK_CLOSED_ERR Infusion chamber isolation valve detected to be stuck closed during power up diagnostics. | • Infusion 5 Valve PCB • Cable W88 • Fluidics Infusion PCB • Fluidics Controller PCB • Fluidics Module |
3434 | Advisory | Infusion chamber overflow error. Control of infusion pressure has possibly been lost. The current pressure, in mmHg, could be as high as: | INF_CHAMBER_OVERFLOW_ADV Infusion chamber overflow error. The fluid level in the infusion chamber reached above the overflow level. | • User Advisory - See Discrepancy description |
3436 | Error | FAX valve error. Infusion/irrigation and extraction functions will be disabled. | INF_FAX_VLV_STUCK_CLOSED_ERR FAX valve detected to be stuck closed during power up diagnostics. | • Infusion 5 Valve PCB • Cable W88 • Fluidics Infusion PCB • Fluidics Controller PCB • Fluidics Module |
3437 | Error | Infusion level sensor error. Infusion/irrigation and extraction functions will be disabled. | INF_LVL_STATUS_ERR The level sensor status signals indicate a level sensor hardware problem | • Infusion Level Sensor • Cable W94 • Fluidics Controller PCB • Fluidics Module |
3438 | Advisory | No more infusion fluid available. Press [Change] to change the fluid container. | INF_FLUID_OUT_ADV The infusion chamber in the cassette is empty. | • User Advisory - See Discrepancy description |
3440 | Advisory | No more irrigation fluid available. Press [Change] to change the fluid container.. | IRR_FLUID_OUT_ADV The irrigation chamber in the cassette is empty. | • User Advisory - See Discrepancy description |
3442 | Advisory | Irrigation chamber overflow detected. Control of irrigation pressure has possibly been lost. The current pressure, in mmHg, could be as high as: | IRR_CHAMBER_OVERFLOW_ADV Irrigation chamber overflow error. The fluid level in the irrigation chamber reached above the overflow level. | • User Advisory - See Discrepancy description |
3444 | Advisory | Low infusion pressure. Please check connections. Select 30 mmHg air backup, or ignore low pressure condition. Backup mode will not allow extraction. | INF_FAX_MS_LO_PRSR_ADV Infusion pressure too low while in FAX with a Manual Stopcock cassette. | • User Advisory - See Discrepancy description |
3460 | Error | Infusion backup valve error. Infusion/irrigation, extraction, and Ultrasound functions will be disabled. | INF_BACKUP_VLV_ERR Infusion backup valve error. | • Infusion 5 Valve PCB • Cable W88 • Fluidics Infusion PCB • Fluidics Controller PCB • Fluidics Module |
3461 | Error | Infusion LPAS pump error detected. Infusion/ irrigation, extraction, and Ultrasound functions will be disabled. | INF_LPAS_TAC_ERR The infusion LPAS pump tachometer indicate that the LPAS pump is not operating correctly. | • LPAS Pump • Fluidics Controller PCB • Fluidics Module |
3462 | Advisory | Infusion flow sensor communication error. IOP Control functions will be disabled. | INF_FLOW_SENSOR_COM_ERR Infusion flow sensor communication error. | • NIFS Flow Sensor • Fluidics Controller PCB • Fluidics Module |
3464 | Advisory | Infusion flow data invalid: IOP Control functions will be disabled. Check infusion tubing for air bubbles. | INF_FLOW_DATA_INVALID_ADV Infusion flow sensor readings are not valid. | • Bubbles • Cassette Connections • Flow Sensor • Fluidics Controller PCB • Fluidics Module |
3466 | Advisory | Infusion flow sensor accuracy error. IOP Control functions will be disabled. | INF_FLOW_ACCURACY_ADV Infusion flow sensor accuracy error. Flow sensor and chamber fluid volume measurements are out of range of each other. Detected during flow sensor calibration. | • Cassette • Infusion Level Sensor • Cable W94 • Flow Sensor • Flow PCB • Fluidics Controller PCB • Fluidics Module |
3467 | Advisory | Infusion flow sensor signal amplitude is low. IOP Control functions will be disabled. Please eject and reinsert the cassette. | INF_FLOW_SIGNAL_AMPLITUDE_ADV Infusion flow sensor signal amplitude error. Detected during flow sensor calibration | • Wrong infusion cannula selected • Cassette • Flow Sensor • Flow PCB • Fluidics Controller PCB • Fluidics Module |
3469 | Advisory | Tubing calibration offset error. IOP Control functions will be disabled. Please position the infusion cannula at the height of the center of the cassette and re-prime. | INF_TUBING_CAL_OFFSET_ADV Infusion tubing calibration 0 offset advisory. The infusion cannula was not positioned within the correct vertical range of the cassette during calibration. | • Cassette • Flow Sensor • Flow PCB • Fluidics Controller PCB • Fluidics Module |
3470 | Advisory | Calibration verification error: calculated pressures are not within the expected range. IOP Control functions will be disabled. Please re-prime. | INF_TUBING_CAL_CHECK_PT_ADV Infusion tubing calibration check point advisory. The pressure drop calculated using the acquired calibration profile was not within the expected range. | • Cassette • Flow Sensor • Flow PCB • Fluidics Controller PCB • Fluidics Module |
3471 | Advisory | Noisy calibration flow readings. IOP Control functions will be disabled. Please re-prime. | INF_TUBING_CAL_STDDEV_ADV Infusion tubing calibration standard deviation advisory. The standard deviation between the acquired flow measurements and the calculated calibration profile was larger than the specified max limit. | • Wrong infusion cannula selected • Cassette • Flow Sensor • Flow PCB • Fluidics Controller PCB • Fluidics Module |
3472 | Advisory | Infusion chamber leak detected. Please eject and replace the cassette. | INF_CHAMBER_LEAK_ADV – A leak in the infusion chambers was detected during priming. | • Cassette |
3473 | Error | Infusion pressure transducer reference voltage out of range. Infusion/irrigation and extraction functions will be disabled. | INF_TRANSDUCER_REFERENCE_ERR Infusion transducer reference voltage error. | • Cable W88 • Fluidics Infusion PCB • Fluidics Controller PCB • Fluidics Module |
3474 | Error | Infusion NIFS valve error detected. Infusion/ irrigation and extraction functions will be disabled. | INF_NIFS_VLV_ERR Infusion NIFS valve error. | • Cable W88 • Fluidics Infusion PCB • Fluidics Controller PCB • Fluidics Module |
3475 | Advisory | Infusion prime failed. | INF_PRIME_FAILED_ADV Infusion prime failed. | • Infusion line clamped • Cassette drip chamber check valve stuck • Cassette • Cassette Connectors • Stuck AI or BI pincher • Loss of 45psi • 5 Valve PCB, Yellow Lower • Cable W90 • Fluidics Controller PCB • Infusion Level Sensors • Faulty MT5 “S1” Pressure Sensor • Fluidics Module |
3476 | Advisory | The infusion chambers did not fill with fluid. Please check the infusion bottle and connections or press [Change] to replace the infusion bottle. | INF_CHAMBER_FILL_TIMEOUT_ADV Infusion chamber did not fill within the specified timeout period. Source container is out of fluid. | • User Advisory - See Discrepancy description |
3477 | Error | Infusion LPAS pressure transducer offset error. Infusion/irrigation and extraction functions will be disabled. | INF_LPAS_PT_OFFS_ERR Infusion LPAS pressure transducer 0 offset out of range. | • Cable W88 • Fluidics Infusion PCB • Fluidics Controller PCB • Fluidics Module |
3478 | Error | Infusion LPAS pressure error. Infusion/irrigation and extraction functions will be disabled. | INF_LPAS_PRSR_ERR Infusion LPAS source pressure too high. | • Proportional valve • Cable W88 • Fluidics Infusion PCB • Fluidics Controller PCB • Fluidics Module |
3479 | Advisory | Infusion LPAS pressure is low. Infusion pressure loss is possible. | INF_LPAS_PRSR_LOW_ADV Infusion LPAS pump output pressure is low. | • Proportional valve • Cable W88 • Fluidics Infusion PCB • Fluidics Controller PCB • Fluidics Module |
3481 | Advisory | Infusion flow sensor is disconnected. IOP Control functions will be disabled. | INF_FLOW_SENSOR_DISCONNECT_ADV Infusion flow sensor is disconnected or the sensor connection has failed. | • User Advisory - See Discrepancy description |
3482 | Advisory | Auto infusion valve failed to open during prime. Please replace infusion tubing and attempt to re- prime. | The AIV valve flow check performed as part of the cassette prime sequence failed. | • User Advisory - See Discrepancy description |
4100 | Error | U/S voltage failure (+5 analog). Ultrasound and Diathermy functions will be disabled. | US_5_VOLT_ANALOG_ERR - US Kernel Analog 5 volt feedback is bad. US Submodule is non- functional. | • U/S Controller PCB |
4101 | Error | U/S voltage failure (+2.5). Ultrasound and Diathermy functions will be disabled. | US_PLUS_2_5_VOLT_ERR - US Kernel +2.5 volt feedback is bad. US submodule is non-functional. | • U/S Controller PCB |
4102 | Error | U/S voltage failure (-2.5). Ultrasound and Diathermy functions will be disabled. | US_MINUS_2_5_VOLT_ERR US Kernel -2.5 volt feedback is bad. US submodule is non-functional. | • U/S Controller PCB |
4103 | Error | U/S voltage failure (+8.5). Ultrasound and Diathermy functions will be disabled. | US_8_5_VOLT_POWER_ERR US Kernel 2.5 volt feedback is bad. US submodule is non-functional. | • U/S Controller PCB |
4111 | Error | US failure: SPI driver write timeout. Ultrasound and Diathermy functions will be disabled. | SPI_WRITE_TIMEOUT_ERR SPI driver timed out waiting for a write to complete. US submodule is non-functional. | • U/S Controller PCB |
4200 | Advisory | Handpiece EEPROM CRC is invalid. Please replace the handpiece. | US_HP_BAD_CRC_ADV The handpiece EEPROM CRC is invalid. Handpiece needs to be replaced. | • User Advisory - See Discrepancy description |
4201 | Advisory | Only one US handpiece may be connected at a time. Please remove one of the handpieces. | US_HP_MULTIPLE_ADV Two US handpieces are connected. One of the handpieces must be removed before the other handpiece can be used. | • User Advisory - See Discrepancy description |
4202 | Advisory | Handpiece current is too low. Please replace handpiece and re-tune. | US_HP_CURRENT_ERR U/S handpiece current is too low. A short circuit in the handpiece can cause this. | • User Advisory - See Discrepancy description |
4203 | Advisory | Handpiece voltage is too low. Please replace handpiece and re-tune. | US_HP_VOLTAGE_ERR U/S handpiece voltage is too low. An open circuit in the handpiece can cause this. | • User Advisory - See Discrepancy description |
4204 | Advisory | Handpiece power is too high. Please replace handpiece and re-tune. | US_HP_EXCESSIVE_POWER_ERR U/S handpiece power output is too high. | • User Advisory - See Discrepancy description |
4206 | Advisory | Handpiece power DC2DC output was out of range. Please release the footswitch treadle to reset. If problem persists, please contact Field Services. | US_HP_DC2DC_ERR While powering the handpiece, the DC2DC voltage for handpiece power is out of range. | • User Advisory - See Discrepancy description |
4207 | Advisory | Handpiece was removed while powered. Please reconnect handpiece and re-tune. | US_HP_REMOVED_WHILE_POWERED_ADV User has disconnected the handpiece while it’s being powered. | • User Advisory - See Discrepancy description |
4208 | Advisory | Data in handpiece EEPROM is out of range. Please replace the handpiece. | US_HP_OUT_OF_RANGE_DATA_ADV Data contained in the handpiece is out of range. | • User Advisory - See Discrepancy description |
4209 | Advisory | Unknown US handpiece connected. Please connect a known handpiece. | US_HP_UNKNOWN_HANDPIECE_ADV The system has detected that a US handpiece was connected but it cannot determine its type. | • User Advisory - See Discrepancy description |
4210 | Advisory | Unsupported US handpiece connected. Please connect a supported handpiece. | US_HP_UNSUPPORTED_HANDPIECE_ADV The system has detected the connection of a recognized US handpiece but that handpiece is not supported by this system. Handpiece is not tuned. | • User Advisory - See Discrepancy description |
4220 | Advisory | Tune failure – attempted while handpiece was in air. Please re-tune the handpiece. | US_TUNE_IN_AIR_ADV The handpiece was tuned while in air. Handpiece is not tuned. | • User Advisory - See Discrepancy description |
4221 | Advisory | Tune failure: handpiece was removed before tuning. Please connect a handpiece and re-tune. | US_TUNE_NO_HANDPIECE_ADV A handpiece tune was requested but no handpiece is connected. | • User Advisory - See Discrepancy description |
4222 | Advisory | Tune failure: handpiece is an unknown type. Please connect a known handpiece and re-tune. | US_TUNE_UNKNOWN_HP_ADV A handpiece tune was requested but an unknown type of handpiece is connected. | • User Advisory - See Discrepancy description |
4223 | Advisory | Tune failure: handpiece has a loose tip. Please tighten the tip and re-tune. | US_TUNE_LOOSE_TIP_ADV The handpiece tip was loose when tuned. Handpiece is not tuned. | • User Advisory - See Discrepancy description |
4224 | Advisory | Tune failure: handpiece current is low. Please replace handpiece and re-tune. | US_TUNE_HP_CURRENT_ADV The handpiece current was too low (open circuit). Handpiece is not tuned. | • User Advisory - See Discrepancy description |
4225 | Advisory | Tune failure: handpiece voltage is low. Please replace handpiece and re-tune. | US_TUNE_HP_VOLTAGE_ADV The handpiece voltage was too low when tuned (short circuit). Handpiece is not tuned. | • User Advisory - See Discrepancy description |
4226 | Advisory | Tune failure: handpiece frequency order error. Please re-tune. | US_TUNE_FREQ_ORDER_ADV The series (low impedance) and parallel (high impedance) frequencies were out of order when the handpiece was tuned. | • User Advisory - See Discrepancy description |
4228 | Advisory | Tune failure: handpiece series frequency margin error. Please re-tune. | US_TUNE_FS_MARGIN_ERR The series tune frequency was too close to the tune start frequency while tuning the handpiece. The handpiece is not tuned. | • User Advisory - See Discrepancy description |
4229 | Advisory | Tune failure: handpiece parallel frequency margin error. Please re-tune. | US_TUNE_FP_MARGIN_ERR The parallel tune frequency was too close to the tune end frequency while tuning the handpiece. The handpiece is not tuned. | • User Advisory - See Discrepancy description |
4231 | Advisory | Tune failure: handpiece frequency bandwidth too low. Please re-tune. | US_TUNE_BW_LOW_ERR The difference between the series and parallel tune frequencies was too small while tuning the handpiece. The handpiece is not tuned. | • User Advisory - See Discrepancy description |
4232 | Advisory | Tune failure: handpiece frequency bandwidth too high. Please re-tune | US_TUNE_BW_HIGH_ERR The difference between the series and parallel tune frequencies was too large while tuning the handpiece. The handpiece is not tuned. | • User Advisory - See Discrepancy description |
4234 | Advisory | Tune failure: handpiece DC2DC output out of range. Please re-tune. If problem persists, please contact Field Service. | US_TUNE_DC2DC_ERR The DC2DC voltage was out of range while tuning the handpiece. The handpiece is not tuned. | • User Advisory - See Discrepancy description |
4235 | Advisory | Tune failure: handpiece removed while tuning. Please connect handpiece and re-tune. | US_TUNE_HP_REMOVED_ADV The handpiece was removed while tuning. The handpiece is not tuned. | • User Advisory - See Discrepancy description |
4240 | Advisory | The requested Frag continuous power is too high. The Power level will be limited. | US_FRAG_POWER_TOO_HIGH_ADV A request for more than 60% frag power was made while not in a pulsed mode. The power will be limited to 60%. | • User Advisory - See Discrepancy description |
4400 | Advisory | AutoSert* handpiece EEPROM CRC is invalid. Please replace the handpiece. | IOL_BAD_CRC_ADV The AutoSert* handpiece EEPROM CRC is invalid. Handpiece needs to be replaced. | • User Advisory - See Discrepancy description |
4401 | Advisory | AutoSert* handpiece impeded. Retry handpiece operation. If problem persists, please contact Field Service. | IOL_MOTOR_NOT_MOVING_ADV The AutoSert* handpiece motor is not moving – a move check timeout has occurred. | • User Advisory - See Discrepancy description |
4402 | Advisory | AutoSert* handpiece speed out of range. Retry handpiece operation. If problem persists, please contact Field Service. | IOL_SPEED_OUT_OF_RANGE_ADV The AutoSert* handpiece motor speed is out of range. | • User Advisory - See Discrepancy description |
4403 | Advisory | AutoSert* handpiece travel out of range. Reinsert/ replace handpiece. If problem persists, please contact Field Service. | IOL_SPEED_OUT_OF_RANGE_ADV The AutoSert* handpiece motor speed is out of range. | • User Advisory - See Discrepancy description |
4404 | Advisory | AutoSert* handpiece calibration failed. Reinsert/ replace handpiece. If problem persists, please contact Field Service. | IOL_CALIBRATION_FAILED_ADV Calibration of the AutoSert* handpiece failed. Possible causes: 1. The calibration did not complete before the 1 minute calibration timeout 2. The calculated handpiece speed is out of range 3. The handpiece travel is out of range 4. The handpiece motor is not moving | • User Advisory - See Discrepancy description |
4405 | Advisory | AutoSert* handpiece must be in fully retracted position prior to autoclave. Reinsert handpiece to correct. | IOL_HP_IMPROPER_DISCONNECT_ADV The AutoSert* handpiece was disconnected improperly with the plunger not in the retracted position. | • User Advisory - See Discrepancy description |
4406 | Advisory | AutoSert* handpiece data corrupted. Replace handpiece. If problem persists, please contact Field Service. | IOL_CORRUPT_HP_DATA_ADV The AutoSert* handpiece EEPROM contains corrupted data. Handpiece needs to be replaced. | • User Advisory - See Discrepancy description |
4407 | Advisory | Unknown AutoSert* handpiece connected. Please connect a known handpiece. | IOL_UNKNOWN_HP_ADV The system has detected that an AutoSert* handpiece was connected but it cannot determine its type. | • User Advisory - See Discrepancy description |
4408 | Advisory | Unexpected AutoSert* handpiece movement detected. Replace handpiece. If problem persists, please contact Field Service. | IOL_UNEXPECTED_MOTOR_MOVE_ADV Unexpected AutoSert* handpiece movement \ feedback was detected. Handpiece needs to be replaced. | • User Advisory - See Discrepancy description |
4409 | Advisory | AutoSert* handpiece motor over current detected. Retry handpiece operation. If problem persists, replace handpiece and\or contact Field Service. | IOL_MOTOR_OVERCURRENT_ADV AutoSert* handpiece motor over current condition has been detected. | • User Advisory - See Discrepancy description |
4410 | Advisory | Preloading of IOL using the treadle is not allowed until the AutoSert* plunger has been fully retracted. | IOL_TRDLE_NOT_ALLOW_LPSTART_ADV User pressed the treadle to perform IOL Preloading when a Load Plunger command was previously started but not yet completed (and currently paused). | • User Advisory - See Discrepancy description |
4411 | Error | Submodule Failure (1.8 volt). AutoSert* functions will be disabled. | IOL_1_8_VOLTS_IS_BAD_ERR IOL Injector CPLD +1.8 Voltage is bad. | • User Advisory - See Discrepancy description |
4250 | Error | Ultrasound failure: ADC calibration. Ultrasound functions will be disabled. | US_ADC_CALIBRATE_ERR The ADC feedback reading with power off for DC2DC voltage, handpiece voltage, or handpiece current was too high. | • User Advisory - See Discrepancy description |
4300 | Error | Diathermy failure: DC2DC output was out of range. Diathermy functions will be disabled. | DIA_AMPLIFIER_ERR While powering the handpiece, the DC2DC voltage for handpiece power is out of range. | • User Advisory - See Discrepancy description |
4301 | Advisory | Diathermy power is too high. Please release the footswitch treadle / button and try again. | DIA_EXCESSIVE_POWER_ADV Too much power was being delivered to the diathermy handpiece. Diathermy power is turned off. The operator must release the treadle/switch and the depress the treadle/switch to re-activate power. | • User Advisory - See Discrepancy description |
5100 | Error | Ballast failure (voltage). Illuminator functions will be disabled. | ILM_VOLTAGE_SENSOR_ERR - Ballast transducer has reading that is out of specified safety range. | • Cable W50 • Illuminator Controller PCB • Table Top Illuminator Module |
5101 | Error | Ballast failure (current). Illuminator functions will be disabled. | ILM_CURRENT_SENSOR_ERR – Ballast transducer has reading that is out of specified safety range. | • Cable W50 • Illuminator Control PCB • Table Top Illuminator Module |
5102 | Advisory | Lamp calibration data is corrupted: lamp needs to be calibrated. Please contact Field Service. | ILM_CALIBRATION_DATA_ADV – The calibration data saved in the flash contains incorrect CRC checksum. | • Calibration Data corrupt |
5103 | Advisory | Failure to turn lamp on: lamp needs to be replaced. Please contact Field Service. | ILM_LAMP_SOURCE_FAULT_ADV – Lamp status reported by hardware is not the same as software status. | • Lamp • Illuminator Controller PCB • Table Top Illuminator Module |
5104 | Advisory | The lamp has exceeded its rated life: lamp needs to be replaced. Please contact Field Service. | ILM_LAMP_REPLACEMENT_ADV – The lamp has exceeded its rated life. The new lamp is expected. | • Lamp |
5105 | Advisory | The lamp has exceeded its rated maximum life: lamp needs to be replaced. Please contact Field Service. | ILM_LAMP_INTENSITY_LOW_ADV – The lamp has exceeded its rated safe life. The new lamp should be installed immediately. | • Lamp |
5107 | Advisory | The Table-Top Illuminator drawer is ejected. Please close the drawer to continue. | ILM_MODULE_DRAWER_OUT_ADV – The module has been pulled out of drawer. | • User Advisory - See Discrepancy description |
5109 | Information | The calibration data for the Illuminator has changed: due to recalibration or replacement with a new unit. | ILM_MODULE_REPLACED_INFO – The calibration data is changed.. | • User Advisory - See Discrepancy |
5113 | Advisory | Metrics data is corrupted; Lamp hour is invalid. Please contact Field Service. | ILM_METRICS_CORRUPT_ADV – The metrics data for lamp hours is invalid. Either the Flash is corrupted or the checksum is incorrect. | • Illuminator Controller PCB • Table Top Illuminator Module |
5200 | Advisory | Illuminator optics temperature is high. The lamp will be turned off if the temperature continues to rise. | ILM_OPTICS_TEMP_HIGH_ADV – Optics temperature is too high. Lamp is going to be shut down if temperature continues to rise. | • Dirty Air Filter - located below Illuminator • Optics Fan (W29) • Illuminator Controller PCB • Table Top Illuminator Module |
5201 | Error | Illuminator optics temperature has exceeded its limit. Illuminator functions will be disabled. | ILM_OPTICS_TEMP_DOWN_ERR – Illuminator is shut down because Optics temperature is too high. | • Dirty Air Filter - (located below Illuminator) • Optics Fan (W29) • Illuminator Controller PCB • Table Top Illuminator Module |
5202 | Advisory | Illuminator optics fan is at full speed. Optics unit may be overheating. | ILM_OPTICS_FAN_ALARM_ADV – Warning that the Optics fan is full on. | • Dirty Air Filter - (located below Illuminator) • Optics Fan (W29) • Illuminator Controller PCB • Table Top Illuminator Module |
5203 | Error | Illuminator optics thermo-cut-off has been triggered. Illuminator functions will be disabled. | ILM_OPTICS_OVER_TEMP_ERR – Lamp is turned off because thermo cut-off. | • Dirty Air Filter - (located below Illuminator) • Optics Fan (W29) • Optics Thermal Switch • Illuminator Controller PCB • Table Top Illuminator Module |
5204 | Advisory | Communication failure with the Illuminator optics fan. The fan may not work properly. | ILM_OPTICS_FAN_COMMS_ADV – Communication Error with optics fan. | • Communication - Eject and reinsert Illuminator • Cable W81 • Illuminator Controller PCB |
5300 | Advisory | Illuminator ballast temperature is high. The lamp will be turned off if the temperature continues to rise. | ILM_BALLAST_TEMP_HIGH_ADV – Ballast temperature is too high. Lamp is going to be shut down if temperature continues to rise. | • Dirty Air Filter - (located below Illuminator) • Ballast Fan (W28) • Illuminator Controller PCB • Table Top Illuminator Mod |
5301 | Error | Illuminator ballast temperature has exceeded its limit. Illuminator functions will be disabled. | ILM_BALLAST_TEMP_DOWN_ERR – Illuminator is shut down because Ballast temperature is too high. | • Dirty Air Filter - (located below Illuminator) • Ballast Fan (W28) • Illuminator Controller PCB • Table Top Illuminator Module |
5302 | Advisory | Illuminator ballast fan is at full speed. Ballast unit may be overheating. | ILM_BALLAST_FAN_ALARM_ADV – Warning that the Ballast fan is full on. | • Dirty Air Filter - (located below Illuminator) • Ballast Fan (W28) • Illuminator Controller PCB • Table Top Illuminator Module |
5303 | Error | Illuminator ballast thermo-cut-off has been triggered. Illuminator functions will be disabled. | ILM_BALLAST_OVER_TEMP_ERR – Lamp is turned off because thermo cut-off. | • Dirty Air Filter - (located below Illuminator) • Ballast Fan (W28) • Illuminator Controller PCB • Table Top Illuminator Module |
5304 | Advisory | Communication failure with the Illuminator ballast fan. The fan may not work properly. | ILM_BALLAST_FAN_COMMS_ADV – Communication Error with ballast fan. | • Communication - Eject and reinsert Illuminator • Cable W09 • Illuminator Controller PCB • Interface (Floating) PCB |
5400 | Error | Lamp louver failure: unable to move to home position. Left port will be disabled. | ILM_STEP_MOTOR1_HOME_ERR – Step motor at port 1 failed to move to home position. | • Home Sensor PCB. • Home Sensor cable • Port 1 Stepper Motor assy. • Interface PCB • Illuminator Controller PCB • Table Top Illuminator Module |
5401 | Error | Lamp louver failure: unable to move to home position. Right port will be disabled. | ILM_STEP_MOTOR2_HOME_ERR – Step motor at port 2 failed to move to home position. | • Home Sensor PCB. • Home Sensor cable • Port 1 Stepper Motor assy. • Interface PCB • Illuminator Controller PCB • Table Top Illuminator Module |
5402 | Error | Lamp louver failure: unable to move to specified position. Left port will be disabled. | ILM_STEP_MOTOR1_MOVE_ERR – Step motor at port 1 failed to move to specified position. | • Port 1 Stepper Motor Assy • Illuminator Controller PCB • Table Top Illuminator Module |
5403 | Error | Lamp louver failure: unable to move to specified position. Right port will be disabled. | ILM_STEP_MOTOR2_MOVE_ERR – Step motor at port 2 failed to move to specified position. | • Port 2 Stepper Motor Assy • Illuminator Controller PCB • Table Top Illuminator Module |
6101 | Error | Pressure reading is too high. Cutting and Pneumatics functions will be disabled. | PNU_CMN_PRESSURE_TOO_HIGH_ERR – Pneumatics pressure is too high. | • Air Distribution Assy • Pneumatic Controller PCB • Cable W47 • Pneumatic Module |
6103 | Error | Pneumatics pressure transducers reference voltage is unacceptable. Pneumatics functions are disabled. | PNU_CMN_ADC_2048_ERR – Pneumatics ADC 2048 mv. Reference voltage is beyond acceptable range. | • Air Distribution Assy • Pneumatic Controller PCB • Cable W47 • Pneumatic Module |
6104 | Error | Pneumatics Air Distribution transducers offset voltage is unacceptable. Pneumatics functions are disabled. | PNU_CMN_AIR_ADC_OFFSET_ERR – Pneumatics Air Distribution PCB reference offset voltage is beyond acceptable range. | • Air Distribution Assy • Pneumatic Controller PCB • Cable W47 • Pneumatic Module |
6105 | Error | Pneumatics Main Manifold transducers offset voltage is unacceptable. Pneumatics functions are disabled. | PNU_CMN_MAIN_ADC_OFFSET_ERR – Pneumatics Main Manifold PCB reference offset voltage is beyond acceptable range. | • Air Distribution Assy • Pneumatic Controller PCB • Cable W47 • Pneumatic Module |
6201 | Error | Air Pressure valves have high transition faults. Cutting and Pneumatics functions will be disabled. | PNU_AIR_VALVE_FAULT_HI_ERR - AirPressure valves have high level faults (fail to open). | • Air Distribution Assy • Pneumatic Controller PCB • Cable W47 • Pneumatic Modul |
6202 | Error | Air Pressure valves have low transition faults. Cutting and Pneumatics functionals will be disabled. | PNU_AIR_VALVE_FAULT_LO_ERR - AirPressure valves have low level faults (fail to close). | • Air Distribution Assy • Pneumatic Controller PCB • Cable W47 • Pneumatic Mod |
6203 | Advisory | Air Pressure inlet filter may be dirty and needs to be replaced. Please contact Field Service. | PNU_AIR_FILTER_DIRTY_ADV - The unusual high pressure drop cross the Air Filter. This is caused by either the Air Filter is too dirty and need to be replaced or the system may be leaking air. | • Filter needs changing • Air Distribution Assy • Pneumatic Controller PCB |
6204 | Advisory | Inlet pressure is too low; adjust to between 90-120 psi. System may have reduced performance between 58-90 psi.. | PNU_AIR_LOW_SWITCH_ON_ADV - The Air source (wall pressure) is too low (below 58 psi) to turn on. On startup, Pneumatics turns on the pressure automatically if the pressure is between 58 psi and 120 psi. | • User Advisory - See Discrepancy description • Air Distribution Assy • Pneumatic Controller PCB • Cable W47 • Pneumatic Module |
6206 | Advisory | Inlet pressure is too high; adjust to between 90-120 psi. | PNU_AIR_HIGH_SWITCH_ON_ADV - The Air source (wall pressure) is too high (above 120 psi) to turn on. On startup, Pneumatics turns on the pressure automatically if the pressure is between 58 psi and 120 psi. | • User Advisory - See Discrepancy description • Air Distribution Assy • Pneumatic Controller PCB • Cable W47 • Pneumatic Module |
6209 | Advisory | Inlet pressure is below 90 psi; system may have reduced performance. Adjust to between 90-120 psi. | PNU_AIR_PRESSURE_90PSI_ADV - The Air source pressure is below 90 psi. System performance will be diminished. | • User Advisory - See Discrepancy description |
6211 | Advisory | The source pressure transducer is out of specification. Please contact Field Service. | PNU_AIR_SENSOR_CALIBRATED_ADV - The source pressure transducer is out of specification. | • Air Distribution Assy • Pneumatic Controller PCB • Cable W47 • Pneumatic Module |
6301 | Error | Cutting errors; valves have high transition faults (failed to open). Cutting functions will be disabled. | PNU_CUT_VALVE_FAULT_HI_ERR - Cutters valves have high level faults (fail to open). | • Cutter Circuit valves • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6302 | Error | Cutting errors; valves have low transition faults (failed to close). Cutting functions will be disabled. | PNU_CUT_VALVE_FAULT_LO_ERR - Cutters valves have low level faults (fail to close). | • Cutter Circuit valves • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6303 | Error | Cutting error: redundant transducers discrepancy error. Cutting functions will be disabled. Please contact Field Service. | PNU_CUT_REDUNDANT_SENSORS_ERR – Cutters redundant transducers discrepancy error. | • Cutter Circuit Transducers • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6304 | Error | Cutting error: redundant transducers calibration. Cutting functions will be disabled. Please contact Field Service. | PNU_CUT_SENSOR_CALIBRATION_ERR – Cutters redundant transducers calibration error. | • Cutter Circuit Transducers • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6305 | Advisory | Cutting pressure is oscillating beyond the specified range. Please release the footswitch treadle to reset. | PNU_CUT_PRESSURE_OSCIL_ADV – Cutters actual pressure is oscillating beyond the specified tolerance. | • Unstable pressure - possible leak • Proportional Valve • Cutter Circuit Transducers • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6306 | Advisory | Cutting pressure is surging beyond the specified tolerance. Please release the footswitch treadle to reset. | PNU_CUT_PRESSURE_OSCIL_ADV – Cutters actual pressure is oscillating beyond the specified tolerance. | • Unstable pressure - possible leak • Proportional Valve • Cutter Circuit Transducers • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6307 | Advisory | Vitrectomy port transducers are out of tolerance. The Vit Probe is still available. Please contact Field Service. | PNU_CUT_VIT_SENSOR_FAULT_ADV – Vit. port transducers discrepancy error. SmartVit function is disabled. | • Calibration needed • Vit Transducers • Vit Valve • Pneumatic Controller PCB • Pneumatic Module |
6308 | Advisory | Pneumatics Flash is in need of Service. The Vit Probe is still available. Please contact Field Service. | PNU_CUT_VIT_FLASH_FAULT_ADV – The Pneumatics Flash memory is in need of Service. | • Pneumatic Controller PCB • Pneumatic Module |
6401 | Error | Utilities valves have high transition faults (failed to open). Pneumatics functions will be disabled. | PNU_UTL_VALVE_FAULT_HI_ERR - Utilities valves have high level faults (fail to open). | • Utility Circuit valves • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6402 | Error | Utilities valves have low transition faults (failed to close). Pneumatics functions will be disabled. | PNU_UTL_VALVE_FAULT_LO_ERR - Utilities valves have low level faults (fail to close). | • Utility Circuit valves • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6403 | Error | Utilities redundant transducers discrepancy error. Pneumatics functions will be disabled. Please contact Field Service. | PNU_UTL_REDUNDANT_SENSORS_ERR – Utilities redundant transducers discrepancy error. | • Utilities Circuit Transducers • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6404 | Error | Utilities redundant transducers calibration error. Pneumatics functions will be disabled. Please contact Field Service. | PNU_UTL_SENSOR_CALIBRATION_ERR – Utilities redundant transducers calibration error. | • Utilities Circuit Transducers • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6405 | Error | Vacuum redundant transducers discrepancy error. Pneumatics functions will be disabled. Please contact Field Service. | PNU_VAC_REDUNDANT_SENSORS_ERR – Vacuum redundant transducers discrepancy error. | • VFC Vacuum Transducers • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6406 | Error | Vacuum redundant transducers calibration error. Pneumatics functions will be disabled. Please contact Field Service. | PNU_VAC_SENSOR_CALIBRATION_ERR – Vacuum redundant transducers calibration error. | • VFC Vacuum Transducers • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6407 | Error | Pressure is unstable. Release the footswitch treadle or restart AGF. | PNU_UTL_PRESSURE_OSCIL_ADV – Utilities actual pressure is oscillating beyond the specified tolerance. | • Unstable pressure - possible leak • Proportional Valve • Cutter Circuit Transducers • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6408 | Advisory | Utilities pressure is surging beyond the specified tolerance. Please release the footswitch treadle to reset. | PNU_UTL_PRESSURE_SURGE_ADV – Utilities actual pressure is surging beyond the specified tolerance. | • Unstable pressure - possible leak • Proportional Valve • Cutter Circuit Transducers • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6501 | Error | Auto Gas valves have high transition faults (failed to open). Pneumatics functions will be disabled. | PNU_AGF_VALVE_FAULT_HI_ERR - Auto Gas valves have high level faults (fail to open). | • AGF valves • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6502 | Error | Auto Gas valves have low transition faults (failed to close). Pneumatics functions will be disabled. | PNU_AGF_VALVE_FAULT_LO_ERR - Auto Gas valves have low level faults (fail to close). | • AGF valves • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6503 | Error | Auto Gas redundant transducers discrepancy error. Pneumatics functions will be disabled. Please contact Field Service. | PNU_AGF_REDUNDANT_SENSORS_ERR – Auto Gas redundant transducers discrepancy error. | • AGF Transducers • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6504 | Error | Auto Gas redundant transducers calibration error. Pneumatics functions will be disabled. Please remove AGF syringe if connected. | PNU_AGF_SENSOR_CALIBRATION_ERR – Auto Gas redundant transducers calibration error. | • AGF Transducers • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6505 | Advisory | C3F8 bottle may be empty and needs to be replaced. Please press [Replaced] to confirm bottle replacement. | PNU_AGF_GAS_1_EMPTY_ADV – Gas 1 bottle may be empty and needs to be filled up. | • C3F8 Valve • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6506 | Advisory | SF6 bottle may be empty and needs to be replaced. Please press [Replaced] to confirm bottle replacement. | PNU_AGF_GAS_2_EMPTY_ADV - Gas 2 bottle may be empty and needs to be filled up. | • C3F8 Valve • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
6507 | Advisory | Unable to run Auto Gas Filling without sufficient source pressure. | PNU_AGF_NO_AIR_PRESSURE_ADV – There is not enough air pressure to support Auto Gas Filling functions. | • Check Rear panel connection • AGF Transducers • Transducer Interface PCB • Cable W122 • Pneumatic Controller PCB • Pneumatic Module |
7100 | Error | Ballast failure (voltage). Illuminator functions will be disabled. | ILM_VOLTAGE_SENSOR_ERR - Ballast transducer has reading that is out of specified safety range. | • Cable W25 • Illuminator Controller PCB • Auxiliary Illuminator Module |
7101 | Error | Ballast failure (current). Illuminator functions will be disabled. | ILM_CURRENT_SENSOR_ERR – Ballast transducer has reading that is out of specified safety range. | • Cable W25 • Illuminator Controller PCB • Auxiliary Illuminator Module |
7102 | Advisory | Lamp calibration data is corrupted: lamp needs to be calibrated. Please contact Field Service | ILM_CALIBRATION_DATA_ADV – The calibration data saved in the flash contains incorrect CRC checksum. | • Calibration Data corrupt |
7103 | Advisory | Failure to turn lamp on: lamp needs to be replaced. Please contact Field Service. | ILM_LAMP_SOURCE_FAULT_ADV – Lamp status reported by hardware is not the same as software status. | • Lamp • Illuminator Controller PCB • Auxillary Illuminator Module |
7104 | Advisory | The lamp has exceeded its rated life: lamp needs to be replaced. Please contact Field Service. | ILM_LAMP_REPLACEMENT_ADV – The lamp has exceeded its rated life. The new lamp is expected. | • Lamp |
7105 | Advisory | The lamp has exceeded its rated maximum life: lamp needs to be replaced. Please contact Field Service. | ILM_LAMP_INTENSITY_LOW_ADV – The lamp has exceeded its rated safe life. The new lamp should be installed immediately. | • Lamp |
7113 | Advisory | Metrics data is corrupted; Lamp hour is invalid. Please contact Field Service. | ILM_METRICS_CORRUPT_ADV – The metrics data for lamp hours is invalid. Either the Flash is corrupted or the checksum is incorrect. | • Illuminator Controller PCB |
7200 | Advisory | Illuminator optics temperature is high. The lamp will be turned off if the temperature continues to rise. | ILM_OPTICS_TEMP_HIGH_ADV – Optics temperature is too high. Lamp is going to be shut down if temperature continues to rise. | • Dirty air filter. (located below the Illuminator). • Optics fan W20. • Illuminator Controller PCB • Auxillary Illuminator Module |
7201 | Error | Illuminator optics temperature has exceeded its limit. Illuminator functions will be disabled. | ILM_OPTICS_TEMP_DOWN_ERR – Illuminator is shut down because Optics temperature is too high. | • Dirty air filter. (located below the Illuminator). • Optics fan W20. • Illuminator Controller PCB • Auxillary Illuminator Module |
7202 | Advisory | Illuminator optics fan is at full speed. Optics unit may be overheating. | ILM_OPTICS_FAN_ALARM_ADV – Warning that the Optics fan is full on. | • Dirty air filter. (located below the Illuminator). • Optics fan W20. • Illuminator Controller PCB • Auxillary Illuminator Module |
7203 | Error | Illuminator optics thermo-cut-off has been triggered. Illuminator functions will be disabled. | ILM_OPTICS_OVER_TEMP_ERR – Lamp is turned off because thermo cut-off. | • Dirty air filter. (located below the Illuminator). • Optics fan W20. • Optics Thermal Switch • Illuminator Controller PCB • Auxillary Illuminator Module |
7204 | Advisory | Communication failure with the Illuminator optics fan. The fan may not work properly. | ILM_OPTICS_FAN_COMMS_ADV – Communication Error with optics fan. | • Communication - Eject and reinsert Illuminator • Cable W11 • Illuminator Controller PCB • Auxillary Illuminator Module |
7300 | Advisory | Illuminator ballast temperature is high. The lamp will be turned off if the temperature continues to rise. | ILM_BALLAST_TEMP_HIGH_ADV – Ballast temperature is too high. Lamp is going to be shut down if temperature continues to rise. | • Dirty air filter. (located below the Illuminator). • Ballast Fan (W33) • Illuminator Controller PCB • Auxillary Illuminator Module |
7301 | Error | Illuminator ballast temperature has exceeded its limit. Illuminator functions will be disabled. | ILM_BALLAST_TEMP_DOWN_ERR – Illuminator is shut down because Ballast temperature is too high. | • Dirty air filter. (located below the Illuminator). • Ballast Fan (W33) • Illuminator Controller PCB • Auxillary Illuminator Module |
7302 | Advisory | Illuminator ballast fan is at full speed. Ballast unit may be overheating. | ILM_BALLAST_FAN_ALARM_ADV – Warning that the Ballast fan is full on. | • Dirty air filter. (located below the Illuminator). • Ballast Fan (W33) • Illuminator Controller PCB • Auxillary Illuminator Module |
7303 | Error | Illuminator ballast thermo-cut-off has been triggered. Illuminator functions will be disabled. | ILM_BALLAST_OVER_TEMP_ERR – Lamp is turned off because thermo cut-off. | • Dirty air filter. (located below the Illuminator). • Ballast Fan (W) • Illuminator Controller PCB • Auxillary Illuminator Module |
7304 | Advisory | Communication failure with the Illuminator ballast fan. The fan may not work properly. | ILM_BALLAST_FAN_COMMS_ADV – Communication Error with ballast fan. | • Communication - Eject and reinsert Illuminator • cable W11 • Illuminator Controller PCB • Auxillary Illuminator Module |
7400 | Error | Lamp louver failure: unable to move to home position. Left port will be disabled. | ILM_STEP_MOTOR3_HOME_ERR – Step motor at port 3 failed to move to home position. | • Home Sensor PCB. • Home Sensor cable • Port 3 Stepper Motor assy. • Interface PCB • Illuminator Controller PCB • Auxiliary Illuminator Module |
7401 | Error | Lamp louver failure: unable to move to home position. Right port will be disabled. | ILM_STEP_MOTOR4_HOME_ERR – Step motor at port 4 failed to move to home position. | • Home Sensor PCB. • Home Sensor cable • Port 4 Stepper Motor assy. • Interface PCB • Illuminator Controller PCB • Auxiliary Illuminator Module |
7402 | Error | Lamp louver failure: unable to move to specified position. Left port will be disabled. | ILM_STEP_MOTOR3_MOVE_ERR – Step motor at port 3 failed to move to specified position. | • Port Stepper Motor assy. • Interface PCB • Illuminator Controller PCB • Auxiliary Illuminator Module |
7403 | Error | Lamp louver failure: unable to move to specified position. Right port will be disabled. | ILM_STEP_MOTOR4_MOVE_ERR – Step motor at port 4 failed to move to specified position. | • Port 3 Stepper Motor assy. • Interface PCB • Illuminator Controller PCB • Auxiliary Illuminator Module |
8100 | Error | Laser controller software error. Laser functions will be disabled. | LSR_SOFTWARE_ERR - bad process communication | • Reload system software and recalibrate |
8101 | Error | Laser controller shutter open for too long error. Laser functions will be disabled. | LSR_SHUTR_OPN_TOO_LONG_ERR - laser beam not properly detected at port | • Photocell Alignment • Top Sensor PCB |
8102 | Error | Laser controller shutter unexpected error. Laser functions will be disabled. | LSR_SHUTR_UNEXPECTED_ERR - shutter open or position sense problem (at startup) | • Shutter position or sense |
8103 | Error | Laser controller shutter open between firing error. Laser functions will be disabled. | LSR_SHUTR_OPN_BTWN_FIRING_ERR - shutter closed or position sense problem (at startup) | • Shutter jammed mechanically • Top Sensor PCB |
8104 | Error | Laser controller mirror incorrect position error. Laser functions will be disabled. | LSR_MIRROR_INCORRECT_POS_ERR - Sensor reporting both or no Port Mirror position (at startup) | • Port Mirror Switch • Sensor • Top Sensor PCB • Bottom Sensor PCB |
8105 | Error | Laser controller mirror incorrect position error. Laser functions will be disabled. | LSR_MIRROR_POS_UNEXPECTED_ERR - Sensor reporting both or no Port Mirror position (at startup) | • Port Mirror Switch • Sensor • Top Sensor PCB • Bottom Sensor PCB |
8106 | Error | Laser controller power over the limit error. Laser functions will be disabled. | LSR_PWR_OVER_LIMIT_ERR - laser power measure flucuates >20% | • PMON calibration • Power Driver PCB |
8107 | Advisory | Laser controller maximum current exceeded. | LSR_MAX_CURRENT_EXCEEDED_ADV - laser diode current exceeds max limit - | • Temperature Calibration • Laser head; |
8108 | Error | Laser controller port power detection error. Laser functions will be disabled. | LSR_PORT_PWR_DETECTION_ERR - laser power incorrectly detected (during normal run time) | • Photocell beam alignment • Photocells • Top Sensor PCB |
8109 | Error | Laser controller power detected at wrong port error. Laser functions will be disabled. | LSR_PWR_AT_WRONG_PORT_ERR - laser detected at wrong port, or not detected at selected port (during normal run time) | • Port Mirror Switch jammed mechanically; • Sensor • Top Sensor PCB • Bottom Sensor PCB |
8110 | Error | Laser controller power reading mismatch error. Laser functions will be disabled. | LSR_PWR_READING_MISMATCH_ERR - pmons disagree by >20% | • PMONs calibration • Top Sensor PCB; |
8111 | Error | Laser controller LBO crystal limit error. Laser functions will be disabled. | LSR_LBO_CRYSTAL_LIMIT_ERR - LBO temp out of limit | • LBO temp calibration • Mid-Extension PCB; |
8112 | Error | Laser controller pump temperature limit error. Laser functions will be disabled. | LSR_PUMP_TEMP_LIMIT_ERR - laser diode temp out of limit | • Temperature related - Allow unit to reach normal room temp • Mid-Extension PCB; |
8113 | Error | Laser controller power detect digital I/O mismatch error. Laser functions will be disabled. | LSR_PWR_DETECT_DIO_MISMATCH_ERR - laser photo detect digital input problem | • Controller PCB |
8114 | Error | Laser controller footswitch digital I/O mismatch error. Laser functions will be disabled. | LSR_FTSW_DIO_MISMATCH_ERR - footswitch digital inputs problem | • Controller PCB |
8115 | Error | Laser controller footswitch no NC (normally closed) error. Laser functions will be disabled. | LSR_FTSW_NO_NC_ERR - footswitch inputs show both NO and NC active | • Footswitch • Back Panel PCB |
8116 | Error | Laser controller system data CRC error | LC_SYSTEM_DATA_CRC_ERR - calculated CRC incorrect (for one of the .dat data files) | • Temperatures, PMONs, term efficiencies, etc. Calibration, |
8117 | Advisory | Laser controller internal parameter error. | LSR_INTERNAL_PARAMETER_ADV - Bad parameter call | • Wipe memory, reload all software and recalibrate |
8118 | Error | Laser controller process hunt error. Laser functions will be disabled. | LSR_PROCESS_HUNT_ERR - startup process issue | • Wipe memory, reload all software and recalibrate |
8119 | Error | Laser controller process attach error. Laser functions will be disabled. | LSR_PROCESS_ATTACH_ERR - startup process issue | • Wipe memory, reload all software and recalibrate |
8120 | Error | Laser controller process attach error. Laser functions will be disabled. | LSR_START_UP_TIME_OUT_ERR - state machine error | • Wipe memory, reload all software and recalibrate |
8121 | Error | Laser controller process attach error. Laser functions will be disabled. | LSR_JEM_CRC_ERR - crc for JEM incorrect | • Wipe memory, reload all software and recalibrate |
8122 | Error | Laser controller process attach error. Laser functions will be disabled. | LSR_KERNEL_CRC_ERR - crc for Kernel incorrect | • Wipe memory, reload all software and recalibrate |
8123 | Error | Laser controller process attach error. Laser functions will be disabled. | LSR_FLASH_FILE_ERR - crc for Laser App incorrect | • Wipe memory, reload all software and recalibrate |
8124 | Advisory | Laser LBO diode over temperature error | LSR_PUMP_OVERTEMP_ADV - Advisory Only for temperature setpoints | • Recalibrate temperatures |
8125 | Advisory | Laser LBO diode under temperature error. | LSR_PUMP_UNDERTEMP_ADV - Advisory Only for temperature setpoints | • Recalibrate temperatures |
8126 | Advisory | Laser controller crystal over-temperature error. | LSR_CRYSTAL_OVERTEMP_ADV - Advisory Only for temperature setpoints | • Recalibrate temperatures |
8127 | Advisory | Laser controller crystal under-temperature error. | LSR_CRYSTAL_UNDERTEMP_ADV - Advisory Only for temperature setpoints | • Recalibrate temperatures |
8129 | Advisory | Laser controller probe connection error. | LSR_PPP_PROBE_CONNECTION_ADV - Probe Detection; mis-match between sensor pairs | • Sensors or sensor position • Probe Detect PCB |
8131 | Error | Laser controller power monitor POST error. Laser functions will be disabled. | LSR_PMON_POST_FAIL_ERR - pmons fail POST | • PMON calibration • Top Sensor PCB; |
8132 | Error | Laser controller +2.5 volt out-of-range error. Laser functions will be disabled. | LSR_PLUS_2_5_VOLT_ERR - 2.5V on Laser Controller out of range | • Controller PCB |
8134 | Error | Laser controller 12 volt power error. Laser functions will be disabled. | LSR_12_VOLT_POWER_ERR - 12V on Power Driver PCB out of range | • Power Driver PCB |
8135 | Error | Laser shutter timer expired error. Laser functions will be disabled. | LSR_SHUTTER_TIMER_OFF_ERR - shutter timer timed out | • Exposure time calibration • shutter is jammed mechanically • shutter sensor • Top Sensor PCB |
8138 | Error | Laser controller diode thermal electric cooler error. Laser functions will be disabled. | LSR_DIODE_TEC_FAULT_ERR - Main TEC (thermo-electric-cooler) problem | • TEC • Power Driver PCB |
8139 | Error | Laser probe port process POST error. Laser functions will be disabled. | LSR_PPP_POST_FAULT_ERR - unable to switch port mirror or detect position | • Port mirror jammed mechanically • Top Sensor PCB; |
8140 | Error | Laser controller startup timeout error (laser engine). Laser functions will be disabled. | LC_LE_STARTUP_TIMEOUT_ERR - state machine error | • Wipe memory, reload all software and recalibrate |
8141 | Error | Laser controller startup timeout error (shutter control process). Laser functions will be disabled. | LC_SCP_STARTUP_TIMEOUT_ERR - state machine error | • Wipe memory, reload all software and recalibrate |
8142 | Error | Laser controller startup timeout error (supervisor process). Laser functions will be disabled. | LC_SUP_STARTUP_TIMEOUT_ERR - state machine error | • Wipe memory, reload all software and recalibrate |
8143 | Error | Laser controller startup timeout error (peripheral management process). Laser functions will be disabled. | LC_PMP_STARTUP_TIMEOUT_ERR - state machine error | • Wipe memory, reload all software and recalibrate |
8144 | Error | Laser controller startup timeout error (probe port process). Laser functions will be disabled. | LC_PPP_STARTUP_TIMEOUT_ERR - state machine error | • Wipe memory, reload all software and recalibrate |
8145 | Error | Laser controller startup timeout error (laser system controller). Laser functions will be disabled. | LC_LSC_STARTUP_TIMEOUT_ERR - | • Wipe memory, reload all software and recalibrate |
8147 | Error | Laser controller TMP crystal startup timeout error. Laser functions will be disabled. | LC_TMP_CRYSTAL_POST_ERR - LBO temp not stable (on startup) | • Recalibrate temperatures |
8148 | Error | Laser controller TMP diode startup timeout error. Laser functions will be disabled. | LC_TMP_DIODE_POST_ERR - laser diode temp not stable (on startup) | • Recalibrate temperatures |
8149 | Error | Laser controller Dr Filter error. Laser functions will be disabled. | LC_DRFILTER_ERR - reads both NC and NO contacts active | • Dr filter • Rear Panel PCB |
8150 | Error | Laser controller invalid maximum power error. Laser functions will be disabled. | LC_INVALID_MAX_POWER_ERR - Invalid power request | • Wipe memory, reload all software and recalibrate |
8153 | Error | Laser controller load module CRC error. Laser functions will be disabled. | LC_LOADMODULE_CRC_ERR - Invalid software load | • Wipe memory, reload all software and recalibrate |
8154 | Error | Laser controller module requires calibration. Laser functions will be disabled. | LC_CALIBRATION_REQUIRED_ERR - A valid calibration flag is missing | • Calibration |
8155 | Error | Laser Hardware and Software version incompatibility error. Laser functions will be disabled. | LC_VERSION_INCOMPATIBLE_ERR - Version Incompatibility | • Check all Software and Hardware Revision levels, • Wipe memory, reload all software and recalibrate as needed |
8156 | Error | Laser controller power monitor saturation error. Laser functions will be disabled. | LC_PMON_SATURATION_ERR - PMON Voltage problem | • PMON calibration • PMON assembly • Top Sensor PCB |
8200 | Advisory | Laser controller probe removed while firing. | LSR_PRBE_REMOVE_WHEN_FIRING_ADV - Laser controller probe removed while firing. | User Advisory - See Discrepancy description |
8202 | Advisory | Laser controller Dr. Filter disengaged while firing. | LSR_FLTR_DISENG_WHEN_FIRING_ADV - Laser controller Dr. Filter disengaged while firing. | User Advisory - See Discrepancy description |
8203 | Advisory | Laser controller Dr. Filter disconnected while firing. | LSR_FLTR_DISCON_WHEN_FIRING_ADV - Laser controller Dr. Filter disconnected while firing. | User Advisory - See Discrepancy description |
8204 | Advisory | Laser controller interlock opened while firing. | LSR_INTRLK_OPN_WHEN_FIRING_ADV - Laser controller interlock opened while firing. | User Advisory - See Discrepancy description |
8205 | Advisory | Laser controller footswitch disconnected while firing. | LSR_FTSW_DISCON_WHEN_FIRING_ADV - Laser controller footswitch disconnected while firing. | User Advisory - See Discrepancy description |
8207 | Advisory | Laser controller port changed in Ready State. | LSR_PORT_CHANGE_IN_RDYSTE_ADV - Laser controller port changed in Ready State. | User Advisory - See Discrepancy description |
8208 | Advisory | Laser controller Dr. Filter disengaged in Ready State. | LSR_FLTR_DISENG_IN_RDYSTE_ADV - Laser controller Dr. Filter disengaged in Ready State. | User Advisory - See Discrepancy description |
8209 | Advisory | Laser controller Dr. Filter disconnected in Ready State. | LSR_FLTR_DISCON_IN_RDYSTE_ADV - Laser controller Dr. Filter disconnected in Ready State. | User Advisory - See Discrepancy description |
8210 | Advisory | Laser controller interlock opened in Ready State. | LSR_INTRLK_OPN_IN_RDYSTE_ADV - Laser controller interlock opened in Ready State. | User Advisory - See Discrepancy description |
8211 | Advisory | Laser controller footswitch removed in Ready State. | LSR_FTSW_REMOVE_IN_RDYSTE_ADV - Laser controller footswitch removed in Ready State. | User Advisory - See Discrepancy description |
8212 | Advisory | Laser controller footswitch engaged when Ready requested. | LSR_FTSW_ENGAGE_WHEN_RDYREQ_ADV - Laser controller footswitch engaged when Ready requested. | User Advisory - See Discrepancy description |
8213 | Advisory | Laser controller user data CRC error. | LSR_USER_DATA_CRC_ERROR_ADV - Laser controller user data CRC error. | User Advisory - See Discrepancy description |
8214 | Advisory | Laser controller Ready State denied: the filter is disengaged. | LcRdyDeniedFilterDisAdv - Laser controller Ready State denied: the filter is disengaged. | User Advisory - See Discrepancy description |
8215 | Advisory | Laser controller Ready State denied: there is no interlock. | LcRdyDeniedNoInterlockAdv - Laser controller Ready State denied: there is no interlock | User Advisory - See Discrepancy description |
8216 | Advisory | Laser controller Ready State denied: there is no valid probe for the active port. | LcRdyDeniedNoPortSelAdv - Laser controller Ready State denied: there is no valid probe for the active port. | User Advisory - See Discrepancy description |
8217 | Advisory | Laser controller Ready State denied: the footswitch is depressed. | LcRdyDeniedFtswPressAdv - Laser controller Ready State denied: the footswitch is depressed. | User Advisory - See Discrepancy description |
8218 | Advisory | Are all necessary Dr. Filters properly installed and connected? | LcRdyDeniedFltrVerAdv - User requests going to Ready Mode when the doctor filters haven’t been verified. | User Advisory - See Discrepancy description |
8219 | Advisory | Laser controller Ready State denied: there is no footswitch present. | LC_RDY_DENIED_NO_FTSW_ADV - User presses the Laser Ready button when no footswitch is connected. | User Advisory - See Discrepancy description |
8220 | Advisory | Laser controller Dr. Filter connected while in Transition, Ready or Firing State. | LC_DRFILTER_CON_TRF_ADV - User connects an engaged Dr. Filter while in Transition / Ready or Firing mode and and Endo probe is connected to the active port. | User Advisory - See Discrepancy description |
8221 | Advisory | Laser controller detected a port Hardware failure. | LC_PORT_HW_FAILURE_ADV - Laser controller detected a port Hardware failure. | User Advisory - See Discrepancy description |
8222 | Advisory | Port can't be selected: the probe type for the port isn't valid. | LC_PORT_DENIED_INVLD_PROBE_ADV - The user tries to make a laser port the active port, using one of the Purepoint console port selection buttons, but the currently selected probe type for that port is invalid. | User Advisory - See Discrepancy description |
MEDiSURG knows the Alcon Infiniti Vision System is being phased out. But Infiniti is not finished yet!
by Medisurg Staff | Feb 8, 2024 | Alcon, Centurion, Constellation, Infiniti, MEDiSURG, Online Training, Parts, Surgical Set Up, Team MEDiSURG | 0 Comments
In the world of ophthalmology, procedures are always evolving. Having a group of consultants and specialists on hand can be a huge benefit when it comes time to make adjustments to your units, surgical procedures, or find new equipment.The Alcon Infiniti Vision...
Phaco Physics & Fluidics, a series produced by MEDiSURG
by Medisurg Staff | Nov 16, 2023 | Alcon, Centurion, Constellation, Infiniti, MEDiSURG, Surgical Set Up | 0 Comments
Phaco Physics & Fluidics, a video series produced by MEDiSURG, will premiere this month. If you're interested in the most cutting-edge tools and education for eye surgery, subscribe to our channel on YouTube.Modern cataract surgery, known as phacoemulsification,...
MEDiSURG Live Training for Alcon Vision Systems in 2024
by Medisurg Staff | Nov 8, 2023 | Alcon, ALUMNI, Centurion, Constellation, Infiniti, MEDiSURG, Online Training, Surgical Set Up | 0 Comments
In the field of eye surgical equipment, MEDiSURG has been at the forefront for over 20 years. MEDiSURG has everything the ophthalmology community needs for the continual success of practicing eye surgeries, including service, repair, components, upgrades to surgical...
MEDiSURG is offering a 50% discount on the Alcon Centurion Preventative Maintenance Online Course until the end of 2023
by Medisurg Staff | Nov 1, 2023 | Accurus, Alcon, ALUMNI, B&L, Centurion, Constellation, Error Code App, Infiniti, MEDiSURG, Online Training, Surgical Set Up, Team MEDiSURG | 0 Comments
Numerous MEDiSURG alumni students and newcomers to the biotech industry have benefited from our best-selling Preventative Maintenance (PM) for the Alcon Centurion Vision System online course. MEDiSURG is currently providing this course at a 50% discount through the...
MEDiSURG Has All Your Eye Surgical Needs Covered for 2024!
by Medisurg Staff | Oct 27, 2023 | Accurus, Alcon, Centurion, Constellation, Infiniti, MEDiSURG, Online Training, Parts | 0 Comments
With 2023 coming to a close, MEDiSURG is here to help your team with unit updates, parts & repair, and training on how to use eye surgical equipment in preparation for 2024. In many facilities, older, fully functional equipment is still used for eye surgery....